Skip to main content
. 2024 Jan 25;8:e51770. doi: 10.2196/51770

Table 2.

Screening accuracy of DigiBel for hearing threshold >20 dB identified by automated pure tone audiometry.

Frequency (Hz) Sensitivity (%), 95% CI Specificity (%), 95% CI
250 95.00 (75.13 to 99.87) 85.00 (70.16 to 94.29)
500 91.30 (71.96 to 98.93) 89.19 (74.58 to 96.97)
1000 100.00 (86.28 to 100.00) 85.71 (69.74 to 95.19)
2000 100.00 (86.77 to 100.00) 79.41 (62.10 to 91.30)
4000 100.00 (86.28 to 100.00) 88.57 (73.26 to 96.80)
8000 100.00 (86.28 to 100.00) 88.57 (73.26 to 96.80)
Screening frequencies (250, 1000, 2000, and 4000) 100.00 (87.23 to 100.00) 72.73 (54.48 to 86.70)