Hand Loading, Rates of Perceived Exertion, and Usability During Assisted Walking in Patients with Type 2 Diabetes Mellitus

Jose G Anguiano-Hernandez 1, Joseph W Harrington 1, David C Kingston 1,*
PMCID: PMC10872897  NIHMSID: NIHMS1940977  PMID: 37864920

Abstract

Background:

Patients suffering from a diabetic foot ulcer often receive a non-weight bearing prescription of the affected limb to promote healing. Total unilalteral offloading of the affected foot necessitates walking aids that require loading at the hands during ambulation. Excessive loading at the hands can increase of the risk of crutch palsy. In addition, certain walking aids can also be more strenuous and less comfortable to use than others, resulting in lower prescription compliance. This study aimed to investigate hand loading, rates of perceived exertion, and usability of typically prescribed walking aids in patients with Type 2 Diabetes.

Methods:

Twenty patients (12 F ∣ 8 M, 61.0 ± 10.36 yrs, 90.54 ± 13.34 kg, 1.71 ± 0.08 m) walked as much as 200 m without assistance and with crutches, walkers, and a wheeled knee walker instrumented with flexible force-sensing pads on walking aid handles. Patients rated exertion using a Modified Borg Scale and completed a System Usability Scale questionnaire after each walking condition with or without walking aids.

Findings:

Results show that using a wheeled knee walker required 94% less hand loading than crutches and walkers. Patients reported 45% lower exertion with the WKW compared to crutches and walkers, and scored the usability of the wheeled knee walker 106% higher than crutches and walkers.

Interpretation:

The wheeled knee walker could be the preferred walking aid for total unilateral offloading because of the reduced loading demand at the hands, lower exertion during use, and greater usability compared to crutches and walkers.

Keywords: Bodyweight Loading, Hand Loading, Rate of Perceived Exertion, Usability, Walking Aids, Type 2 Diabetes Mellitus

1. Introduction

Approximately 50% of the Type 2 Diabetes Mellitus (DM) population will suffer from a diabetic foot ulcer in their lifetime1-3, with estimates predicting that diabetic foot ulcers will develop in 9.1 million to 26.1 million patients worldwide4. Over half of diabetic foot ulcers become infected5, potentially leading to amputation or premature death6-8. One intervention to promote ulcer healing is total offloading of the affected foot via walking aids, such as crutches, walkers, or wheeled knee walkers (WKW)9. These devices allow total offloading of the affected foot by supporting body weight (BW) through the hands and propulsive foot during walking. The WKW has a design advantage compared to other walking aids, as the ipsilateral shin and knee can also support BW. Previous research comparing peak plantar force in the propulsive foot during normal and assisted gait with crutches, walkers, and WKWs in patients with Type 2 DM found that locomotion with the WKW reduced net plantar impulse in the propulsive foot by 43% - 77% compared to unassisted walking, crutches, and walkers but did not reduce impulse at the hands10. Although this study could not attribute reduced plantar impulse to reduced hand impulse, peak loading at the hands could be indicative of injury risk at the hand and wrist. Excessive BW loading onto the hands during ambulation may also negatively influence patient comfort. Excessive hand loading increases the risk of crutch palsy, which occurs due to compression of the axillary and radial nerves in the upper arm and hand11-13. Patients with Type 2 DM may be more susceptible to crutch palsy as this population has reduced upper body strength, endurance, and higher BW compared to healthy counterparts14.

Usability of walking aids could be an important factor in patient compliance with non-weight-bearing prescriptions. One verified tool to quantify feedback and usability of novel systems or devices is the System Usability Scale (SUS)15,16. Although system usability of crutches, walkers, and WKWs has not yet been investigated, previous research has found that healthy adults reported lower rates of perceived exertion (RPE) and walked faster when using a WKW to assist with ambulation compared to crutches17. Additionally, prior work with healthy older adult reports similar self-reported RPE using crutches (5.7) and walkers (6.1) using a Modified Borg 0-10 scale18. The same sample population reported a 54% lower RPE after using a WKW18. While these findings may have some transferability to patients with Type 2 DM, patients could have other comorbidities, lower cardiovascular endurance, and reduced overall strength that could affect the way they use crutches, walkers, and WKWs. Therefore, there is a need to investigate RPE, SUS, and peak hand loading during walking aid use in patients with Type 2 DM.

The primary purpose of this study was to compare peak BW supported by the hands during assisted walking with crutches, walkers, and WKWs in patients with Type 2 DM. This is a necessary first step to determine differences between patients with Type 2 DM and good disease management to prior work in healthy older adults before investigating clinical outcomes, such as ulcer healing time, in a cohort of patients suffering from diabetic foot ulcers in a clinical trial. We hypothesize that locomotion with the WKW will result in the lowest peak hand loading compared to crutches and walkers. The secondary purpose of this study was to compare patient-reported RPE and usability of crutches, walkers, and WKWs in patients with Type 2 DM. We hypothesize that patients will report the lowest RPE after using the WKW compared to crutches and walkers. We also hypothesize that patients will report the greatest usability after using the WKW compared to crutches and walkers.

2. Methods

2.1. Participants

20 participants (12 F ∣ 8 M) were recruited for this study with demographics reported in Table 1. This sample size was chosen to facilitate comparisons between our sample population and previously reported findings18. Participants were pre-screened for eligibility by clinical support staff from the University of Nebraska Medical Center’s Department of Endocrinology, Diabetes, and Metabolism. Eligible participants were referred to the investigators for screening prior to participation using a medical history questionnaire.

Table 1.

Participant characteristics.

Characteristic
mean (SD)
Male (n = 8) Female (n = 12) Total (n = 20)
Age (years) 64.25 (10.36) 58.33 (10.38) 61.0 (10.36)
Mass (kg) 96.24 (15.55) 86.74 (10.68) 90.54 (13.34)
Height (m) 1.75 (0.06) 1.69 (0.09) 1.71 (0.08)
Time Since Diagnosis (yr) 15.00 (7.98) 13.10 (5.49) 13.94 (6.57)
HbA1c (%) 7.44 (0.73) 6.65 (0.84) 7.00 (0.87)
HbA1c (mmol/mol) 57.88 (7.77) 49.20 (9.09) 53.06 (9.40)

Inclusion criteria for this study were: individuals older than 40 years of age with Type 2 DM, A1c < 10%, and a self-reported ability to walk for 5 minutes at a self-selected speed. We elected to only include participants with an A1C < 10% (< 86 mmol/mol) in our study to exclude those that may present with additional medical conditions that could hinder their ability to complete our study. Exclusion criteria were: having a lower extremity joint replacement or amputation, unexplained dizziness during the previous 6 months, use of a walking aid during the previous 12 months, patients currently in treatment for a Diabetic Foot Ulcer, and BW exceeding 136 kg (weight capacity of the walking aids). Potential participants who had experienced an unexpected bout of dizziness 6 months prior to recruitment were excluded to maximize participant safety and minimize risk of falling during the experiment, as ambulating with one foot totally offloaded while using a walking aid would require a certain degree of balance.

2.2. Experimental Protocol

This study required one visit to the University of Nebraska Omaha Biomechanics Research Building. After informed consent participants completed a familiarization period to fit walking aids to participants’ bodies as all self-reported as novice users. The familiarization period ended when participants stated that each walking aid was adjusted to an appropriate height for their bodies to walk with while totally offloading a foot of their choice. The study consisted of four walking conditions: 1) normal walking, 2) crutches, 3) walker, and 4) WKW. The order of conditions was randomized. Participants were instructed to walk a maximum of 200 m along a 10 m straight path in each walking condition while unloading one foot. Compliance with unloading of the foot was checked via force-sensing insoles (loadsol, Novel Electronics Inc., St. Paul, MN, USA), however, contralateral loading results were previously reported10. A total distance of 800 m was chosen as this distance is representative of typical community-based walking bouts19. To minimize potential fatigue, participants could elect to finish walking conditions before reaching 200 m and distance covered was recorded.

Flexible force-sensing pads sized 158.75 mm x 107.95 mm (loadpad, Novel Electronics Inc., St. Paul, MN, USA) were used to measure force loaded through the hands onto the crutches and walker. Force-sensing pads were calibrated to each participant’s BW using static unilateral toe standing. Participants were instructed to stand on a force-sensing pad while balancing on their metatarsal-phalangeal joints of one foot. The contralateral foot was offloaded during this time. A walker was given to each participant to aid with balance, but minimal contact was encouraged. Force-sensing pads were wrapped around the handles of each walking aid (Figure 1) to measure force at 100 Hz syncronized with the insoles. A custom-made WKW instrumented with a strain gauge under the handlebar (TAL206, HT Sensor Technology Co., Ltd, China, capacity: 20-300kg) and knee cushion (TAL208, HT Sensor Technology Co., Ltd, China, capacity: 60-800 kg) was used in this study (Figure 2). The strain gauges recorded force at 50 Hz from both sites with synchronization procedures detailed in section 2.3.

Figure 1.

Figure 1.

Flexible force-sensing pads (loadpad, Novel Electronics Inc., St. Paul, MN, USA) wrapped around walking aid handles.

Figure 2.

Figure 2.

Instrumented wheeled knee walker used in this study. Red arrows point to strain gauges under the handlebars (TAL206, HT Sensor Technology Co., Ltd, China) and knee cushion (TAL208, HT Sensor Technology Co., Ltd, China) used to record force data.

Patient-reported RPE were quantified using a Modified Borg Scale, commonly used for rating exertion in clinical rehabilitation and exercise20,21. This tool is a 0-10 scale, where 0 equates to ‘at rest’ and 10 equates to ‘very, very hard’. Participants provided an RPE after each walking condition. Walking aid usability was quantified by completing a SUS Questionnaire after each walking aid condition22. The SUS is a mixed-tone questionnaire in which odd-numbered questions had positive tones, and even-numbered questions had negative tones. A copy of one device’s SUS questionnaire is provided in Appendix A. Participants rated their agreement with each statement on a 1-5 scale, where 1 equates to ‘Strongly Disagree’ and 5 equates to ‘Strongly Agree’. SUS Scores were calculated using Equation 122.

Equation 1.

Equation used to calculate System Usability Scale (SUS) Score. Higher scores indicate greater usability.

SUS Score=(Q1+Q3+Q5+Q7+Q9−5+25−Q2−Q4−Q6−Q8−Q10)∗2.5

2.3. Data Processing

Data were processed using custom MATLAB software (R2021b, The Mathworks Inc., Natick, MA, USA). During crutches and walker conditions, the force-sensing insoles and pads synchronously recorded data via the manufacturer’s software. For the WKW condition, signals from the force-sensing insoles, pads, and strain gauges were synchronized by ‘slapping’ a pad placed over the handlebar strain gauge. This ‘slap’ produced a spike in both equipment’s signals that was used to synchronize devices (Figure 3). Strain gauge signals were then upsampled to 100 Hz in MATLAB using a custom function.

Figure 3.

Figure 3.

Visual of synchronization 'slap' during wheeled knee walker Conditions. Peak force magnitudes of the slap were used to synchronize the pad and strain gauge signals.

A 50 N threshold was used to identify gait events from insole force measurements23. This threshold was chosen to avoid false heel strike or toe-off gait event identification due to deflections of the force-sensing insoles during gait. Gait event frame numbers were used to define stance periods in force-sensing pad and strain gauge data during all conditions. Raw force profiles were normalized to participant BW in magnitude and 101 points in time (0-100% stance). Peak BW supported was calculated by summing normalized force profiles from the offloaded and propulsive side hand in crutches and walker conditions, and summing normalized force profiles from the hands and knee strain gauges in the WKW condition. A peak BW loading value was then identified for each step and averaged acrossed all loading cycles (Table 2).

Table 2.

Mean (standard deviation) of outcome variables of interest. Separate one-way repeated measures ANOVA were used to compare peak hand loading, rate of perceived exertion, and System Usability Scale Score across all conditions. Values that share letters are not significantly different. Noted ηpartial2 values suggest large effect sizes for all outcome variables.

Variable Normal Crutches Walker WKW P-value ηpartial2
Peak Hand Loading (BW) - 0.93 (0.33) 1.58 (0.51) 0.09 (0.03) <0.001 0.859
RPE 1.48 (0.95) 6.26 (2.32)a 5.96 (1.85)a 3.43 (2.04) <0.001 0.745
SUS Score - 45.11 (26.56)a 40.23 (19.94)a 83.26 (17.80) <0.001 0.664
# of Steps 189(33) 123(102) 149(128) 175(57) - -

2.4. Statistical Analyses

Statistical analyses were performed using R (RStudio 2022, R Core Team, Vienna, Austria). Separate one-way repeated measures ANOVA were used to compare BW loading, RPE, and SUS for all four conditions. Significant main effects of walking condition were further investigated using Bonferroni-adjusted pairwise comparisons. All statistical tests were performed using an a priori α = 0.05.

3. Results

There was a significant main effect of walking condition on all variables of interest (Table 2). Hand loading during locomotion with the WKW produced 90.32% and 94.30% lower forces compared to crutches and walkers, respectively (Table 2). Post hoc pairwise comparisons found significant difference between all conditions (P < 0.001).

A significant main effect of walking condition was found when comparing RPE across all conditions (Table 2). Participants reported 45.21% and 42.45% lower RPE during the WKW condition compared to the crutches and walker conditions, respectively (Table 2). Although, RPE during the WKW condition was 131.76% greater than in the normal walking condition. Post hoc pairwise comparisons found significant differences between all conditions (P < 0.001) except between crutches and walkers (P = 1).

A significant main effect of walking condition was found when comparing SUS Scores across conditions (Table 2). Participants scored the WKW 84.57% and 106.96% higher than crutches and walkers, respectively (Table 2). Post hoc pairwise comparisons found significant difference between all conditions (P < 0.001) except the comparison between the crutches and walker (P = 0.474).

4. Discussion

This study compared peak BW supported and patient-reported exertion and usability during assisted walking with crutches, walkers, and WKWs in patients with Type 2 DM. Results suggest that locomotion with the WKW requires 88% and 94% less body weight loading via the hands compared to crutches and walkers, respectively. Additionally, patients reported the lowest RPE and greatest usability with the WKW, while reporting similarly high RPE and usability with the crutches and walker.

One reason we observed greater loading at the hands during the walker condition compared to crutches could be due to differences in walking strategies between the two walking aids. During locomotion with bilateral crutches, subjects exhibited walking patterns similar to a pendulum, where the crutch bases are planted on the ground and the patient swings their body forward while supporting their BW with the hands. This strategy could require less loading at the hands to support BW compared to walkers. Alternately, using a walker to assist with ambulation could elicit a walking pattern more similar to a ‘hop’, where the patient must use their hands and propulsive foot to propel their body up and forward to take a step. Our results suggest that using a walker could increase the risk of crutch palsy in Type 2 DM patients, who could be more susceptible to developing this condition compared to healthy older adults due to reduced upper body strength and endurance14. Due to the design of the WKW, significantly less BW was supported by the hands during ambulation. Instead, the offloaded side knee and shin supported the majority of BW during walking. This finding suggests that the WKW could be the preferred walking aid to reduce the risk of crutch palsy in patients with Type 2 DM during a non-weight-bearing prescription.

Although peak hand loading magnitudes were different between the crutches and walker, participants reported similar RPE and usability during these conditions. The WKW condition yielded the lowest RPE and SUS Score. These results agree with previous findings that RPE during crutches use was significantly greater compared to using a WKW in a healthy older population17, and suggest that if a patient is able to choose between using bilateral crutches or a walker for ulcer offloading, crutches could be the better choice due to reduced loading at the hands, greater locomotion speed10, and reduced plantar loading compared to walkers10. However, if the patients struggles with balance while using crutches, a walker could be the better option due to the larger base of support.

Limitations of this study include the size of the pads used to measure hand forces and the diabetic sample population. The pads could be wrapped entirely around the handles of the crutches and walker, thus grip forces could influence force measurements during ambulation that cannot be separated to only vertical loading. Patients in this study had good disease management (A1c < 10%; 86 mmol/mol) and no active diabetic foot ulcers. Patients needing to offload an active ulcer may have worse disease management and other comorbidities that affect the way they use walking aids, or ambulate in general, that may limit the transferability of our findings. In addition, there is a paucity of data for walking aid movement strategies in patients with pereipheral neuropothy. The loss of sensory feedback in the propulsive limbs may result in greater risk for tissue injury and is a needed area of future investigation.

5. Conclusion

This study found that the WKW could be the preferred walking aid for offloading a diabetic foot ulcer due to the least amount of loading at the hands required for ambulation and the greatest ease of use compared to crutches and walkers.

Supplementary Material

1
  • Less hand loading occurs in wheeled knee walkers than crutches and walkers

  • Lower exertion was reported with wheeled knee walkers compared to other devices

  • Diabetic patients reported the greatest usability with the wheeled knee walker

  • A wheeled knee walker was the most comfortable walking aid for plantar unloading

7. Acknowledgements

This work was supported by the University of Nebraska Omaha University Committee on Research and Creative Activity (UCRCA) Award [grant number 26073]. Equipment for this study was supported by the Center of Research in Human Movement Variability of the University of Nebraska at Omaha and the National Institutes of Health [grant number P20GM109090].

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Conflict of interest

The authors declare that there are no conflicts of interests.

Declaration of interests

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

8. References

  • 1.Boulton AJM. Chapter 8 - Diabetic neuropathy and foot complications. In: Zochodne DW, Malik RA, eds. Handbook of Clinical Neurology. Vol 126. Diabetes and the Nervous System. Elsevier; 2014:97–107. doi: 10.1016/B978-0-444-53480-4.00008-4 [DOI] [PubMed] [Google Scholar]
  • 2.Boulton AJM. The Pathway to Foot Ulceration in Diabetes. Medical Clinics of North America. 2013;97(5):775–790. doi: 10.1016/j.mcna.2013.03.007 [DOI] [PubMed] [Google Scholar]
  • 3.Kumar S, Ashe HA, Parnell LN, et al. The prevalence of foot ulceration and its correlates in type 2 diabetic patients: a population-based study. Diabet Med. 1994;11(5):480–484. doi: 10.1111/j.1464-5491.1994.tb00310.x [DOI] [PubMed] [Google Scholar]
  • 4.Atlas D. International diabetes federation. IDF Diabetes Atlas, 7th edn Brussels, Belgium: International Diabetes Federation. 2015;33. [Google Scholar]
  • 5.Prompers L, Huijberts M, Apelqvist J, et al. High prevalence of ischaemia, infection and serious comorbidity in patients with diabetic foot disease in Europe. Baseline results from the Eurodiale study. Diabetologia. 2007;50(1):18–25. doi: 10.1007/s00125-006-0491-1 [DOI] [PubMed] [Google Scholar]
  • 6.Walsh JW, Hoffstad OJ, Sullivan MO, Margolis DJ. Association of diabetic foot ulcer and death in a population-based cohort from the United Kingdom. Diabetic Medicine. 2016;33(11):1493–1498. doi: 10.1111/dme.13054 [DOI] [PubMed] [Google Scholar]
  • 7.Lipsky BA, Berendt AR, Cornia PB, et al. 2012 Infectious Diseases Society of America Clinical Practice Guideline for the Diagnosis and Treatment of Diabetic Foot Infectionsa. Clinical Infectious Diseases. 2012;54(12):e132–e173. doi: 10.1093/cid/cis346 [DOI] [PubMed] [Google Scholar]
  • 8.Lavery LA, Armstrong DG, Wunderlich RP, Tredwell J, Boulton AJM. Diabetic Foot Syndrome: Evaluating the prevalence and incidence of foot pathology in Mexican Americans and non-Hispanic whites from a diabetes disease management cohort. Diabetes Care. 2003;26(5):1435–1438. doi: 10.2337/diacare.26.5.1435 [DOI] [PubMed] [Google Scholar]
  • 9.Bus SA, Armstrong DG, Gooday C, et al. Guidelines on offloading foot ulcers in persons with diabetes (IWGDF 2019 update). Diabetes/Metabolism Research and Reviews. 2020;36(S1):e3274. doi: 10.1002/dmrr.3274 [DOI] [PubMed] [Google Scholar]
  • 10.Anguiano-Hernandez JG, Harrington JW, Shivaswamy V, Kingston DC. Alterations to plantar loading and ankle range of motion of the contralateral foot during assisted walking in patients with Type 2 Diabetes Mellitus. Gait & Posture. 2022;98:56–61. doi: 10.1016/j.gaitpost.2022.08.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Feldman DR, Vujic I, McKay D, Callcott F, Uflacker R. Crutch-induced axillary artery injury. Cardiovasc Intervent Radiol. 1995;18(5):296–299. doi: 10.1007/BF00203679 [DOI] [PubMed] [Google Scholar]
  • 12.Poddar SB, Gitelis S, Heydemann PT, Piasecki P. Bilateral predominant radial nerve crutch palsy. A case report. Clin Orthop Relat Res. 1993;(297):245–246. [PubMed] [Google Scholar]
  • 13.Raikin S, Froimson MI. Bilateral Brachial Plexus Compressive Neuropathy (Crutch Palsy). Journal of Orthopaedic Trauma. 1997;11(2):136–138. [DOI] [PubMed] [Google Scholar]
  • 14.Orlando G, Balducci S, Bazzucchi I, Pugliese G, Sacchetti M. Muscle fatigability in type 2 diabetes. Diabetes/Metabolism Research and Reviews. 2017;33(1):e2821. doi: 10.1002/dmrr.2821 [DOI] [PubMed] [Google Scholar]
  • 15.Bangor A, Kortum P, Miller J. Determining what individual SUS scores mean: adding an adjective rating scale. J Usability Studies. 2009;4(3):114–123. [Google Scholar]
  • 16.Lewis J, Sauro J. Revisiting the Factor Structure of the System Usability Scale. Journal of Usability Studies. 2017;12:183–192. [Google Scholar]
  • 17.Kocher BK, Chalupa RL, Lopez DM, Kirk KL. Comparative Study of Assisted Ambulation and Perceived Exertion With the Wheeled Knee Walker and Axillary Crutches in Healthy Subjects. Foot Ankle Int. 2016;37(11):1232–1237. doi: 10.1177/1071100716659748 [DOI] [PubMed] [Google Scholar]
  • 18.Kingston DC, Ferwerda S, Fontaine C, et al. Implications of Walking Aid Selection for Nonweightbearing Ambulation on Stance Limb Plantar Force, Walking Speed, Perceived Exertion, and Device Preference in Healthy Adults 50 Years of Age and Older. Foot Ankle Orthop. 2021;6(1):2473011421998939. doi: 10.1177/2473011421998939 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Yang Y, Diez-Roux AV. Walking Distance by Trip Purpose and Population Subgroups. American Journal of Preventive Medicine. 2012;43(1):11–19. doi: 10.1016/j.amepre.2012.03.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Lamb KL, Eston RG, Corns D. Reliability of ratings of perceived exertion during progressive treadmill exercise. Br J Sports Med. 1999;33(5):336–339. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Skinner JS, Hutsler R, Bergsteinová V, Buskirk ER. The validity and reliability of a rating scale of perceived exertion. Med Sci Sports. 1973;5(2):94–96. [PubMed] [Google Scholar]