TABLE 2:
Sequence | Plane | Slice Thickness (mm) | Matrix | FOV (mm) | Interslice Distance (%) | Voxel Size (mm) | Acq Time (min:s) | TR/TE (ms) | Accel Fac | Interpolation |
---|---|---|---|---|---|---|---|---|---|---|
| ||||||||||
3D FS T1-weighted MP-RAGE | Sagittal | 0.3 | 224 × 224 | 150 | 50 | 0.3 × 0.3 × 0.3 | 6:32 | 3000/2.49 | GRAPPA 3 | Yes |
FS FLAIR | Axial | 2.0 | 320 × 256 | 220 | 20 | 0.7 × 0.7 × 2.0 | 3:38 | 9000/74 | GRAPPA 3 | No |
3D FLAIR SPACEa | Sagittal | 0.6 | 272 × 272 | 160 | NA | 0.6 × 0.6 × 0.6 | 13:14 | 9000/248 | GRAPPA 3 | No |
FS T2-weighted TSE | Axial | 3.0 | 704 × 704 | 210 | 30 | 0.1 × 0.1 × 3.0 | 6:08 | 3900/61 | GRAPPA 3 | Yes |
SWI | Axial | 1.4 | 640 × 640 | 220 | 20 | 0.2 × 0.2 × 1.4 | 6:28 | 22/15 | GRAPPA 3 | Yes |
3D T2-weighted SPACEb | Sagittal | 0.8 | 224 × 224 | 180 | NA | 0.8 × 0.8 × 0.8 | 7:50 | 5420/118 | CAIPIRINHA 3 | No |
After IV injection of contrast medium (0.05 mL/kg) followed by normal saline flush (10 mL) IV, both at 2 mL/s | ||||||||||
DSC perfusiona,c | Axial | 1.6 | 146 × 146 | 225 | 0 | 1.5 × 1.5 × 1.6 | 3:24 | 1500/24 | Slice accel 3 | No |
DWI RESOLVE | Axial | 2.0 | 192 × 144 | 230 | 20 | 1.2 × 1.2 × 2.0 | 2:49 | 6200/46 and 72 | Slice accel 3 | No |
3D FS T1-weighted MP-RAGE | Sagittal | 0.3 | 224 × 224 | 150 | 50 | 0.3 × 0.3 × 0.3 | 6:32 | 3000/2.49 | GRAPPA 3 | Yes |
3D FS T1-weighted MP-RAGEb,d | Sagittal | 0.6 | 224 × 224 | 150 | 50 | 0.6 × 0.6 × 0.6 | 6:32 | 3000/2.49 | GRAPPA 3 | No |
Note—Total acquisition time is approximately 48 minutes for brain tumor protocol and approximately 30 minutes for radiotherapy planning protocol. Acq = acquisition, Accel Fac = acceleration factor, FS = fat-saturated, NA = not applicable, TSE = turbo spin-echo, SWI = susceptibility-weighted imaging, CAIPIRINHA = controlled aliasing in parallel imaging results in higher acceleration, DSC = dynamic susceptibility contrast, RESOLVE = readout segmentation of long variable echo trains.
Sequence included in both brain tumor protocol and radiotherapy planning protocol.
Sequence included only in radiotherapy planning protocol.
Optional sequence, performed at discretion of ordering clinician and neuroradiologist.
Included 3D distortion correction algorithm.