TABLE 4:

Pituitary Protocol for 7-T Brain MRI With and Without IV Contrast Medium

Sequence Plane Slice Thickness (mm) Matrix FOV (mm) Interslice Distance (%) Voxel Size (mm) Acq Time (min:s) TR/TE (ms) Accel Fac Interpolation

3D FS T1-weighted MP-RAGEa Sagittal 0.3 224 × 224 150 50 0.3 × 0.3 × 0.3 6:32 3000/2.49 GRAPPA 3 Yes
T1-weighted TSEb Sagittal 2.0 384 × 384 160 10 0.4 × 0.4 × 2.0 5:35 500/8.5 GRAPPA 2 No
FS T2-weighted TSEb Coronalc 1.5 704 × 704 210 20 0.2 × 0.2 × 1.5 4:41 4580/61 GRAPPA 3 Yes
T1-weighted TSEb Coronalc 1.5 400 × 320 160 20 0.2 × 0.2 × 1.5 4:03 500/8.7 None Yes
DWI RESOLVEa Axial 2.0 192 × 144 230 20 1.2 × 1.2 × 2.0 2:49 6200/46 and 72 Slice accel 3 No
After IV injection of contrast medium (0.05 mL/kg) followed by normal saline flush (10 mL) IV, both at 2 mL/s
Dynamic T1-weighted FLASHb,d Coronalc 2.0 228 × 228 200 10 0.9 × 0.9 × 2.0 2:59 130/3.23 None No
T1-weighted TSEb Sagittal 2.0 368 × 368 160 10 0.4 × 0.4 × 2.0 5:35 500/8.5 GRAPPA 2 No
T1-weighted TSEb Coronalc 1.5 400 × 320 160 20 0.2 × 0.2 × 1.5 4:03 500/8.7 None Yes
3D FS T1-weighted MP-RAGEa Sagittal 0.3 224 × 224 150 50 0.3 × 0.3 × 0.3 6:32 3000/2.49 GRAPPA 3 Yes

Note—Total acquisition time is approximately 43 minutes. Acq = acquisition, Accel Fac = acceleration factor, FS = fat-saturated, TSE = turbo spin-echo, RESOLVE = readout segmentation of long variable echo trains, accel = acceleration.

a

Whole-brain anatomic coverage.

b

Limited anatomic coverage, focused on the pituitary gland.

c

Oblique plane in orientation of pituitary stalk.

d

Optional sequence, performed at discretion of ordering clinician and neuroradiologist.