Table 3.

Guide to assay troubleshooting.

Problem Potential cause Protocol step Corrective action
Plating pattern (striping pattern, checkerboard pattern) Dispensing cassette tip obstruction during dispensing Cells and cells plus virus dispensing Obtain a new dispensing cassette
Dispenser failure Cells and cells plus virus dispensing QC dispensing instrument
Edge effect Wrong incubator humidity setting Incubation Ensure the incubator humidity is 95%. For added humidity, incubate the plates inside a pan with dampened paper towels
Low variable signal Loss of trypsin activity during storage Antiviral assay setup Re-titration of the trypsin working stock / preparation of a new stock
Decrease in virus titer during storage Antiviral assay setup Re-titration of the virus stock
Bad batch of collagen coated plates Antiviral assay setup Rerun in parallel with a plate from a different batch and obtain a new batch of plates