Skip to main content
. Author manuscript; available in PMC: 2025 Jan 1.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2023 Dec 6;1232:123954. doi: 10.1016/j.jchromb.2023.123954

Table 3:

Accuracy and extraction efficiency of analyte standards in solvent matrixa.

Analytea Measured
(ng/mL)
%CVb REc (%) EEd (%)
25OHD3-S
1 ng/mL 0.90 ± 0.12 13 −12 66 ± 16
5 ng/mL 4.84 ± 0.4 8.2 −3.2 67 ± 12
20 ng/mL 20.6 ± 1.5 7.5 3 77 ± 15
VitD3-S
1 ng/mL 0.95 ± 0.05 5.7 −5 47 ± 15
5 ng/mL 4.78 ± 0.47 9.8 −4.4 48 ± 12
20 ng/mL 21.35 ± 0.93 4.3 7 59 ± 19

Statistical data are mean ± SD, n=4.

a

Analytes were suspended in 60% MeOH for measurement.

b

Inter-assay coefficient of variation (CV), n=4.

c

Relative error (RE) is [(measured - theoretical) /theoretical]*100%.

d

Extraction efficiency (EE) is [(with extraction) /( without extraction)]*100%.