Skip to main content
. Author manuscript; available in PMC: 2024 Feb 5.
Published in final edited form as: Nat Protoc. 2020 Jun 26;15(8):2443–2469. doi: 10.1038/s41596-020-0337-1

Table 1 ∣.

Troubleshooting table

Step Problem Possible reason Solution
‘Equipment setup’, ‘Preparation of the observation chamber’ PDMS support is not properly cured Wrong curing time Find the optimal PDMS curing time by repeating the procedure several times. As an alternative to a PDMS-made support, small ribbons of biocompatible double-sided adhesives can be used (such as the GraceBio SecureSeal Adhesive). However, chamber contamination by the adhesives during membrane painting can occur
‘Equipment setup’, ‘Electrical noise check’ Excessive electrical noise Improper shielding and/or grounding Check that you have connected all the metallic parts of the setup to grounding point(s); avoid ‘ground loops’; ensure that all the equipment’s ground outlets are connected to a single grounding point
6 Thick lipid films fail to spontaneously convert to lipid bilayers Excessive amount of lipids on the grid mesh Clean the brush (Step 3); use the clean brush to remove excess lipid material
Lipid bilayers break spontaneously during the process of/shortly after painting Impurities are triggering nucleation of pores and membrane rupture Verify that lipid reservoirs (Fig. 2b) are not in contact with the PDMS support. Increase the lipid concentration in squalene. If nothing above works, change your lipids and/or solvent
15 Bad gigaseal. Gseal (Fig. 1b) is >100 pS and/or unstable Contaminated pipette tip Use freshly made pipettes and store pipettes in a closed jar to protect the tips from dust. Fire-polish the pipette tip right before use. Verify that a constant positive pressure is applied to the pipette during manipulations in solution; constant efflux for the tip diminishes the tip contamination risk
Inappropriate type of glass used Change the type of the glass capillary you use (see ‘Equipment setup‘, ‘Patch-clamp and delivery pipettes’); if conventional patch-clamp and borosilicate capillary fail, consider using a thick-walled capillary or quartz
18 The measuring current does not increase with the pipette movement toward the membrane reservoir The NT is broken or its lumen is inaccessible or clogged Remove and discard the patch-clamp pipette. Repeat one to three times from Step 7; an NT should form in 50–90% of the attempts
Decrease the deviation of the pipette axis from the vertical
Decrease the speed of the pipette movement.
Consider raising the pipette incrementally, in small steps
See Troubleshooting advice for Step 6
19A(iii) The dependence of Im on changes of the NT length is not hyperbolic; the data fitting (Box 1) is not reliable Nano-positioner is not calibrated Label the tip of a patch-clamp pipette with a fluorescent marker (e.g., soak the pipette in a solution of 100-nm fluorescent microspheres). Using this pipette as a marker, make a calibration curve relating the position of the tip (in the vertical direction) and the reading of the nanoposition controller. Verify that the curve is reproducible in repetitive raising/lowering of the pipette
The NT conductance is unstable at constant length (drift) See Troubleshooting advice for Step 19B(i)
19B(i) Conductance changes at a fixed position of the nanopositioner Patch-clamp pipette and/or grid mesh move slowly Using phase-contrast optics (40× or 60× objective) determine whether the pipette and/or grid move. Identify and eliminate the reason of the drift (loose attachment of the grid/pipette, rotation of the pipette holder, mechanical drift of the micromanipulator and/or microscope stage)
19C(i) IU characteristic is either linear or not symmetric Wrong voltage range used Increase U. Chose the nanotube with high rNT value (low tension)
IU asymmetry is seen with charged membrane due to electrophoresis in the inner lipid monolayer.
Reduce U, apply fast voltage ramps
19C(iii) k is highly variable; the distribution of the k values is not a sharp Gaussian Poor membrane material Change lipid stocks
20 Low SNR Vibrations of the patch-clamp pipette Check that your setup is properly protected against vibrations
25 Displacements of the delivery pipette cause acute and irreproducible changes in the NT conductance Mechanical interference or large efflux from the delivery pipette Check that the delivery pipette is mechanically isolated from the patch-clamp pipette (look for, e.g., entangled cords, physical contact between the micromanipulators; verify that the delivery pipette’s controller is not causing vibrations). Reduce the approach speed. Trace the efflux, for example, by adding fluorescently labeled nanospheres to the delivery pipette; verify that the spheres escape the pipette slowly
27 Upward stepwise changes of Im Membrane poration Change the osmoticant. If the problem persists, it indicates that the poration is caused by the curvature stress; the effect depends on the lipid composition used. Change the lipid composition or avoid high NT constriction
Acute drop of the NT conductance to the background level NT scission Avoid extreme NT constrictions causing scission
35 No NT constriction Delivery pipette is clogged Change the delivery pipette, using a low-magnification microscope verify that the pipette tip is clean
Gradual conductance decrease; no steps are observed Low SNR Decrease the NT length. Verify the SNR curve (Step 20). Use NTs with higher rNT values (lower lateral tension). To further increase SNR high-ionic-strength (1 M) solution can be used in the patch-clamp pipette
Upward stepwise changes of Im Membrane poration by the protein Use dialysis to remove detergent from the protein solution. Decrease the protein concentration. To further assess the effect, the protein can be added to the patch-clamp pipette at Step 7. Continue the protocol until Step 15; do not rupture the patch. Apply the main acquisition protocol (Step 19A(i)), monitor the patch conductance, and look for upward channel-like events. The channel-like activity, if detected, can be minimized by optimizing lipid composition, for example, by adding cholesterol or lipids promoting inverted hexagonal phase appearance
48 The usNT radius variations from membrane to membrane are small Membranes on the different holes of the grid mesh share the same lipid reservoir Prepare another grid and repeat the protocol