Skip to main content
. 2024 Jan 17;20:101737. doi: 10.1016/j.bonr.2024.101737

Table 2.

Within- and between-day precision data for pentosidine in human serum (n = 21).


Within-day
Between-day
Mean concentration (nM) SD CV (%) Mean concentration (nM) SD CV (%)
QC Low 64 1.9 3.0 64 4.2 6.5
QC Medium 162 3.0 1.8 165 7.8 4.7
QC High 509 27.8 5.5 534 23.2 4.4

SD – Standard Deviation.