Table 3.
Guide to assay troubleshooting.
| Problem | Potential cause | Protocol step | Corrective action |
|---|---|---|---|
| Plating pattern (striping pattern, checkerboard pattern) | Dispensing cassette tip obstruction during dispensing | Cells and cells plus virus dispensing | Obtain a new dispensing cassette |
| Dispenser failure | Cells and cells plus virus dispensing | QC dispensing instrument | |
| Edge effect | Wrong incubator humidity setting | Incubation | Ensure the incubator humidity is 95%. For added humidity, incubate the plates inside a pan with dampened paper towels |
| Low variable signal | Loss of trypsin activity during storage | Antiviral assay setup | Re-titration of the trypsin working stock / preparation of a new stock |
| Decrease in virus titer during storage | Antiviral assay setup | Re-titration of the virus stock | |
| Bad batch of collagen coated plates | Antiviral assay setup | Rerun in parallel with a plate from a different batch and obtain a new batch of plates |