Table 2.

Mechanism of cardiac implantable electronic device-related tricuspid regurgitation by 3D echocardiography

Mode of TR Echocardiographic findings Example
Impingement
  • Normal diastolic excursion of the tricuspid valve leaflet with separation from the PPM lead

  • Systolic apposition of the leaflet and lead with reduced systolic leaflet excursion.

graphic file with name ehad783il1.jpg
Adhesions
  • PPM lead adhesion is diagnosed when the leaflet and lead move together throughout the cardiac cycle.

  • Reduced systolic leaflet excursion.

graphic file with name ehad783il2.jpg
Perforation
  • Defect in the body of the leaflet with PPM lead traversing the defect.

graphic file with name ehad783il3.jpg
Subvalvular
  • Direct interference with chordae resulting in abnormal leaflet closure.

graphic file with name ehad783il4.jpg

CIED, cardiac implantable electronic device; PPM, permanent pacemaker; TR, tricuspid regurgitation.