Accuracy and extraction efficiency of analyte standards in solvent matrixa.
| Analytea | Measured (ng/mL) |
%CVb | REc (%) | EEd (%) |
|---|---|---|---|---|
| 25OHD3-S | ||||
| 1 ng/mL | 0.90 ± 0.12 | 13 | −12 | 66 ± 16 |
| 5 ng/mL | 4.84 ± 0.4 | 8.2 | −3.2 | 67 ± 12 |
| 20 ng/mL | 20.6 ± 1.5 | 7.5 | 3 | 77 ± 15 |
| VitD3-S | ||||
| 1 ng/mL | 0.95 ± 0.05 | 5.7 | −5 | 47 ± 15 |
| 5 ng/mL | 4.78 ± 0.47 | 9.8 | −4.4 | 48 ± 12 |
| 20 ng/mL | 21.35 ± 0.93 | 4.3 | 7 | 59 ± 19 |
Statistical data are mean ± SD, n=4.
Analytes were suspended in 60% MeOH for measurement.
Inter-assay coefficient of variation (CV), n=4.
Relative error (RE) is [(measured - theoretical) /theoretical]*100%.
Extraction efficiency (EE) is [(with extraction) /( without extraction)]*100%.