Skip to main content
. 2023 Nov 17;4(3):100428. doi: 10.1016/j.xops.2023.100428

Table 1.

Performance of the Fully Automated AI Model for Detecting nGA in the Combined Validation Test Sets

Training and Tuning Specificity All Visits
At First Development
Sensitivity Specificity Sensitivity Specificity
0.99 0.76 (0.67–0.84) 0.98 (0.96–0.99) 0.53 (0.40–0.65) 0.98 (0.96–0.99)
0.98 0.86 (0.77–0.93) 0.97 (0.95–0.99) 0.75 (0.62–0.87) 0.97 (0.95–0.99)
0.95 0.91 (0.83–0.98) 0.94 (0.91–0.97) 0.87 (0.77–0.97) 0.94 (0.91–0.97)

All values are presented as the estimate (95% confidence interval). AI = artificial intelligence; nGA = nascent geographic atrophy.

Based on threshold determined from the training and tuning sets.

Including all OCT volume scans up until the visit of nGA development, or until the final visit if nGA did not develop in an eye.