Inter-assay accuracy and precision.

Item LLOQ
(1.25 µg/mL)
LQC
(3.00 µg/mL)
MQC
(12.0 µg/mL)
HQC
(22.5 µg/mL)
ULOQ
(30.0 µg/mL)
Inter-run Mean 1.09 2.88 11.9 22.2 28.3
Inter-run % CV 7.6 6.5 7.3 4.0 3.0
Inter-run % Bias −12.8 −4.0 −0.8 −1.3 −5.7
Inter-run % Total Error 20.4 10.5 8.1 5.3 8.7
n 6 6 6 6 6