Table 1.

Description of the utilised laser and the adjustable and calculated laser operating of carbon dioxide laser. Abbreviations: seconds (s); mL: millilitre; min: minute; mm: millimetre; J: joule; W: watt; cm2: square centimetre.

Device Specifications Manufacturer LX-20SP Nova Pulse, Lumenis, Israel
Wavelength (nm) 10,600
Emission mode Gated continuous wave
Duty cycle (%) 50% duty cycle, 5 s on/5 s off
Delivery system Fibre optic
Energy distribution Gaussian
Beam divergence 8 degrees
Irradiation Parameters Peak power output (W) 1.62
Average power output (W) 0.81
Spot diameter (mm) 0.8
Spot diameter at focus (cm2) 0.005
Spot diameter at tissue (cm) 0.1081
Laser-to-tissue distance (mm) <1
Spot area at tissue (cm2) 0.0092
Total energy (J) 73
Peak fluence (J/cm2) 7934.78
Peak irradiance at spot area (W/cm2) 324.0
Peak irradiance at tissue (W/cm2) 176.08
Average irradiance at spot area (W/cm2) 162
Average irradiance at tissue (W/cm2) 88.04
Speed of the movement (mm/s) ~2
Treatment Protocol Total treatment duration (s) 90
Water irrigation (mL/min) 12
Air (mL/min) 20–25 (Manufacturer instructions)
Aspirating airflow (mL/min) 300 (Manufacturer instructions)