Overview of the bias and coefficient of variation for different syringe volumes

Nuclide Syringe volume (mL) Nominal activity (MBq) Bias (%)
(Criteria: <  ± 5%)
Coefficient of variation (%)
(Criteria: <  ± 5%)
177Lu 30 100 0.77 1.70
20 100 0.16 0.74
3 40 − 2.00 3.03
89Zr 30 37 − 0.30 0.62
20 37 − 0.38 0.39
3 5 2.35 1.70
68 Ga 10 500 − 0.25 − 0.59
5 300 − 0.19 0.36
3 100 0.15 0.43
18F 10 1,000 − 0.51 0.19
5 500 0.03 0.53