. 2023 Nov 24;5(1):24–33. doi: 10.1016/j.hroo.2023.11.015

Table 2.

Procedural characteristics (n = 20)

Catheter location
 GCV/AIV and LVOT 17 (85)
 GCV/AIV and RVOT 2 (10)
 Septal perforator vein and LVOT 1 (5)
Polarity of RF application
 Vein to endocardium 11 (55)
 Endocardium to vein 7 (35)
 Alternating 2 (10)
Active catheter
 3.5-mm irrigated 20 (100)
Return catheter
 3.5-mm irrigated 12 (60)
 8-mm nonirrigated 4 (20)
 4-mm nonirrigated 3 (15)
 2F EPstar catheter 1 (5)
Mapping system
 EnSite 13 (65)
 CARTO 7 (35)
Circuit impedance (Ω) 217 ± 77
Impedance drop (Ω) 21 ± 9 (range 8–35)
Power (W) 30 ± 8
No. of bipolar RF applications 5 ± 3
Bipolar RF total time (s) 238 ± 217
Use of half-normal saline 9 (45)

Values are given as n (%) or mean ± SD.

AIV = anterior intraventricular vein; GCV = great cardiac vein; LVOT = left ventricular outflow tract; RF = radiofrequency; RVOT = right ventricular outflow tract.