. 2024 Feb 1;29:100544. doi: 10.1016/j.phro.2024.100544

Table 1.

Alanine and film results filtered per algorithm. The number of beams is written in parentheses. The mean deviation between the measured dose by the alanine and the calculated dose is reported per algorithm. The standard deviation is reported in parentheses. For the film results the mean passing rate per algorithm is reported and the standard deviation on the passing rate is reported in parenthesis.

Alanine results
Algorithm (# beams) Mean dose difference (%) Standard deviation (%)
AAA (6) 2.6 2.2
Acuros Dw,m (11) 2.3 1.2
CCC (4) 2.1 1.2
Monte Carlo (8) 3.2 0.9



Film results

Algorithm (# beams) Mean PR through target 5 %/1mm (st. dev) Mean PR top of lungs 3 %/2mm (st. dev)
AAA (6) 98.7 (1.8) 94.9 (3.2)
Acuros Dw,m (11) 97.6 (2.2) 91.9 (7.7)
CCC (4) 99.9 (0.2) 96.8 (2.6)
Monte Carlo (8) 99.9 (0.2) 97 (2.6)