. 2024 Jan 29;10(3):e25158. doi: 10.1016/j.heliyon.2024.e25158

Table 1.

Comparison data of six assays vs. all-assays median values of CEA results for 40 individual serum samples.

Measuring system Regression parameters

Slope (95 % CI) Intercept (95 %CI) Cusum test rs (95 %CI) Average bias, mg/L (95 % CI) Average bias,
% (95 % CI)a
CCC (95 %CI)
Abbott
Architect i2000
1.0201 (1.000–1.0422) 0.0913 (−0.1325 - 0.4250) P = 0.53 0.996 (0.993–0.998) 2.97 (−0.29 - 6.23) 1.67 (−1.78 - 5.12) 0.9990 (0.9982–0.9994)
Autolumo
A2000
0.9184 (0.8954–0.9363) 2.3815 (1.8409–3.2455) P = 0.15 0.998 (0.997–0.999) −19.30 (−39.74 - 1.15) 6.16 (0.46–11.86) 0.9448 (0.9293–0.9570)
Beckman
DxI800
1.1462 (1.0845–1.2054) −0.4720 (−1.2793 - 0.3001) P = 0.97 0.998 (0.997–0.999) 18.97 (2.10–35.84) 13.18 (8.47–17.89) 0.9737 (0.9556–0.9844)
Diasorin
Liaison XL
1.0508 (1.0209–1.0984) −08629 (−1.3395 to −0.4254) P = 0.15 0.997 (0.995–0.999) 7.26 (0.80–13.73) 1.50 (−1.79 - 4.78) 0.9958 (0.9925–0.9976)
Maccura
IS1200
0.9804 (0.9504–1.0189) −0.3010 (−0.7041 - 0.0813) P = 0.15 0.997 (0.994–0.998) −1.72 (−12.64 - 9.21) −4.17 (−7.34 to −1.00) 0.9891 (0.9797–0.9942)
Roche
Cobas E601
0.8617 (0.8208–0.8883) 0.7981 (0.4616–1.2332) P = 0.53 0.999 (0.997–0.999) −19.72 (−36.83 to −2.61) −9.00 (−11.86 to −6.13) 0.9614 (0.9489–0.9709)
a

Assays not fulfilling optimal bias analytical quality specifications are in bold. CCC, concordance correlation coefficient.