Table II.

Previous works on electrochemical detection of Pb in human blood matrix.

References Sensor Technique * Sample preparation Gold standard ** Calibration range (μg/dL) LOD (μg/dL)
Li et al. 54 cyclodextrin modified gold electrode (MEA-β-CD) DPSV heating at 100 °C with HNO3 ICP-AES 0.35 −19.2 0.14 (standard solutions 1 M HClO4)
(3.5x dilution)
addition of DI H2O
10x dilution with
1 M HClO4
Zhao et al. 53 mesoporous polymer of melamine-formaldehyde/SPE SWASV not reported AAS (n = 6 blood samples) 0.1 − 5 0.1
Yantassee et al. 52 Hg-film on glassy carbon WE FIA / ASV ultrafiltration ICP-MS 0 − 2.0 0.046%RSD = 2.4 (10%blood)
(protein removal)
dilution to 20%blood with 1 M HCl
(5x dilution)
2x dilution to 10%blood
(0.5 M HCl)
Mai et al. 58 glassy carbon WE DPV / Cu2+ enhancement (800 ppb Cu2+) 3x dilution HNO3 AAS (n = 5 blood samples) 0.2 −10.0 (standard 0.01 M HNO3) 0.1%RSD = 4.6 (standard 0.01 M HNO3)
heating at 100 °C
additional dilution HNO3
Kruusma et al. 55 bismuth-film modified boron-doped diamond WE (Bi-BDD) sono-electroanalytical SWASV + ultrasound-assisted deposition pretreatment: 2/3 dilution,centrifugation AAS 0.62 − 8.9 0.87 (2%blood, 0.1 M HNO3 pH0.9)
50x dilution in 0.1 M HNO3 pH0.9, (2%blood)
Benzhi et al. 34 Nafion coated Bismuth-film WE DPSV pretreatment with HNO3 acidification AAS (n = 2 blood samples) 0.17 − 2.79 (standard 0.1 M AB) 0.013%RSD = 3.1 (standard 0.1 M AB)
+ advanced oxidation process (AOP with H2O2 and UV irradiation)
Bannon et al. 15 Hg-coated graphite WE (ESA Trace Metal Analyzer) ASV dilution with reagent to 3.33% blood GFAAS 1.0 − 70.0 (3.33% blood) 1.0%RSD = 11 (3.33% blood in reagent, pH = 1.3–1.4)
(30x dilution, 100 μL blood in 3000 μL solution)
Feeney et al. 43 LeadCare II ASV dilution with proprietory treatment reagent (containing HCl) GFAAS 3.3 − 65 3.3
this work Nafion/MC/SPCE WE SWASV 6x dilution with 0.5 M HCl ICP-MS 0.3 −10 0.39