Skip to main content
. 2024 Feb 8;14:2652. doi: 10.1038/s41598-024-52487-5

Table 3.

Average rate of detected interictal spikes over all channels per minute for the BIC unit and the clinical amplifiers during the 24-h recordings.

Subject BIC spikes Clinical system spikes BIC/clinical system ratio Number of channels (bipolar)
P1 33.6 35 0.96 26
P2 46.4 48.8 0.95 28
P3 59.2 62.1 0.95 28