Abstract
Vapor pressure (psat) data are needed to assess the potential use of terpenes as breath markers of recent cannabis use. Herein, a recently introduced gas-saturation method for psat measurements, known as dynamic vapor microextraction (DVME), was used to measure psat for the terpene (±)-3,7-dimethylocta-1,6-dien-3-ol, commonly known as linalool. The DVME apparatus utilizes inexpensive and commercially available components, a low internal volume, and helium carrier gas to minimize nonideal mixture behavior. In the temperature range from 314 K to 354 K, DVME-based measurements of the psat of linalool ranged from 81 Pa to 1250 Pa. With a measurement period of 30 min, the combined standard uncertainty of these measurements ranged from 0.0358·psat to 0.0584·psat, depending on temperature. The DVME-based measurements agree with a Wagner correlation of available literature data. We demonstrate that DVME produces accurate results for values of psat that are 200 times higher than in the DVME validation study with n-eicosane (C20H42). The oxidative stability of linalool was improved by the addition of 0.2 mass % of the antioxidant tert-butylhydroquinone.
Graphical Abstract

1. Introduction
Dynamic vapor microextraction (DVME) was recently introduced1 as a new type of gas-saturation method2–10 for the measurement of saturated vapor pressure (psat) near 1 Pa. The DVME method was designed to (1) use only commercially available components, (2) minimize operating costs, and (3) achieve state-of-the-art measurement uncertainty by leveraging a careful assessment of the sources of uncertainty. DVME was validated previously with psat measurements on the linear alkane n-eicosane (C20H42) from 344 K to 374 K, where the psat ranged from 0.45 Pa to 5.8 Pa.1 Eicosane is a useful vapor pressure standard because it is stable, non-hygroscopic, can be purchased in high purity, and reference correlations for the psat of n-eicosane exist in the literature.1, 11–13
The goal of the current study was to use DVME to collect psat data for a cannabis-associated monoterpene. We believe that such psat data are a prerequisite for the potential use of cannabis-associated terpenes as breath markers of recent cannabis use. Currently, breathalyzer development strategies typically involve quantitative measurement of Δ9-tetrahydrocannbinol (THC), the main psychoactive cannabinoid in cannabis sativa. However, THC’s extremely low volatility2 requires concentrating breath aerosols from multiple breaths into filter devices for offline analysis with specialized instrumentation such as liquid chromatography with tandem mass spectrometry.14–19 While detecting THC in breath may provide insight into recent cannabis use, it is likely that the detection of multiple cannabis-associated compounds in breath would provide a more definitive identifier of impairment.
Many cannabis-associated terpenes do not have published psat data near body temperature because there are no commercial devices that measure psat in this pressure range. Terpenes are also less stable2, 4 than n-eicosane, which creates an additional barrier to good psat measurements. Difficulties notwithstanding, psat data for these compounds are useful because they define the upper limits of concentration in the vapor phase of breath. The compound chosen for this work, (±)-3,7-dimethylocta-1,6-dien-3-ol (racemic mixture), commonly known as linalool, is an unsaturated monoterpene alcohol that is present in cannabis sativa20, 21 and hemp22. Reliable psat data already exist for linalool,23 which is an advantage in this study because DVME has not been proven reliable for substances more challenging than simple alkanes. Unlike n-eicosane, linalool is prone to oxidation24 and is somewhat hygroscopic (see below).
In addition to cannabis breathalyzer development, psat measurements of linalool and other compounds in this psat range are of interest because indoor air exposure can be estimated from such data.25–27 There is also a more general need for a robust technique to provide low-uncertainty psat data for compounds like terpenes24, 28 and cannabinoids2, 29, which have oxidatively susceptible moieties such as hydroxyl groups and C=C double bonds.4, 30
All gas-saturation methods employ an inert carrier gas that flows through a “saturator” where the carrier gas becomes saturated with analyte vapor before flowing through a “trap” where the analyte is condensed and removed from the carrier gas stream. Helium is the carrier gas used for DVME because it minimizes measurement uncertainty due to nonideal mixture behavior and because of its low solubility in the liquid phase.5, 11, 31 One of the unique features of DVME is that the total internal volume of the carrier gas flow path is only about 4 mL.1 The small internal volume was achieved by making the saturator from a 2 mL gas chromatography (GC) autosampler vial filled with 1 mm glass beads coated in the analyte (with a second, empty GC vial to prevent upstream contamination by the analyte1) and making the helium transfer lines and capillary vapor trap from fused silica capillary tubing with an inner diameter (ID) of 530 μm. A key advantage of the small internal volume of the DVME apparatus is that less total carrier gas flow is required to achieve a given relative uncertainty in the total flow; thus, short measurement periods can still provide low-uncertainty psat data.1
2. Experimental Methods
2.1. Chemicals.
Table 1 shows purities of the HPLC grade acetone, tert-butylhydroquinone (TBHQ), research grade 6 helium, n-tetradecane (C14, olefin-free), analytical grade linalool (racemic mixture, lot BCCB5318, Millipore Sigma), purchased from commercial sources and used as received (except linalool).
Table 1.
Grade and supplier-estimated purity for each chemical. The GC-FID purity for C14 and linalool was determined on each day of psat measurements, so the range of results is given (peak-area basis; assumed to correspond to mole % purity for psat calculations).
| Chemical (CAS #) | Grade/Info | Supplier’s Purity | GC-FID Purity |
|---|---|---|---|
| Linalool (78-70-6) | Analytical Standard | 99.6 % | 99.6 – 99.7 % |
| Acetone (67-64-1) | HPLC grade | 99 % | n/a |
| TBHQ (1948-33-0) | n/a | 97 % | n/a |
| C14 (629-59-4) | Analytical Standard; Olefin-free | ≥99.0 % | 99.1 – 99.3 % |
| Helium (7440-59-7) | Grade 6 | 99.9999 % | n/a |
Because GC-FID is insensitive to water, we also analyzed the linalool by coulometric Karl Fischer titration, which showed a water impurity of 4 mol % in the as-received material. This water impurity was removed from the linalool with a pretreatment step before psat measurements (see Section 2.4.).
2.2. DVME apparatus.
The DVME apparatus is shown in Figure 1. The carrier gas flows from a pressurized helium cylinder through the mass flow controller (MFC, used to set the flow rate), an adsorbent tube filled with the porous polymer poly-2,6-diphenyl-1,4-phenylene oxide, a transfer vial (to prevent upstream contamination), a transfer capillary, the saturator vial (where the carrier gas becomes saturated with linalool vapor), a capillary vapor trap (where the linalool vapor condenses), and finally through the mass flow meter (MFM, used to measure total flow). Saturators were made from 2 mL GC autosampler vials containing 1 mm borosilicate glass beads coated with linalool. The transfer capillary was cut from a spool of deactivated fused silica tubing (530 μm ID). Capillary vapor traps were cut from a spool of Al2O3-coated PLOT GC column (530 μm ID, film thickness 0.25 μm, 1.5 m in length). Most of the capillary trap was wound in a circle and secured on the thermoelectric cooling plate with aluminum tape and foam insulation. A repurposed GC oven controlled the saturator temperature and a 100 Ω platinum resistance thermometer (PRT) measured the saturator temperature. A barometer was used to determine the ambient pressure and the overpressure in the saturator vial.
Figure 1.

The DVME apparatus used to measure the vapor pressures of linalool.
2.3. Preparation and analysis of samples for gas chromatography with flame ionization detection (GC-FID).
Samples for GC-FID analysis were prepared by mass; volumes are given only for the reader’s convenience. The balance had a repeatability of 0.02 mg and an uncertainty of 0.1 mg. The GC-FID method for sample analysis employed a HP-5ms Ultra Inert GC column (17 m, 320 μm ID, film thickness 0.25 μm); an injection volume of 1 μL with a 20:1 split; an inlet temperature of 573 K at 34.5 kPa with a split flow of 28.0 mL/min; an initial oven temperature of 373 K with a ramp of 15 K/min to 463 K (total run time of 6 min); and FID settings of 523 K and 69 kPa.
2.4. Linalool saturator vial preparation.
A 1 mL ampule of linalool was opened, and 0.2 mass % of TBHQ was added to the full volume of linalool. The TBHQ was added as an antioxidant and is discussed further in Section 3.3.4 This mixture was stored in a −20 °C freezer in an amber GC vial and used within a week of preparation. The saturator was prepared by filling a 2 mL autosampler vial with 1.25 – 1.5 mL of glass beads (1 mm diameter) and then adding 0.10 g – 0.15 g of the linalool + TBHQ mixture with a glass pipette. The vial was capped, and the beads were shaken to ensure an even coating of linalool. The cap was then removed and replaced with a clean cap; this was done to prevent physical transfer of linalool from the septum cap to the capillary vapor trap.1 A fresh linalool saturator was prepared for each day of measurements. To remove the water impurity in the linalool, the saturator was pretreated by flowing 700 scc of warm (354 K) helium through it at 10 sccm before making psat measurements (see Supporting Information, Section 1 and Figure S1).
2.5. Vapor pressure measurements.
The working equations for DVME are shown in eqs 1–4:1, 32
| (eq 1) |
| (eq 2) |
| (eq 3) |
| (eq 4) |
The psat of linalool was calculated from the pressure in the saturator vial (psaturator), the mole fraction of linalool in the saturated vapor phase (y2), and the mole fraction of linalool in the liquid phase in the saturator (x2), (eq 1). The value of psaturator was not determined directly but was instead calculated as the sum of the ambient pressure (pambient) and the overpressure (poverpressure) (eq 2). The determination of is poverpressure described in section 2.6. For the calculation of y2 (eq 3), the molar masses of helium (M1 = 4.002602 g/mol) and linalool (M2 = 154.2493 g/mol) were taken from the NIST Chemistry WebBook.33 The total mass of helium gas (m1) for each experiment was determined from the MFM, where 1.0 standard cubic centimeter (scc) is the mass of 1.0 cm3 of helium at 298.15 K and 1.0 atm (i.e., 1.0 scc of helium is 0.16353 mg of helium34). The mass of condensed linalool in the capillary vapor trap (m2) was determined by GC-FID using an internal standard of n-tetradecane (C14).35 For the calculation of the mole fraction purity of linalool, x2 (eq 4), the helium solubility in linalool (x1) was estimated to be 0.00017, which is actually the helium solubility in 1-decanol31 at 314.49 K and 1.0 atm (used instead due to the lack of helium solubility data in linalool). The value for 1-decanol was chosen because, like linalool, it is a ten-carbon alcohol. Excluding water and helium, the mole fraction of impurities in linalool, ximpurities, was estimated from the GC-FID purity analysis of linalool plus the 0.2 mass % of TBHQ (~0.002 mol fraction) added as an antioxidant4.
Each psat measurement began with a 10 min thermal equilibration period, during which the helium flow rate was 0.20 standard cubic centimeters per minute (sccm).1 Then the flow rate was increased to 6.00 sccm for (20 ± 1) min, for a total helium flow of approximately 120 scc (i.e., ~20 mg of helium). Control experiments showed no influence of flow rate on the psat measurements between 2.00 sccm and 10.00 sccm. The temperature was recorded at the midpoint of the experiment (after 60 scc of total flow). At the end of the flow period, the helium flow was stopped, the total flow at the MFM was recorded, and the capillary vapor trap was immediately removed from the DVME apparatus.
The condensed linalool in the capillary vapor trap was eluted immediately after it was removed from the DVME. This was done by using helium pressure to force 1.3 mL of acetone through the trap and into a GC vial containing 0.2 mL of C14 internal standard solution. The internal standard solution was created by adding 30 μL of C14 to 3.5 mL of acetone (approximately 8,500 ppm by mass or 4.2×10−8 mol/kg). A second elution was done with 1.5 mL of acetone into an empty autosampler vial. Both eluent fractions were analyzed by GC-FID. The concentration of linalool in the second fraction never exceeded 0.1 % of the concentration in the first fraction, which demonstrates complete linalool recovery. The capillary vapor trap was rinsed with 1 mL of acetone and left to dry under flowing helium for at least 5 minutes before being used again. Two capillary vapor traps were made for each day of measurements. Each trap was used for two or three psat measurements and then discarded. The relative sensitivity of GC-FID to linalool and C14 (mass-to-area ratios) was used to determine the mass of linalool eluted from the capillary vapor trap. GC-FID showed a linear response across all sample concentrations35 (see SI, Section 2, Figure S2, Table S1).
2.6. Determination of overpressure in the saturator vial.
A separate experiment was conducted to determine the overpressure (poverpressure, i.e., pressure build-up) in the saturator vial due to viscous flow through the capillary vapor trap. We did not directly measure psaturator during psat measurements for two reasons. First, such an arrangement creates an additional puncture in the septum of the saturator vial, which has the potential to leak. Second, and more important, is that the barometer capillary creates dead volume that is mostly at room temperature. Thus, the barometer capillary could act as a vapor trap during psat measurements. Therefore, we measured poverpressure after psat measurements were completed for the day, and then calculated psaturator for each psat measurement from eq. 2.
To measure poverpressure the barometer inlet was fitted with a septum and connected to the saturator vial (with linalool present) with a fused silica capillary. Besides the connection between the saturator and the barometer, the DVME apparatus was configured in the same way as for a psat measurement, with the thermoelectric cooling plate at 263 K, a flow rate of 6.00 sccm, and one of the capillary vapor traps used for that day of psat measurements. No flow was lost to the barometer, as verified by the same flow rate on the MFC and the MFM.
At each temperature setpoint, the saturator pressure (psaturator) and temperature were recorded after 6 min of thermal equilibration. The barometer was also used to measure ambient pressure (pambient) at the beginning and end of the temperature ramp. The average value of pambient was subtracted from the psaturator to determine poverpressure (eq 2). An exponential fit to a representative set of poverpressure vs. temperature data (from 18 NOV 2021; see SI Section 3, Figure S3) for a 530 μm ID capillary vapor trap is given in eq 5,
| (eq 5) |
Finally, as shown in eq 2, the value of psaturator for each psat measurement was calculated from the average value of pambient (as measured with the barometer during the psat measurement) plus the value of poverpressure from the curve fit of the poverpressure vs. temperature data for that day.
2.7. Analysis of measurement uncertainties in the experimental variables.
A detailed analysis of uncertainties in temperature (uT), mass of helium carrier gas (um1), mass of trapped linalool (um2), pressure in the saturator vial (upsaturator), linalool purity (uimpurities), and helium solubility (ux1) is available in our initial publication on DVME1. Section 4 of the SI describes how the uncertainties for these variables, which are given in Tables 2 and 3, were derived from the specific parameters of the linalool psat measurements.
Table 2.
DVME-based vapor pressure (psat) measurements for linalool from 314 K – 354 K. The values of psat were calculated with eq 1–4 from the mass of helium carrier gas (m1), the mass of trapped linalool vapor (m2), and the pressure in the saturator vial (psaturator). All measurements were made with a flow rate of 6.00 sccm and a thermal equilibration period of 10 min. Absolute standard (k=1) uncertainties are given for each experimental variable (see Section 2.7). The absolute combined standard uncertainty for psat (upsat) was calculated as ucombined· psat from the relative combined standard uncertainties in Table 3. For ease of comparison, correlated vapor pressures from eq 6 (pcorr) are also given.
| Date | uT (K) | um1 (mg) | um2 (mg) | upsaturator (kPa) | upsat (Pa) | pcorr (Pa) | 100*(psat−pcorr)/pcorr |
|---|---|---|---|---|---|---|---|
| 21 SEP 2021 | 0.11 | 0.36 | 0.004 | 0.21 | 4.8 | 88.0 | −6.15% |
| 0.12 | 0.37 | 0.013 | 0.21 | 7.0 | 183.6 | −5.59% | |
| 0.13 | 0.36 | 0.036 | 0.21 | 13 | 364 | −5.21% | |
| 0.15 | 0.36 | 0.039 | 0.21 | 23 | 679 | −6.18% | |
| 0.16 | 0.36 | 0.081 | 0.21 | 42 | 1215 | −6.40% | |
| 23 SEP 2021 | 0.11 | 0.36 | 0.006 | 0.21 | 5.1 | 88.0 | −0.35% |
| 0.12 | 0.36 | 0.010 | 0.21 | 7.2 | 183.7 | −3.58% | |
| 0.13 | 0.36 | 0.021 | 0.21 | 13 | 363 | −3.85% | |
| 0.15 | 0.36 | 0.033 | 0.21 | 23 | 679 | −4.38% | |
| 0.16 | 0.36 | 0.065 | 0.21 | 43 | 1215 | −3.84% | |
| 10 NOV 2021 | 0.11 | 0.36 | 0.010 | 0.21 | 4.8 | 88.1 | −6.94% |
| 0.12 | 0.37 | 0.024 | 0.21 | 7.3 | 184.2 | −2.21% | |
| 0.13 | 0.36 | 0.045 | 0.21 | 13 | 364 | 0.18% | |
| 0.15 | 0.36 | 0.082 | 0.21 | 24 | 682 | −0.24% | |
| 0.16 | 0.36 | 0.173 | 0.21 | 45 | 1219 | 0.95% | |
| 15 NOV 2021 | 0.11 | 0.37 | 0.013 | 0.21 | 4.9 | 88.4 | −5.70% |
| 0.12 | 0.37 | 0.032 | 0.21 | 7.4 | 184.4 | −1.41% | |
| 0.13 | 0.37 | 0.044 | 0.21 | 13 | 365 | −2.39% | |
| 0.15 | 0.37 | 0.084 | 0.21 | 24 | 683 | −1.52% | |
| 0.16 | 0.37 | 0.158 | 0.21 | 44 | 1221 | −2.09% | |
| 17 NOV 2021 | 0.11 | 0.37 | 0.011 | 0.21 | 5.1 | 88.3 | −0.52% |
| 0.12 | 0.37 | 0.023 | 0.21 | 7.5 | 184.3 | −0.07% | |
| 0.13 | 0.37 | 0.047 | 0.21 | 13 | 365 | 1.74% | |
| 0.15 | 0.37 | 0.084 | 0.21 | 25 | 682 | 1.25% | |
| 0.16 | 0.37 | 0.167 | 0.21 | 46 | 1220 | 2.35% | |
| 0.11 | 0.37 | 0.010 | 0.21 | 5.0 | 88.3 | −3.68% | |
| 18 NOV 2021 | 0.11 | 0.37 | 0.011 | 0.21 | 4.9 | 88.3 | −5.84% |
| 0.12 | 0.36 | 0.021 | 0.21 | 7.3 | 184.4 | −2.42% | |
| 0.13 | 0.37 | 0.044 | 0.21 | 13 | 365 | −0.70% | |
| 0.15 | 0.37 | 0.083 | 0.21 | 24 | 682 | 0.06% | |
| 0.16 | 0.36 | 0.145 | 0.21 | 45 | 1220 | −0.43% | |
| 0.11 | 0.37 | 0.010 | 0.21 | 4.8 | 88.4 | −7.66% |
Table 3.
Uncertainty budget for DVME-based vapor pressure (psat) measurements for linalool. Seven sources of uncertainty in the psat measurement are considered (see Sections 2.7 and 2.8). The relative combined standard (k=1) uncertainty (ucombined) in psat is the quadrature sum of all these sources of uncertainty.
| Source of Uncertainty | 314 K | 324 K | 334 K | 344 K | 354 K |
|---|---|---|---|---|---|
| m 1 | 0.0178·psat | 0.0177·psat | 0.0179·psat | 0.0179·psat | 0.0177·psat |
| m 2 | 0.0124·psat | 0.0126·psat | 0.0124·psat | 0.0112·psat | 0.0119·psat |
| T | 0.0539·psat | 0.0337·psat | 0.0284·psat | 0.0282·psat | 0.0295·psat |
| eq 1 simplifications | 0.005·psat | 0.005·psat | 0.005·psat | 0.005·psat | 0.005·psat |
| p saturator | 0.0023·psat | 0.0023·psat | 0.0023·psat | 0.0023·psat | 0.0023·psat |
| x impurities | 0.0029·psat | 0.0029·psat | 0.0029·psat | 0.0029·psat | 0.0029·psat |
| x 1 | 0.00017·psat | 0.00017·psat | 0.00017·psat | 0.00017·psat | 0.00017·psat |
| u combined | 0.0584·psat | 0.0406·psat | 0.0364·psat | 0.0358·psat | 0.0369·psat |
2.8. Uncertainty budget for psat.
The uncertainty budget given in Table 3 shows the uncertainty in the psat measurement that results from uncertainty in each experimental variable. The effects of uT, um1, um2, upsaturator, uimpurities, and ux1 on the value of psat were determined from a sensitivity analysis. For example, the Wagner correlation (see Table 4 and eq 6) was used to determine the vapor pressure at pcorr(T) and pcorr(T+uT). The fractional difference between the two values, calculated as |(1-(pcorr(T)/pcorr(T+uT))|, is reported in Table 3 as the uncertainty in the psat measurement that results from uncertainty in the experimental temperature. For the other experimental variables, the sensitivity analysis was performed with eqs 1–4. The low solubility of helium31 means that ux1 is an insignificant source of uncertainty in the psat measurement, but we report it for completeness. In Table 3, the entry called “eq 1 simplifications” refers to the absence of the Poynting correction or the effect of the vapor-phase fugacity coefficient. For helium, these are expected roughly cancel each other32 but a systematic error of about 0.005·psat is introduced by ignoring them.
Table 4.
Coefficients for Wagner correlations of linalool vapor pressure (pcorr) where Tc = 668 K, pc = 2907 kPa, τ = 1 – T/Tc, and p0 = 1 kPa. Eq 6 is a Wagner correlation of all literature data from 270 K to 425 K (excluding the present work), eq 7 is a Wagner correlation of the same literature data plus the DVME data, and eq 8 is a Wagner correlation of the DVME data alone.
| pcorr = p0 · exp((Tc/T)·(Aτ+Bτ1.5+Cτ2.5+Dτ5)+ln(pc/p0)) | |||
|---|---|---|---|
| Coefficient | Eq 6 | Eq 7 | Eq 8 |
| A | −14.882 | −14.395 | −14.617 |
| B | 18.521 | 17.472 | 15.039 |
| C | −20.822 | −20.137 | −12.121 |
| D | 2.4678 | 2.5165 | −11.851 |
3. Results and Discussion
One of the primary goals of this work was to establish that DVME can provide low-uncertainty measurements at higher values of psat than the maximum of ~6 Pa reported in our previous work with n-eicosane.1 For gas-saturation methods, nonideal mixture behavior limits psat measurements to ≤7000 Pa.8 Measurements near this high-pressure limit generate relatively large quantities of analyte vapor, which is a consideration for instrument design and operational parameters. This is especially important given that DVME uses much smaller capillary vapor traps and saturators than conventional designs. The present work demonstrates psat measurements up to about 1250 Pa with linalool. Note that the use of DVME at lower pressures does not require the same kind of demonstration because the ability to measure lower values of psat is mainly a function of the total carrier gas flow and the sensitivity of the analytical method used to determine the recovered mass of analyte (e.g., GC-FID, GC-MS, etc.).
3.1. Plugging, breakthrough, and overpressure.
One challenge with using DVME to measure higher psat values is that the capillary vapor trap can become plugged with condensate. Even a thin coating of condensate on the inner wall of the capillary is visible to the eye—it appears as a section of darkened capillary (compared to the lighter, tan color of clean capillary). During psat measurements, a stationary band of darkened capillary is typically observed just outside the oven where the capillary temperature drops to ambient. However, during the initial linalool experiments with a 320 μm ID capillary vapor trap (as used for the n-eicosane study1), we visually observed multiple darker bands moving along the capillary in the direction of gas flow. This suggests that liquid condensate had entirely filled sections of the capillary trap and that slugs of liquid linalool were being pushed along by the carrier gas. This is undesirable because it creates a significant flow restriction that is not accounted for in the psaturator determination. Increasing the capillary vapor trap ID from 320 μm to 530 μm eliminated the formation of liquid slugs.
The relatively high psat of linalool raised concerns about vapor trapping efficiency. Consequently, we increased the capillary length to 1.5 m (from 1.0 m used in the n-eicosane study1), and a thermoelectric cooling plate was installed to cool the capillary vapor trap to 263 K (Figure 1). Sufficient trapping efficiency was verified in triplicate at 354 K with a total flow of 240 scc and a flow rate of 10 sccm—more demanding conditions than any of the psat measurements. During helium flow, the outlet of the capillary vapor trap was immersed in a “breakthrough vial” containing 1 mL of acetone (instead of being inserted into the MFM) to collect any linalool that was not trapped. Afterward, the acetone in the breakthrough vial was analyzed by GC-FID (Figure 2). No linalool was detected in the breakthrough vial for any of the three replicates, demonstrating the chilled stationary phase in the capillary vapor trap operated at 100% efficiency under conditions that were more extreme than the actual psat measurements.
Figure 2.

GC-FID chromatograms demonstrating the absence of linalool degradation and the trapping efficacy of the capillary vapor trap when cooled to 263 K. Chromatograms are shown for the following samples: (1) neat acetone solvent; (2) the eluent of a capillary vapor trap after a typical psat measurement at 334 K with a flow rate of 6.00 sccm and a total flow of 120 scc; (3) a breakthrough experiment performed at 354 K with a flow rate of 10.00 sccm and a total flow of 240 scc; (4) a purity check of linalool from the saturator vial after a series of psat measurements over the entire temperature range.
The increase in capillary vapor trap ID had a secondary benefit: poverpressure, (eq 2), was reduced by a factor of three. Given that poverpressure is caused by viscous flow through the capillary vapor trap, it varies with flow rate, temperature, and capillary ID and length. Figure 3 shows poverpressure curves from 304 K to 364 K with a 320 μm or a 530 μm ID capillary at a helium flow rate of 6.00 sccm. Smaller poverpressure values were observed with the 530 μm ID capillary, which makes helium leaks less likely, makes the overpressure correction smaller, and potentially allows for significantly higher flow rates than were used herein.
Figure 3.

The overpressure (poverpressure) in the saturator vial versus oven temperature for capillary vapor traps with an ID of 320 μm (triangles) or 530 μm (circles). See Figure S4 of the SI for a plot of the overpressure data for the capillary vapor trap with 530 μm ID with an expanded y-axis.
3.2. Sample Purity.
Sample purity is important for obtaining high-quality psat measurements. The presence of a volatile impurity in linalool was initially inferred from the observation that the first psat measurement with a new saturator vial always yielded an anomalously low value. The identity of the impurity, which did not appear in the GC-FID chromatogram, was verified as water by coulometric Karl Fisher titration. The initial water content in a freshly opened ampule of linalool was approximately 4 mol %, which depresses the measured psat by about 4 %. Fortunately, the relatively high psat of water allowed it to be removed by pretreatment with a flow of warm helium. The required quantity and temperature of helium for the drying procedure was informed with a simple numerical model of the drying process (see SI, section 1). The effectiveness of the drying procedure was then confirmed with replicate measurements of psat, which showed that, after the drying procedure, the initial psat measurement with a new saturator vial was not significantly lower than later replicates (e.g., see the 314 K replicates on 18 NOV 2021 and 17 NOV 2021 in Table 2).
3.3. Sample stability.
DVME allows for relatively short measurement periods (30 min for the measurements reported herein), which is advantageous for psat measurements with compounds such as linalool, which are less stable than alkanes.2, 4 However, instability is difficult to predict and it is convenient to make multiple measurements on the same sample material. Therefore, three precautions against sample decomposition were taken for the measurements on the linalool. First, to minimize the potential for oxidative decomposition, 0.2 mass % of the antioxidant TBHQ was added to the linalool before the psat measurements. The specific decomposition pathway of concern was the autoxidation of the C=C double bonds.4, 30 The small decrease in measured psat caused by the addition of TBHQ was corrected via the ximpurities term (eq 4). The addition of an antioxidant has been used successfully in the past for psat measurements of unsaturated molecules.4 The second precaution was to measure from the lowest to the highest temperature, then repeat the initial 314 K (lowest) data point at the end of the measurement series.4, 36, 37 These replicate data points, shown in Table 2, were within 3.2 % of the initial values, which is less than the combined standard uncertainty (see Section 3.4) of the psat measurement. Third, a given saturator was used for only a single day (5 or 6 measurements). With these precautions in place, no evidence of linalool decomposition was observed, even in the linalool remaining in the saturator vial after a complete series of psat measurements (Figure 2, panel 4).
3.4. Linalool psat measurements, uncertainty budget, correlations, and enthalpy of vaporization.
The psat of linalool was measured by the DVME method over two months, Table 2. The absolute uncertainties in each variable are given in Table 2 (see Section 2.7 for more details about how these were derived). A full uncertainty budget for the psat measurements is given in Table 3, as described in Section 2.8.
The psat measurements were repeatable over daily, weekly, and monthly timeframes. Such repeatability is a key indicator of quality for less stable compounds.4 To facilitate comparisons of our measurements with existing data, all published psat data for linalool between 270 K and 425 K23, 38–42 were taken from the NIST Thermodata Engine (TDE)34 and fit to the Wagner equation43 with equal weighting of all experimental data points. This Wagner correlation (eq 6 from Table 4) used a critical temperature of Tc = 668 K and a critical pressure of pc = 2907 kPa (where τ = 1 – T/Tc, and p0 = 1 kPa), both taken from the NIST TDE.34 Figure 4 is a deviation plot of experimental linalool psat data (multiple colored data points) from this Wagner correlation (black line). In Table 2, values of vapor pressure from the correlation (pcorr) are given at each measurement temperature.
Figure 4.

The percent deviation plot for experimental linalool psat values from literature (varied colors and shapes) and the current measurements (pink stars). The black zero line represents the Wagner correlation of all published psat data for linalool between 270 K to 425 K (excluding the present work, Table 4 eq 6). Panel A is the full range, and panel B is a zoom in to ±10 %.
In Figure 4, our measurements of linalool psat with DVME are the pink stars. The Wagner correlation, represented by the black zero line in Figure 4, is heavily influenced by the two data sets from Zaitsau et al. (gold triangles and teal triangles)23 because of its numerous data points. The data from Zaitsau et al. appear to be of exceptionally high quality, as indicated by the relatively small scatter and the good agreement obtained with two different measurement methods: a gas saturation method (called “transpiration” by Zaitsau et al.) and a static method.23 The transpiration data from Zaitsau et al. are, on average, a little higher and a little more scattered than their static data, but it is not clear which data set is more accurate. In any case, the DVME-based measurements have a similar level of scatter as the transpiration data from Zaitsau et al., Figure 4. Notably, 22 of 32 data points in the DVME data set (Table 2) deviate from the Wagner correlation by less than the ucombined of the DVME-based measurements. Such agreement is evidence that the uncertainty estimates for the DVME method are reasonable. The large deviations and scatter for the other published data sets in Figure 4 are typical of psat measurements in this temperature and pressure range.
Of course, the literature data and, therefore, the Wagner correlation have uncertainty as well. See the SI, Section 5 for more details about the Wagner correlation, including Tables S4 (literature data) and S5 (correlated values). In addition to the Wagner correlation of the literature data alone (eq 6), we created a Wagner correlation of the literature data plus the DVME data (Table 4, eq 7). Finally, we created a Wagner correlation of the DVME data alone (Table 4, eq 8). These are provided to facilitate comparisons with any future measurements.
As shown in Table 3, the DVME-based psat measurements had the three dominant sources of uncertainty: the temperature of the saturator (T, as measured by the PRT), the mass of helium carrier gas (m1, as measured by the MFM), and the mass of linalool (m2, as measured by GC-FID).1 Because of the drying procedure, water contamination was assumed to not contribute to the uncertainty in linalool purity. Even though the solubility of helium in linalool was estimated from helium solubility in 1-decanol,31 this does not introduce significant measurement uncertainty because helium’s solubility is low compared to the concentration of other impurities. The relative combined standard uncertainty (ucombined) at 314 K is estimated at 0.0584·psat and ranged from 0.0358·psat to 0.0406·psat at the other temperatures, Table 3.
The measurement uncertainties for linalool psat are higher than our previous measurements for n-eicosane, which ranged from 0.0203·psat to 0.0282·psat at temperatures from 344 K to 374 K.1 One reason for the difference is that, with n-eicosane, more carrier gas flow (240 scc to 800 scc) was needed to obtain optimal quantities of condensed analyte in the capillary vapor trap to analyze by GC-FID, which lengthened the measurement period but minimized the uncertainty in the mass of helium, m1. Also, for the GC oven used, the temperature setpoint had relatively high uncertainty near ambient temperature; for this reason, measurements below 314 K were not made. This limitation could presumably be overcome with different hardware (such as adding a chiller to balance the heater in the oven).
The enthalpy of vaporization (ΔHvap) was determined from a Clausius-Clapeyron plot of the DVME data, ΔHvap(T range 314 K – 354 K) = 61.8 ± 0.4 kJ/mol (see Figure S5). This value agrees with (but slightly outside the standard uncertainty intervals for) the value determined from the Wagner correlation (Table 4 eq 6) for the same temperature range, ΔHvap(corr) = 60.8 ± 0.4 kJ/mol (see Figure S4). At the median experimental temperature of 334 K, NIST TDE web application predicts a ΔHvap(334 K) = 59.3 ± 2.2 kJ/mol (https://wtt-pro.nist.gov; subscription to TDE required). The experimentally determined ΔHvap value is within the uncertainty of the predicted value.
3.5. Implications for cannabis breathalyzer development.
It is instructive to compare the psat curve for linalool to the psat curves for ethanol and THC (Figure 5). Breathalyzer tests for ethanol intoxication are relatively easy because of ethanol’s high psat at body temperature. Cannabis breathalyzer development that is based (solely) on the detection of THC is much more challenging, in large part because the psat of THC is about one millionth that of ethanol.2 Monoterpenes like linalool have psat curves closer to ethanol than to THC (the psat of linalool is about one hundredth that of ethanol). We propose that this fundamental behavior makes cannabis-associated terpenes potentially valuable as additional breath markers of recent cannabis use. Breath studies for terpenes are lacking, but it is noteworthy that linalool has been detected in the vapor phase above human whole blood following dermal exposure.44 A better understanding of terpene metabolism is also needed. However, for linalool, gastric exposure in rats45 or dermal exposure in humans44 resulted in peak blood concentrations of linalool after 15−40 minutes. Therefore, we believe that further research into the concentration of linalool in both breath and blood after recent cannabis use may be useful in the development or validation of meaningful breath-based measurements for law enforcement.
Figure 5.

Vapor pressure (psat) data for ethanol34, linalool (current work), and THC2 over a similar temperature range.
4. Conclusions
The DVME method was optimized to measure the psat of linalool up to 1250 Pa, which is about 200 times higher psat than previous measurements with DVME on n-eicosane. This greatly extends the utility of the method for measuring a variety of compounds important to criminal justice and public health. Two key changes were made to the apparatus to allow this increase in measurement range: the capillary vapor trap had a larger inner diameter and was cooled to 263 K with a thermoelectric cooling plate to increase trapping efficiency. Measurements at even higher psat are presumably possible; the suggested strategy would be to use lower flow rates and less total helium, although this would increase the uncertainty of the mass of helium carrier gas. Additionally, multiple precautions were taken to ensure that linalool purity did not affect the measurement results: a fresh saturator was made daily, a drying step was included, and TBHQ was added as an antioxidant. With these precautions, there was no evidence that impurities or sample decomposition significantly influenced the psat measurements.
Supplementary Material
Acknowledgements
Dr. Megan Harries provided useful advice in the planning of this work. Dr. Adam Friss installed the thermoelectric cold plate on the DVME apparatus.
Funding Sources
C.N. Beuning was supported by a National Research Council (US National Academy of Sciences) Postdoctoral Associateship Award. This research was partly supported by funding from the National Institute of Justice, Office of Justice Programs, US Department of Justice (DJO-NIJ-19-RO-0008). The funders had no role in study design, data collection, analysis, publication decision, or manuscript preparation. The opinions, findings, conclusions, or recommendations expressed in this publication are those of the authors and do not necessarily reflect those of the Department of Justice.
Abbreviations
- DVME
dynamic vapor microextraction
- GC-FID
gas chromatography with flame ionization detection
- ID/OD
inner and outer diameter
- MFC
mass flow controller
- MFM
mass flow meter
- PRT
platinum resistance thermometer
- scc(m)
standard cubic centimeters (per minute)
- C14
tetradecane
- TBHQ
tert-butylhydroquinone
- THC
Δ9-tetrahydrocannbinol
Footnotes
Specific commercial equipment, instruments, or materials are identified in this paper to specify the experimental procedure adequately. Such identification is not intended to imply recommendation or endorsement by NIST, nor is it intended to suggest that the materials or equipment identified are necessarily the best available for the purpose.
The authors declare no competing financial interest.
Supporting Information
The SI is available free of charge on the ACS website.
The SI includes details about the pre-treatment of the saturator vial to remove water contamination, the GC-FID sensitivity correlation, an expanded overpressure curve for the 530 μm trap, a Clausius-Clapeyron plot of linalool data, and all of the literature data points that were used to determine the Wagner fit for percent deviation calculations.
References
- (1).Harries ME; Beuning CN; Johnston BL; Lovestead TM; Widegren JA Rapid Vapor-Collection Method for Vapor Pressure Measurements of Low-Volatility Compounds. Anal. Chem 2020, 92 (24), 16253–16259. DOI: 10.1021/acs.analchem.0c04131. [DOI] [PubMed] [Google Scholar]
- (2).Lovestead TM; Bruno TJ Determination of Cannabinoid Vapor Pressures to Aid in Vapor Phase Detection of Intoxication. Forensic Chemistry 2017, 5, 79–85. DOI: 10.1016/j.forc.2017.06.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (3).Widegren JA; Bruno TJ Vapor Pressure Measurements on Low-Volatility Terpenoid Compounds by the Concatenated Gas Saturation Method. Environ. Sci. Technol 2010, 44 (1), 388–393, Article. DOI: 10.1021/es9026216. [DOI] [PubMed] [Google Scholar]
- (4).Widegren JA; Beall CE; Tolbert AE; Lovestead TM; Bruno TJ The Use of Antioxidants to Improve Vapor Pressure Measurements on Compounds with Oxidative Instability: Methyl Oleate with tert-Butylhydroquinone. Journal of Chemical & Engineering Data 2017, 62 (1), 539–546. DOI: 10.1021/acs.jced.6b00821. [DOI] [Google Scholar]
- (5).Widegren JA; Harvey AH; McLinden MO; Bruno TJ Vapor Pressure Measurements by the Gas Saturation Method: The Influence of the Carrier Gas. Journal of Chemical and Engineering Data 2015, 60 (4), 1173–1180. DOI: 10.1021/je500865j. [DOI] [Google Scholar]
- (6).Macknick AB; Prausnitz JM Vapor-Pressures of High-Molecular-Weight Hydrocarbons. Journal of Chemical & Engineering Data 1979, 24 (3), 175–178. DOI: 10.1021/je60082a012. [DOI] [Google Scholar]
- (7).Mokbel I; Razzouk A; Hajjaji A; Msakni N; Jose J A Gas Saturation Apparatus for Very Low Vapor or Sublimation Pressure Measurements (10–3 Pa): Vapor-liquid Equilibria of n-alkanes (n-C10, n-C24, n-C28). Journal of Chemical and Engineering Data 2007, 52 (5), 1720–1725. DOI: 10.1021/je7001122. [DOI] [Google Scholar]
- (8).Verevkin SP; Emel’yaneriko VN Transpiration Method: Vapor Pressures and Enthalpies of Vaporization of Some Low-boiling Esters. Fluid Phase Equilib. 2008, 266 (1–2), 64–75, Article. [Google Scholar]
- (9).Carruth GF; Kobayashi R Vapor Pressure of Normal Paraffins Ethane through n-Decane from Their Triple Points to about 10 mm Hg. Journal of Chemical & Engineering Data 1973, 18 (2), 115–126. DOI: 10.1021/je60057a009. [DOI] [Google Scholar]
- (10).Wu Y; Eichler CMA; Chen S; Little JC Simple Method To Measure the Vapor Pressure of Phthalates and Their Alternatives. Environ. Sci. Technol 2016, 50 (18), 10082–10088. DOI: 10.1021/acs.est.6b02643. [DOI] [PubMed] [Google Scholar]
- (11).Lemmon EW; Goodwin ARH Critical Properties and Vapor Pressure Equation for Alkanes CnH2n+2: Normal Alkanes with n ≤ 36 and Isomers for n=4 through n=9. Journal of Physical and Chemical Reference Data 2000, 29 (1), 1–39. DOI: 10.1063/1.556054. [DOI] [Google Scholar]
- (12).Růžička K; Majer V Simultaneous Treatment of Vapor Pressures and Related Thermal Data Between the Triple and Normal Boiling Temperatures for n-Alkanes C5–C20. Journal of Physical and Chemical Reference Data 1994, 23 (1), 1–39, Article. DOI: 10.1063/1.555942. [DOI] [Google Scholar]
- (13).Myers HS; Fenske MR Measurement and Correlation of Vapor Pressure Data for High Boiling Hydrocarbons. Industrial and Engineering Chemistry 1955, 47 (8), 1652–1658. DOI: 10.1021/ie50548a052. [DOI] [Google Scholar]
- (14).Himes SK; Scheidweiler KB; Beck O; Gorelick DA; Desrosiers NA; Huestis MA Cannabinoids in Exhaled Breath following Controlled Administration of Smoked Cannabis. Clinical Chemistry 2013, 59 (12), 1780–1789. DOI: 10.1373/clinchem.2013.207407. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (15).Kintz P; Mura P; Jamey C; Raul J-S Detection of Δ9-tetrahydrocannabinol in exhaled breath after cannabis smoking and comparison with oral fluid. Forensic Toxicology 2017, 35 (1), 173–178. DOI: 10.1007/s11419-016-0333-x. [DOI] [Google Scholar]
- (16).Lynch KL; Luo YR; Hooshfar S; Yun C Correlation of Breath and Blood Δ9-Tetrahydrocannabinol Concentrations and Release Kinetics Following Controlled Administration of Smoked Cannabis. Clinical Chemistry 2019, 65 (9), 1171–1179. DOI: 10.1373/clinchem.2019.304501. [DOI] [PubMed] [Google Scholar]
- (17).Luo YR; Yun C; Lynch KL Quantitation of Cannabinoids in Breath Samples Using a Novel Derivatization LC–MS/MS Assay with Ultra-High Sensitivity. Journal of Analytical Toxicology 2019, 43 (5), 331–339. DOI: 10.1093/jat/bkz023. [DOI] [PubMed] [Google Scholar]
- (18).Hubbard JA; Smith BE; Sobolesky PM; Kim S; Hoffman MA; Stone J; Huestis MA; Grelotti DJ; Grant I; Marcotte TD; et al. Validation of a liquid chromatography tandem mass spectrometry (LC-MS/MS) method to detect cannabinoids in whole blood and breath. Clin Chem Lab Med 2020, 58 (5), 673–681. DOI: 10.1515/cclm-2019-0600. [DOI] [PubMed] [Google Scholar]
- (19).Jeerage KM; Beuning CN; Friss AJ; Bidwell LC; Lovestead TM THC in Breath Aerosols Collected with an Impaction Filter Device Before and After Legal-market Product Inhalation—A Pilot Study. Journal of Breath Research 2023, 17 (3), 037103. DOI: 10.1088/1752-7163/acd410. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (20).Shapira A; Berman P; Futoran K; Guberman O; Meiri D Tandem Mass Spectrometric Quantification of 93 Terpenoids in Cannabis Using Static Headspace Injections. Anal. Chem 2019, 91 (17), 11425–11432. DOI: 10.1021/acs.analchem.9b02844. [DOI] [PubMed] [Google Scholar]
- (21).Liktor-Busa E; Keresztes A; LaVigne J; Streicher JM; Largent-Milnes TM Analgesic Potential of Terpenes Derived from Cannabis Sativa. Pharmacological Reviews 2021, 73 (4), 98–126. DOI: 10.1124/pharmrev.120.000046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (22).Eržen M; Košir IJ; Ocvirk M; Kreft S; Čerenak A Metabolomic Analysis of Cannabinoid and Essential Oil Profiles in Different Hemp (Cannabis sativa L.) Phenotypes. Plants (Basel) 2021, 10 (5), 966. DOI: 10.3390/plants10050966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (23).Zaitsau DH; Verevkin SP; Sazonova AY Vapor Pressures and Vaporization Enthalpies of 5-nonanone, Linalool and 6-methyl-5-hepten-2-one. Data Evaluation. Fluid Phase Equilib. 2015, 386, 140–148. DOI: 10.1016/j.fluid.2014.11.026. [DOI] [Google Scholar]
- (24).Matura M; Sköld M; Börje A; Andersen KE; Bruze M; Frosch P; Goossens A; Johansen JD; Svedman C; White IR; et al. Selected Oxidized Fragrance Terpenes are Common Contact Allergens. Contact Dermatitis 2005, 52 (6), 320–328. DOI: 10.1111/j.0105-1873.2005.00605.x From Nlm. [DOI] [PubMed] [Google Scholar]
- (25).Salthammer T; Goss K-U Predicting the Gas/Particle Distribution of SVOCs in the Indoor Environment Using Poly Parameter Linear Free Energy Relationships. Environ. Sci. Technol 2019, 53 (5), 2491–2499. DOI: 10.1021/acs.est.8b06585. [DOI] [PubMed] [Google Scholar]
- (26).Weschler CJ; Nazaroff WW Semivolatile Organic Compounds in Indoor Environments. Atmos. Environ 2008, 42 (40), 9018–9040, Review. DOI: 10.1016/j.atmosenv.2008.09.052. [DOI] [Google Scholar]
- (27).Goss K-U; Schwarzenbach RP Empirical Prediction of Heats of Vaporization and Heats of Adsorption of Organic Compounds. Environ. Sci. Technol 1999, 33 (19), 3390–3393. DOI: 10.1021/es980812j. [DOI] [Google Scholar]
- (28).Martins M; Carvalho P; Palma A; Domanska U; Pinho S Selecting Critical Properties of Terpenes and Terpenoids Through Group-contribution Methods and Equations of State. Industrial & Engineering Chemistry Research 2017, 56. DOI: 10.1021/acs.iecr.7b02247. [DOI] [Google Scholar]
- (29).Caprioglio D; Mattoteia D; Pollastro F; Negri R; Lopatriello A; Chianese G; Minassi A; Collado JA; Munoz E; Taglialatela-Scafati O; et al. The Oxidation of Phytocannabinoids to Cannabinoquinoids. Journal of Natural Products 2020, 83 (5), 1711–1715. DOI: 10.1021/acs.jnatprod.9b01284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (30).Deatherage FE; Mattill HA Antioxidants and the Autoxidation of Fats Autoxidation of Oleic Acid, Methyl Oleate, Alcohol, and cis-9-Octadecene. Industrial & Engineering Chemistry 1939, 31 (11), 1425–1431. DOI: 10.1021/ie50359a024. [DOI] [Google Scholar]
- (31).Wilcock RJ; Battino R; Danforth WF; Wilhelm E Solubilities of Gases in Liquids II. The Solubilities of He, Ne, Ar, Kr, O2, N2, CO, CO2, CH4, CF4, and SF6 in n-octane 1-octanol, n-decane, and 1-decanol. The Journal of Chemical Thermodynamics 1978, 10 (9), 817–822. DOI: 10.1016/0021-9614(78)90154-4. [DOI] [Google Scholar]
- (32).Yang S; Schultz AJ; Kofke DA; Harvey AH Interpreting Gas-Saturation Vapor Pressure Measurements Using Virial Coefficients Derived from Molecular Models. Journal of Chemical and Engineering Data 2014, 59, 3183–3192. DOI: 10.1021/je500245f. [DOI] [Google Scholar]
- (33).NIST Chemistry Webbook, Standard Reference Database 69,. National Institute of Standards and Technology, Gaithersburg, MD. https://webbook.nist.gov/chemistry/ (accessed 2023). [Google Scholar]
- (34).NIST ThermoData Engine, Standard Reference Database 103a/103b, Version 10.4,. National Institute of Standards and Technology, Gaithersburg, MD. https://www.nist.gov/mml/acmd/trc/thermodata-engine/srd-nist-tde-103b (accessed 02/01/2021). [Google Scholar]
- (35).McNair HM; Miller JM; Snow NH Basic Gas Chromatography; Wiley, 2019. [Google Scholar]
- (36).Rose A; Schrodt VN Vapor-Liquid Equilibria for the Methyl Oleate and Methyl Stearate Binary System. Journal of Chemical & Engineering Data 1964, 9 (1), 12–16. DOI: 10.1021/je60020a006. [DOI] [Google Scholar]
- (37).Scott TA; Macmillan D; Melvin EH Vapor Pressures and Distillation of Methyl Esters of Some Fatty Acids. Industrial & Engineering Chemistry 1952, 44 (1), 172–175. DOI: 10.1021/ie50505a049. [DOI] [Google Scholar]
- (38).Clará RA; Marigliano ACG; Sólimo HN Density, Viscosity, and Refractive Index in the Range (283.15 to 353.15) K and Vapor Pressure of α-Pinene, d-Limonene, (±)-Linalool, and Citral Over the Pressure Range 1.0 kPa Atmospheric Pressure. Journal of Chemical & Engineering Data 2009, 54 (3), 1087–1090. DOI: 10.1021/je8007414. [DOI] [Google Scholar]
- (39).Deng DS; Li HR; Han SJ Isobaric (vapour plus liquid) Equilibria of (linalool+1propanol) and (linalool+1butanol). The Journal of Chemical Thermodynamics 2002, 34 (9), 1431–1437. DOI: 10.1016/s0021-9614(02)00189-1. [DOI] [Google Scholar]
- (40).Espinosa Díaz MA; Guetachew T; Landy P; Jose J; Voilley A Experimental and Estimated Saturated Vapour Pressures of Aroma Compounds. Fluid Phase Equilib. 1999, 157 (2), 257–270. DOI: 10.1016/S0378-3812(99)00016-3. [DOI] [Google Scholar]
- (41).García-Abarrio SM; Haya L; Pardo JI; Urieta JS; Mainar AM Isobaric VLE of the Mixture {(±)-linalool+ethanol}: A Case Study for the Distillation of Absolute and Volatile Oils. The Journal of Chemical Thermodynamics 2013, 64, 182–186. DOI: 10.1016/j.jct.2013.05.022. [DOI] [Google Scholar]
- (42).Zaretskii MI; Chartov EM; Rusak VV; Pushkina LA Liquid-vapor Phase Equilibria in Binary Systems of Methylheptenone, Geranylacetone, and Linalool with Methyl Acetoacetate. Russian Journal of Applied Chemistry 1999, 72 (9), 1512–1515. [Google Scholar]
- (43).Wagner W New Vapour Pressure Measurements for Argon and Nitrogen and a New Method for Establishing Rational Vapour Pressure Equations. Cryogenics 1973, 13 (8), 470–482. DOI: 10.1016/0011-2275(73)90003-9. [DOI] [Google Scholar]
- (44).Friedl SM; Oedendorfer K; Kitzer S; Reznicek G; Sladek G; Heuberger E Comparison of Liquid-Liquid Partition, HS-SPME and Static HS GC/MS Analysis for the Quantification of (−)-Linalool in Human Whole Blood Samples. Nat Prod Commun 2010, 5 (9), 1447–1452. DOI: 10.1177/1934578X10005009 From Nlm. [DOI] [PubMed] [Google Scholar]
- (45).Shi F; Zhao Y; Firempong CK; Xu X Preparation, Characterization and Pharmacokinetic Studies of Linalool-loaded Nanostructured Lipid Carriers. Pharmaceutical Biology 2016, 54 (10), 2320–2328. DOI: 10.3109/13880209.2016.1155630. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
