Dilution linearity for a plasma sample analyzed with the new TDP-43 assay.

Fold dilution Mean measured concentration (pg/mL) CV (%) Dilution corrected concentration (pg/mL) Recovery rate (%)
Sample 2 79.00 3.7 79.0 100.0
4 46.01 4.9 92.02 116.5
8 24.87 7.6 99.48 125.9
16 14.63 7.5 117.04 148.2
32 8.00 1.2 128.0 162.0

CV = coefficient of variation.