Review of Flow Cytometry as a Tool for Cell and Gene Therapy

Christopher A Lazarski 1,2,*, Patrick J Hanley 1,2
PMCID: PMC10872958  NIHMSID: NIHMS1945272  PMID: 37943204

Abstract

Quality control testing and analytics are critical for the development and manufacture of cell and gene therapies, and flow cytometry is a key quality control and analytical assay that is used extensively. However, the technical scope of characterization assays and safety assays must keep apace as the breadth of cell therapy products continues to expand beyond hematopoietic stem cell products into producing novel adoptive immune therapies and gene therapy products. Flow cytometry services are uniquely positioned to support the evolving needs of cell therapy facilities, as access to flow cytometers, new antibody clones, and improved fluorochrome reagents becomes more egalitarian. This report will outline the features, logistics, limitations, and the current state of flow cytometry within the context of cellular therapy.

Keywords: Flow Cytometry, Cellular Therapy, Quality Control

Introduction

Cell and gene therapies represent a broad pipeline of bioengineered treatments for disease, encompassing gene editing and de novo expression, hematopoietic stem cell transplantation and immune reconstitution, targeted immunotherapy, and tissue regeneration. The translation of cell and gene therapies from preclinical research into formulations appropriate for clinical trials, standard of care, of licensed products is a complex process which bridges foundational science with bio-manufacturing quality cellular products safely and at scale. This manufacturing draws upon various assays to assess gene expression, cell expansion, viability, phenotype, function, and an array of other metrics. This process becomes exceedingly more complicated when the cell and gene therapies are manufactured specifically for each patient as they require strict protocols, training, validations, and adherence to good manufacturing practices (GMP). Identifying purity and safety of these products is critical after receipt of patient samples, during the manufacturing process, and during follow-up monitoring of the health and immune reconstitution of the patient after infusion. Flow cytometry as an assay can assess the phenotype, reactivity, and activation status of in process samples from products during manufacture, the final drug product itself, and follow up samples derived from patients enrolled in a clinical study or treated with a licensed drug. The versatility and speed of flow cytometry assays have established their utility as integral parts of many cell therapy manufacturing processes. Understanding the capabilities and limitations of flow cytometry assays is essential for effective integration into manufacturing and process development.

Flow cytometry as a service began to see adoption to evaluate the immune status of patients in the 1980s and 1990s as multi parameter systems became commercially available[1]. Flow cytometry was used to identify frequencies of CD4+ cells in HIV patients during treatment[2, 3], to identify the frequency of hematologic malignancies within whole blood[4], and for the initial identification of CD34+ stem cell populations[5, 6]. The versatility of flow cytometry assays thus cemented its use as a foundational component for clinical and pre-clinical research. Modern flow cytometry instruments continue to operate using principles developed on instruments constructed nearly sixty years ago[7]. Cells in suspension are funneled into a single file using hydrodynamic focusing of the sample as it is surrounded by sheath fluid. Fluid dynamics prevents the mixing of the sample with the sheath, and allows for narrow focusing of the cells past lasers of defined wavelengths[8]. These lasers activate fluorochromes associated with the cell, releasing light which is subsequently measured by photonic detectors after passing through filters of defined wavelengths. Data are reported to the user as units of relative signal intensity read out after amplification of the signal, which is not necessarily a specific numerical value of photons. Flow cytometry instruments are thus primarily defined by the lasers used to illuminate cells and the bandwidth of detectors used to measure the intensity of fluorescent markers bound to cells. Although flow cytometry assays may use similar antibodies as immunohistochemistry assays to target surface markers, flow cytometry is not necessarily liquid immunohistochemistry. While immuno-microscopy excels at highlighting interior cellular architecture and exterior anatomy, flow cytometry excels in the analysis and sorting of large numbers of living cells in suspension[8, 9], and both assays can serve independent roles in diagnostic hematopathology.

Common uses of flow cytometry in Cellular Therapy

A clear understanding of the benefits and limitations of flow cytometry assays are critical to integrating flow cytometry services into the production workflow of cell therapy programs. Flow cytometry can offer a diverse set of assays in support of cell therapy programs by not only characterizing the composition of cellular products, but additionally analyzing their purity, potency, and safety (Table I). As an example, flow cytometry assays can provide valuable information on patient and donor samples as part of an allogeneic hematopoietic stem cell transplant (alloHSCT). Complications from alloHSCT may include graft vs. host disease (GVHD), host rejection of the graft, or opportunistic infections due to the immune-ablative treatments needed prior to transplant. Clinicians and cell therapy labs can navigate these complications in part with the information flow cytometry assays can provide. Firstly, central to flow cytometry assays is product characterization through measuring expression of lineage specific markers on cells. Lineage phenotyping assays can characterize the effectiveness of the patient’s immune-ablation preparation before infusion by measuring the frequency of multiple immune subsets present in blood samples in a single panel (e.g. T cell subsets using CD3, CD4, and CD8; B cells using CD19; and natural killer cells using CD56 and CD16). They can also be used to determine whether graft manipulation strategies have effectively selected or depleted the graft. In the case of haploidentical TCRαβ/CD19 depletion, the graft should contain a minimal number of CD3+/TCRαβ+ and CD19+ cells. Secondly, flow cytometry can identify potential reactivity between host and donor cells through co-culture assays either by measuring proliferation due to HLA mismatching in mixed lymphocyte reactivity assays (MLR), or by measuring direct killing in flow based cytotoxicity assays. Finally, flow cytometry can rapidly measure the reconstitution or chimerism of the recipient cells within the patient after transplant by using antibodies against recipient HLA and donor HLA molecules, and subsequently comparing the relative frequency of donor HLA+ cells to recipient HLA+ cells from small samples of blood. In this way, flow cytometry measurements can act as an in-house complement to widely used and standardized molecular assays for chimerism[10], with both types of assays able to analyze the kinetics of immune cell reconstitution to ensure sufficient levels of immune cells are present to fight off opportunistic infections. All of these assays provide essential information for both the clinician investigator and the cell therapy manager, identifying the current immune status of the patient, the quality and safety of the product, and the effectiveness of the infusion. Few other assay systems are as versatile as flow cytometry and can offer such breadth of information through assay customization. Further information on various functional assays are described below.

Table 1.

Flow Cytometry Assays

Types of flow cytometry assays Target cell populations or culture conditions Typical antibodies and staining conditions Assay purpose
Lineage phenotyping[11] Lymphocytes, monocytes, dendritic cells, T cells, B cells, NK cells, CD45, CD14, CD3, CD19, CD56, CD4, CD8, CD83, HLA A/B/C, HLA DR/DQ Characterizing frequency of hematopoietic cells
Multi-parameter analysis[12] Cellular activation, exhaustion, differentiation, memory status PD1, TIM3, CD28, CD56, CD45RA, CD19, CD27, CCR7, Differentiation makers, Chemokine receptors Rare or disease- associated subset identification
Median fluorescence intensity of receptor expression[13] Cultured, transfected, or transduced cells Genetically modified surface receptors, such as CAR-T receptors, humanized antibodies, truncated receptors Identification of expression level of transgenic receptors
Intracellular cytokine staining[14, 15] Activated T cells, Activated B cells Fixation and permeabilization, followed by cytokine and chemokine staining for IFNg, TNFa, IL10, IL4, Granzyme, Perforin Measurement of activation in response to antigen stimulation
Intracellular phosphoflow[16] Activated cells Fixation and permeabilization, followed by staining for phosphorylated proteins pSTAT, pERK, pNFkB, pSyk Measurement of signal transduction
Cell proliferation and division assays[17, 18] In vitro cell culture Vital Dyes (CFSE, Cell Trace, Vybrant), DNA dyes measuring content, lineage markers, absolute counting beads Measurement of division or proliferation after mono or co-culture
Cytotoxicity[19], suppression[20], apoptosis assays[21] In vitro cell culture Viability markers (Live Dead, Zombie, propidium iodide, 7AAD, Annexin V), lineage markers, absolute counting beads Measurement of target survival after mono-culture with drugs or co-culture with effectors
Spectral flow cytometry[22], Cytof cytometry[23], multiplexing[24] Multi-parameter analysis of cell subsets High dimensional combinations of surface and intracellular markers which cannot be run on traditional cytometers due to the large number of markers Deep phenotyping of cell subsets on samples of limited quantity

Lineage Phenotyping

Most prominently, flow cytometry is an exceptionally adaptable means to phenotype cells in suspension, limited only by availability of patient samples and antibody reagents and, for intracellular markers, their ability to penetrate the cell or nuclear membrane as described later. Flow cytometry panels are most frequently assembled from vendor antibodies targeting surface antigens which help define the population of interest. Advances in flow cytometry have steadily increased the number of surface and intracellular markers analyzed as part of a “typical” panel for phenotypic identification of cellular subsets[25, 26]. Commercially available flow cytometers have progressed from single laser systems reporting on two parameters to multiple laser instruments routinely processing data from eight to twelve parameters, with up to forty parameters investigable on specialized instruments[27]. Flow cytometry assays thus excel at population dynamics, evaluating the relative frequency of rare cell populations within large cell suspensions. To support these assays, monoclonal antibody reagents have also expanded in scope and utility, with multiple vendors offering unique fluorochrome conjugates for antibodies specific for both extracellular and intracellular markers. This dramatic increase in the size and complexity of the typical staining panel has enabled the precise characterization of unique cell populations hidden within heterogenous samples using minimal sample volume, often less than one million cells per staining sample.

As an example of incorporating flow cytometry assays for quality in production of a GMP cellular product, the production of polyclonal tumor-associated antigen specific T cell products targeting several types of malignancies, including solid tumors, Hodgkin’s lymphoma, and acute leukemias, utilized multiple flow cytometry panels both before culture initiation and after in vitro expansion of the tumor associated antigen specific T cells (TAATs)[28–31]. Two flow cytometry panels were employed before initiation of cultures (Table II) to both identify the frequency of immune cells within patient samples and characterize the differentiation status of patient T cells in detail. These assays provided an initial confirmation T cells were present for expansion into TAATs and provided a baseline to compare with other products for quality control. After expansion in vitro, products were phenotyped again by flow cytometry to confirm the TAAT product did not contain appreciable numbers of undesirable CD19+ or CD83+CD3− cells. Additionally, phenotyping post-expansion helped to identify the composition of products which suggested a trend towards more efficacious in responders vs. non-responders within the trial. For this trial, patients who received products which contained a greater frequency of CD8+ CD3+ cells versus CD4+ CD3+ cells trended towards a more favorable response than those with a lower frequency of CD8+ CD3+ cells. Further pre-clinical investigations using flow cytometry have subsequently identified alternative cytokine combinations which promote antigen specific CD8+ CD3+ T cell growth in in vitro cell cultures[32]. In this way, flow cytometry services are able to concretely inform on multiple aspects of patient health and product quality throughout the process of development and production.

Table 2.

Flow Cytometry Panels for Phenotyping Patient Samples

Pre-Culture Panel I Pre-Culture Panel II Post-Culture Panel I Post-Culture Panel II
CD14 FITC CD45RA FITC CD19 FITC HLA DR FITC
CD19 PE CD45RO PE CD16 PE CD56 PE CD83 PE
CD3 PerCP Vio700 CD3 PerCP Vio700 CD3 PerCP Vio700 CD3 PerCP Vio700
CD56 PE Vio770 CCR7 PE Vio770 CD4 PE Vio770
CD83 APC CD95 APC CD45 APC CD45 APC
CD8 APC Vio770 TCRγδ APC Vio770 CD8 APC Vio770
CD4 VioBlue CD62L VioBlue CD14 VioBlue CD14 VioBlue
CD83 VioGreen TCRαβ VioGreen Live Dead Aqua Live Dead Aqua

Proliferation

Proliferation assays measure the expansion of cellular subsets within mono or mixed cultures over time using antibodies identifying lineage markers or by tracking cells labeled with fluorescent proteins to quantify cell numbers. Flow cytometry allows for the counting of lineage specific subsets of cells within mixed cultures which otherwise could not be identified using traditional cell counting methods, such as quantifying the proliferation of CD3+ T cells in mixed cultures with CD3− feeder cells. In addition to lineage markers, cells can be labeled before initiation of the culture in several ways to measure division, including labeling protein[17, 33, 34] (CFSE, Cell Trace), membrane[35, 36] (PKH, DiO), or DNA[37–39] (BrdU). These assays measure proliferation by tracking either the distribution of labeled protein or membrane to daughter cells, or the accumulation of non-radioactive tracers within DNA as the cells divide. Lineage markers can also be added to further subdivide cell populations. Protein and lipid dyes can further identify individual cellular divisions through deconvolution of median fluorescence peak intensities as the dye is distributed to daughter cells during division, however protein dyes tend to have more distinct fluorescence peaks within daughter cells for deconvolution. Comparison between membrane and protein labeling procedures has been reviewed elsewhere[35]. Cells labeled with nucleic acid dyes help visualize the content of DNA for calculation of the frequency of cells within each stage of mitosis[40]. Of note is that while strong exogenous signals can stimulate synchronous division within cell populations, many cells, including lymphocytes, undergo asymmetrical division in response to stimulation, especially within mixed co-cultures which may mask the appearance of clear cell division peaks. This and other biological features of growing cells in culture (potential differences in cell death, exponential division, and variances of CFSE distribution) should be considered by biostatisticians when analyzing proliferation by protein dye dilution[41].

Mixed Lymphocyte Reaction (MLR)

Safety assays are essential for all cell therapy products, however traditional safety assays have often relied upon the use of radioactivity to measure both cytotoxicity[42, 43] and proliferation[44]. Flow cytometry has been instrumental in replacing radioactive compounds when evaluating reactivity between donor and recipient cell populations in co-culture assays. As an alternative to radioactive assays, flow cytometry assays use fluorescent compounds to stably label targets and effectors for evaluation in short and long term co-cultures. For both types of assays, fluorescence based assays offer a similar level of specificity

Similar to proliferation assays, mixed lymphocyte reactions (MLR) are specialized co-culture assays which measure one-way cell expansion of effectors due to allogeneic recognition of differences in minor or major histocompatibility antigens between two populations of lymphocytes[45]. Effector and irradiated target lymphocytes may each be labeled with fluorescent tracking dyes and co-cultured over time to measure the division of CD3+ effectors in response to allogeneic stimulation. Additional markers can be added to the flow panel, including lineage subtype markers (CD4, CD8) and other activation markers (CD25), to further identify reactive populations. MLR assays which compare cells with mismatched HLA molecules can stimulate division with 1×105 cells or fewer effectors[46]. MLR assays evaluating cells with matching HLA molecules can be challenging to yield reproducible results because very low frequencies of cells may be reactive (<0.1%), and thus may require substantial numbers (>1×107) of effectors to produce a reaction[47, 48].

Cytotoxicity Assays

Cytotoxicity assays are a critical assay to measure the safety and potency of products by quantifying the potential to kill targets by effector cells when co-cultured in varying ratios[49]. Cytotoxicity assays can be used as an assessment of product safety by using them to measure off target killing by the drug product. For instance, the tumor antigen-specific T cells (TAATs) mentioned above have also been tested in the context of a bone marrow transplant, where TAAT from the donor were expanded and then infused in the recipient of the transplant. In this instance, it would be prudent to test the alloreactivity of the TAAT from the donor with recipient target cells. For this, flow cytotoxicity assays would label targets with fluorescent tracking dye and enumerate the number of viable targets remaining after co-culture with varying numbers of effectors[50, 51]. Flow cytotoxicity assays to evaluate the potency of chimeric antigen receptor T cell (CAR-T cell) responses[52], NK cell responses[53], or other antigen specific T cell responses[54] all have been previously described and follow similar principles. The frequency of reactive effector cells should be referenced when calculating the total number of effectors to targets during setup of the cytotoxicity assay, especially if it is less than 100% of the total effector population. FDA guidance on cytotoxicity assays suggest a 30% reduction in viability as a baseline cytotoxic effect[55]. This would translate to a minimum of 3000 targets killed due to the specific activity of effector cells in an assay with 10,000 targets, thus a polyclonal effector population containing 5% reactive cells would need at least 60,000 effectors to reach a ratio of 1:1 of antigen reactive cells to targets. Positive and negative killing conditions would be necessary to inform the dynamic range of the flow cytotoxicity assay.

Suppression assays:

Similar to the proliferation assay, a suppression assay measures the ability of cells to divide and expand in the presence of either drug or cellular inhibitors. Unlike the proliferation assay where proliferation is typically analyzed based on the dilution of the fluorescent dye, in the suppression assay, the fluorescent dye is not diluted since the cells are not proliferating[20]. Suppressive regulatory T cells and mesenchymal stromal cells (MSCs) can inhibit T cell division and effector functions, B cell maturation, and monocyte differentiation, and are being investigated as therapies for various autoimmune diseases and transplant graft enhancement. Although flow-based suppression assays may utilize similar fluorescent protein probes as proliferation assays, suppression assays are more complex and require additional considerations as detailed for T regulatory assays[56]. Of note, these additional considerations include clear phenotypic characterization measuring viability and lineage markers of both responders and suppressors to ensure inter assay reproducibility from heterologous populations of suppressors; proper controls to demonstrate clear and consistent proliferation of the target cells in the absence of suppression; and titration controls to demonstrate suppressive activity operates at physiologic ratios of effectors to targets. Interestingly, additional flow cytometry assays can satisfy each of these controls.

Evaluation of cytokine production by intracellular cytokine staining (ICS) and cytokine secretion assays

T cells that recognize antigen or are mitogenically stimulated will secrete various cytokines such as IFNγ and TNFα. By blocking release of the intracellular protein from endocytic and golgi apparatus via golgi transport inhibitors (Brefeldin A and Monensin), these cytokines accumulate and can be readily detected after permeabilizing the cellular membrane and staining with fluorochrome conjugated antibodies[57, 58]. ICS can be combined with lineage phenotyping of surface markers, allowing for highly detailed analysis of subpopulation reactivity within mixed cultures[59], unlike ELISA or ELISPOT assays which measure bulk production of cytokine or are only semi-quantitative. For antigen screening assays, ICS allows for immediate identification of CD4 versus CD8 restriction of peptide epitopes when paired with the additional surface markers. Additionally, multiple cytokines can be identified within single co-cultures using ICS depending on channel availability, however the timing of cytokine production must be coordinated as both transport inhibitors are toxic after extended culture. Cells must be fixed and permeabilized to visualize the intracellular accumulation of cytokines, thus to sort live cytokine producing cells, cytokine secretion assays are used. Secretion assays use bi-specific antibodies which target both a cell surface antigen and the cytokine of interest[60]. Cells are labeled with the bi specific antibodies and exposed to the antigen in diluted conditions to limit capture of any neighboring cell cytokines. Fluorescent antibodies to the cytokine or other surface antigens are added before sorting the cytokine positive cells. For cell therapy labs, secretion assays do not yield high numbers of cells, and would be appropriate during development to isolate live antigen reactive cells for further sequencing of the T cell receptor.

Analysis of gene and protein expression

In addition to functional assays, flow cytometry can used to characterize protein and gene expression. Flow cytometry has been traditionally utilized in preclinical research to visualize transgenic expression in vitro and in vivo by genetically engineering fluorescent protein expression under specific promoters of interest[61]. This allows for identification of transfected cell populations, for example using GFP expression for single cell cloning after viral transduction. Other, more specialized fluorescence assays can measure hybridization of fluorescent probes to RNA and DNA on a single cell level (fluorescence in-situ hybridization; FISH). Flow based fluorescence in situ hybridization (flow-FISH) assays have been used to measure both telomere length[62] and HIV latency[63] within cells. Phosphoflow analysis of protein expression can be used during clinical development as a substitute for western blotting to measure activation of specific kinases[16]. Phosphoflow uses antibodies targeting phosphorylated forms of critical signaling molecules, such as MYC, ERK, and STAT, in combination with surface staining. This allows for a highly individualized analysis of cellular activation within mixed cell populations, normally not feasible when cells are processed as whole lysate for western blot. For cell therapy programs, each of these specialized assays may be more appropriate for product development or analysis of follow-up samples rather than as a product release assay.

High dimensional assays incorporating flow cytometry

High dimensional analysis of flow cytometry data can offer unparalleled insight of the activation status of rare and specialized lymphocyte subsets present within samples. These data are collected on specialized instruments designed to address the issues of spillover from fluorochromes into multiple channels present on standard flow instruments, with each instrumental approach unique. These instruments include multiplex immunoassay systems, Cytof cytometry, and spectral cytometry.

Multiplexed flow cytometry assays utilize a gradient of fluorescence intensities to create an array of pre-set analysis conditions. This array may be individual clones of cells which have been barcoded with different combinations of two or more fluorescent dyes, or the array may consist of antibody-bead conjugates which measure a defined set of protein molecules. The intensity of the fluorescence in two channels determines the specific marker or cell being investigated, while additional antibodies added to the array can either phenotype the cell of interest or quantify the concentration of the protein captured by the beads. Barcoding flow multiplex assays act as high throughput measurements of cell markers in response to activation[64, 65], while bead multiplex assays act as large scale sandwich ELISAs[66]. Multiplex assays can be tailored to investigate a wide range of conditions within the provided analyte, including cytokine production[67], induction of apoptosis[68], miRNA expression[69], and viral gene expression[70], and can accommodate up to 500 different analyte targets depending on the instrument and technology[71]. Instruments such as the Bio-plex (Bio-Rad) and Magpix (Luminex), can be integrated into SOPs testing for cell potency after stimulation, or for screening reactivity against drug candidates during development. Importantly, each multiplex kit incorporates standards to translate the detected fluorescence into a quantified level of protein. Most bead multiplex kits require pairing with vendor manufactured instruments to automate the detection and analysis components of the assay, however several vendors offer flow multiplex kits independent of instruments which are compatible on many standard cytometers.

Mass cytometry by time of flight, Cytof[72, 73], avoids the use of fluorochromes completely, and instead tags antibodies with rare heavy metal ions. Cells are vaporized within the machine, and mass spectrometry is performed on the resulting vapor cloud. Mass spectrometry is capable of extremely sensitive measurements of ion quantities on individual cells, more sensitive than traditional cytometry, and lacks the spillover between channels typical in fluorescence-based assays. Cytof data was initially reported using a panel of 20 markers[72] close to the limit for conventional flow cytometry systems, quickly expanding to 34 markers[74], and has continued to increase as techniques for combinatorial panels emerge[75]. As an assay system in support of cell therapy labs, Cytof requires a substantial financial investment in equipment and custom metal ion bound antibody reagents and ion free buffers, as well as a requirement for significant investment in personnel for technical expertise in panel design and bioinformatic data analysis. Additionally, product analyzed by Cytof instruments are vaporized within the chamber, precluding real-time application of the single cell information to recovery of the cells by sorting.

Spectral flow cytometry is an alternative to traditional flow cytometry which retains use of traditional fluorochrome tagged antibodies to label cells in suspension[76]. Unlike traditional flow cytometry which measures fluorescence intensity using single detectors per channel within a defined, narrow wavelength, spectral flow cytometry measures fluorescence intensity across multiple detectors covering a much wider range of wavelengths, from 373 nm to 812 nm. A unique spectral map is created for each fluorochrome, allowing for distinction between fluorescence markers with similar, but not identical, spectral patterns after unmixing of data. Although spectral flow does not require specialized metal-ion antibody conjugates like Cytof requires, it does require extensive analysis of background fluorescence patterns which consumes substantial amounts of sample and antibody during validations. Panels for spectral flow cytometry have reached 40 markers and beyond[77, 78], and instruments for sorting using spectral flow cytometry are now offered by vendors. Spectral flow cytometry assays are currently primarily limited by the availability of unique fluorochromes by vendors, and the intensive training and investment necessary to validate customized panels.

In addition to these instruments focused on high dimensional interrogation of samples, vendors can customize traditional flow instruments tailored for the needs of individual cell therapy programs. These customizations may include specialized violet fluorescence detectors for fine particle differentiation of exosomes[79], high throughput sampling for characterization of drug resistance candidates by flow cytometry[80–82], and unique nozzle tips in sorters to reduce shear forces experienced by sensitive cells during sorting[83]. Overall, several options are available for in depth analysis of patient and product samples.

Limitations with flow cytometry in cellular therapy

Although flow cytometry is a powerful tool in evaluating cellular therapies, it carries with it several technical and foundational challenges that are amplified when using them in a controlled setting such as cellular therapies manufactured according to good manufacturing practices (GMP). It is critical to understand these challenges exist as technical, conceptual, and logistical hurdles as cores initiate product development.

Firstly, flow cytometry protocols have an involved workflow for staining which involves many individual reagents. Maintaining a stock of antibodies and buffers that are in-date and have been qualified and titrated quickly escalates the investment cost of flow cytometry assays and makes proper scheduling and planning paramount. During process development, each antibody must be validated for antigen reactivity and titrated on the specific cell type of interest[84]. Titration is necessary to identify potential non-specific binding and autofluorescence to ensure accurate gating of positive and negative events[85]. These expenses and complications are further exaggerated when supporting disparate clinical protocols and their unique requirements for cell phenotyping[86]. Furthermore, these reagents are not necessarily as interchangeable as other consumable reagents between manufacturers, depending on the antibody clone and fluorochrome conjugate in use. Different antibody clones can react differently to staining and fixation conditions, and some conjugate fluorochromes cannot be subjected to the harsh alcohol fixatives necessary for staining of nuclear proteins. Re-qualification of reagents would be necessary to demonstrate any changes to fluorochrome panels do not negatively impact sample data and workspace templates. For all of these reasons, flow cytometry is exceedingly sensitive to supply chain constraints. Additionally, protocols must include validated controls to ensure confidence in the quality of staining cocktails for each assay performance. These may include vendor-purchased lyophilized cell preps, or previously validated lots of frozen cells, all of which must be stained with the same antibody cocktails as are used on the cell therapy product.

Once data are acquired, they must be analyzed by qualified technicians proficient in flow cytometry. One area that is particularly important and dictates the populations investigated in subsequent steps is cell gating. Because not all markers show a bimodal distribution and largely depend upon the biology of the cell therapy product, placing the gate has inherent subjectivity regardless of the controls used to determine the parameters of the gate. For instance, two main splice variants of CD45 (CD45RA and CD45RO) are used when identifying stage of differentiation within memory T cell populations, among other markers[11]. However, memory T cells may express an intermediate level of CD45RA and CD45RO as the cells progressively differentiate. Thus additional markers are necessary to properly gate different memory populations, including CCR7, CD95, CD28, and CD27. This is in contrast to the expression of lineage markers, such as CD4 and CD8 in T cell populations, or CD3 and CD19 between T and B cell populations, which are bimodal[87]. These differences in staining distribution are inherent in the biology of the markers in question. Thus technicians require not only a high degree of technical competency to perform the staining protocols, but also an understanding of the relationship between marker biology, instrument gains, fluorescence signal, and compensation theory. Changes to instrument gains will alter the relative spillover reported between neighboring channels, and potentially impair the ability to distinguish the true expression of multiple markers co-expressed on a single cell from false positives arising from improperly set gains and spillover error[88]. A thorough review of the actual vs. expected results by all parties during development- technicians, principal investigator, and quality assurance- will set clear benchmarks to refer against during clinical production runs.

Finally, flow cytometry staining protocols can become both time and technically intensive as they frequently involve multiple rounds of staining, with each round followed by washes and resuspensions. Viability, phenotyping, fixation, and intracellular staining each has its own independent set of antibodies, buffers, and blocking reagents. Time spent staining can take anywhere from thirty minutes per set of conditions (e.g. surface phenotyping for lineage markers on lymphocytes) to multi-hour or multi-day staining steps (e.g. marking cell passage number using vital dye dilution), again depending on the antibody and assay panel. Each of these staining steps are performed sequentially, and cell loss is not insignificant after each centrifugation and resuspension. Performing staining for multiple panels at the same time will improve efficiency at the cost of increased complexity and record keeping. Estimates of time needed for flow staining should include time spent running daily QC, template setup, and instrument cleaning steps before and after samples are run, and that samples must be run sequentially, often taking at least two minutes per sample. In total, flow staining protocols can occupy a technician for four hours or more per staining run not including time spent performing any additional pre-processing steps for cellular products, such as lymphocyte enrichment by density centrifugation or lysing of red blood cells. When creating budgets and evaluating personnel needs, leaders of cell therapy programs should also consider the time required for data analysis, inventory management, or document reporting.

Flow cytometry instruments are manufactured by multiple vendors, often equipped with similar laser lines and filter sets common for standard fluorochromes(Table III, [89]). When deciding between vendors, cell program managers should have clear assay requirements in hand before purchasing to ensure equipment meets both their immediate and their future needs based off their panel design; e.g. purchasing a flow cytometer with only one laser is likely to significantly restrict future use. Flow cytometers can remain in service for many years with modern solidstate lasers if the fluidics are properly maintained with yearly preventative maintenance, as proper cleaning procedures before and after every run, along with pre-filtering of samples to remove debris and aggregates, can be sufficient to prevent clogging. Analysis instruments may also be run using water for sheath buffer instead of PBS, avoiding potential salt buildup within internal fluidics since live cell recovery is not needed. Although equipment upgrades are available for equipment in certain circumstances, equipment replacement is often driven more by limitations in the number of available lasers and detectors rather than actual equipment breakdown.

Table 3.

Hardware and software considerations

Manufacturer (Analyzers) Manufacturer (Sorters) Manufacturer (Multiplex/Spectral/Cytof)
Accelix Beckman Coulter Bio-Rad
Beckman Coulter BD Biosciences BD Biosciences
BD Biosciences Luminex
Bio-Rad Cytek Cytek
Cytek Miltenyi Biotec Standard Biotools (Cytof)
Miltenyi Biotec Sony Sony
Sony EMD Millipore Sigma
Thermofisher

In addition to equipment hardware, flow cytometry may require software purchase for further analysis and presentation of data. FlowJo (BD) and FCS Express (Denovo) are two major software platforms for manual analysis of individual flow cytometry files, in addition to manufacturer-specific analysis software bundled with equipment or other independent software platforms. Both software packages have expanded into spectral flow, and both have modules for specialized analysis such as batched output of multiple samples, Boolean gating for multiparameter cytokine production, or best-fit curve analysis of DNA content or cell division peaks. Compensation can also be imported or performed within the software itself depending on user preferences. Data storage for flow cytometry files should be planned in advance with representatives from IT departments, as individual flow files can surpass 100 MB per file.

Analysis of high dimensional data and multiplex data also requires additional software and techniques for transformation to visualize clustering within the large data sets[90, 91], such as from Cytobank (Beckman Coulter), Matlab (Mathworks), and Bioconductor (open source). Incorporating computational gating practices using software during the validation process can standardize population identification and avoid the introduction of subjectivity due to manual gating. Finally, flow cytometry data can be processed using statistical software to perfom basic comparative analysis between experiments, such as Prism (Dotmatics) and Origin (Originlab). Consultation with biostatisticians is appropriate during validation and operation.

Validation of flow processes and panels

The Qualification of complex instruments such as a flow cytometer, including installation qualification (IQ), operational qualification (OQ), and performance qualification (PQ) is often provided by the manufacturer and is therefore not discussed here. Basic validation of cell based assays to characterize cell therapy products, specifically flow cytometry assays, require specialized considerations separate from manufacturer qualification or assays validating drugs or other small molecule compounds. Because reagents for flow cytometry have diversified in the absence of a similarly expansive collection of control cellular reference materials, recommendations and guidelines have been published to minimize the potential for heterogeneity in flow staining arising from different antibody panels, staining methods, equipment, and gating strategies[58, 84, 89, 92]. These recommendations provide in-depth analysis of several of the issues associated with adapting cell based flow assays into a regulatory framework for standardization of procedures, analysis, and documentation. The conditions listed here are for early phase clinical studies and ultimately qualification and validations of equipment and processes is at the discretion of the Quality Unit and must be done according to federal regulations and applicable accrediting standards. The Clinical Laboratory Standards Institute has released a first edition of its Validation of Assays Performed by Flow Cytometry (H62; [93]), while the National Institute of Standards and Technology has established the Flow Cytometry Standards Consortium as part of their Biomarker and Genomic Sciences Group in collaboration with industry and academic representatives[94]. This ongoing collaboration investigates and publishes recommendations for integration of specific flow cytometry assays in clinical trial design. Both of these resources offer extensive and detailed recommendations for assay validation and are excellent resources when designing specific protocols for validation of flow cytometry assays in clinical labs. Sample flow cytometry validation protocols have been published [95, 96], identifying four different validation strategies depending on the regulatory and clinical status of the laboratory in question. Validation of reagents and assays must also include considerations for whether flow analysis of samples will be performed at one central site or among all satellite locations during a clinical trial[97]. These considerations can include recommendations for storage and expiration of reagents, distribution of reference material, and qualification of technician adherence to SOP and training, all in order to ensure inter-site parity when assays performed at multiple sites[98].

Prior to initiating the validation, which is a formal process that has a set of acceptance criteria that are approved by responsible individuals such as the director of the program and quality assurance, staff should perform feasibility studies to identify how a given flow cytometry assay performs and get a general sense of the specificity, sensitivity, and other critical parameters that will need to be formally tested in the validation. Once comfortable with how the assay will perform and once a general understanding of the critical parameters and outcomes of the assay is determined, a formal validation plan for the assay is drafted. This includes a background of the assay with relevant references, the overall goal of the assay and in what context it will be used. It will also include how the validation will be performed and by whom, as well as the acceptance criteria for the assay and controls that will be included. The acceptance criteria is critical and requires careful thought and review of prior data; if subsequent tests do not meet the validation criteria then the validation fails. Finally, a summary of the results should be included as well as any change to the validation plan that occurred, when they occurred, and who approved the changes (prospectively).

Additional Considerations for Panel Design

Briefly, initial feasibility inquiries should identify how a flow cytometry assay can provide useful information for the cell therapy process in question. Some questions which are considered include what specific markers the flow assay must report, how frequent the cells of interest are compared to the total sample population[99], what antibody reagents are available for the assay, what instruments are available for use, and what positive and negative controls are available[100]. Investigators should first collect published studies to establish the utility and biological relevance of the markers chosen for the assay, ideally studies which identify the specific antibody clones used for the assay. These studies would also help identify the expected frequency and working range which the assay will function within, and the lower limit of detection to enumerate the input cell frequency, since flow cytometry assays which focus on rare event identification can be susceptible to errors arising from background noise and inherent statistical deviation if not enough events are collected. These errors can be minimized by increasing the power of the experiment by collecting both more events per sample and more replicates of samples, with a minimum of four independent collections to demonstrate interexperiment reproducibility[101, 102]. Deliberate panel design can also minimize fluorescence spillover from neighboring channels into the marker of interest. This can include pairing dim markers with bright fluorochromes to enhance detection, while avoiding pairing highly overlapping fluorochromes with markers co-expressed on the same cell, depending on available laser lines within the instrument. Common overlapping fluorochromes include PerCP-Cy5.5 and PE-Cy7 off the blue laser, APC, Alexa700, and APC-Cy7 off the red laser, and BV610, BV650, and BV711 off the violet laser. Antibodies and staining conditions are next evaluated for brightness and concentration through titrations on the cells of interest, testing different pairings of fluorochromes and antibodies to see which clone and fluorochrome combinations provide the highest relative staining intensity. Importantly, both positive and negative controls must be evaluated both during the initial stages of panel validation, and included as repeatable measures of quality control. Controls to test the precision and accuracy of flow cytometry assays represent the biggest hurdle for establishing QC and QA measurements for the panel. Vendor supplied lyophilized cellular controls provide fixed populations of cells to test in parallel with samples during staining protocols. These fixed controls provide published ranges for multiple surface markers of interest, with the option to investigate stimulated cells to test for activation marker expression. Aliquots of frozen PBMCs may also be used as an alternative to the vendor supplied lyophilized cells, however lab collected PBMCs may only report on the qualitative success of the staining assay (presence or absence of CD4+ cells) and not on its quantitative precision or accuracy (e.g. whether all CD4+ cells were labeled in the sample). Moreover, for non-standard or genetically altered markers outside of traditional lymphocytic phenotyping, managers will need to design custom fit-for-purpose controls to be assayed by flow[103]. These custom assays will need to undergo additional review by QC and technical staff to identify any revisions needed for the successful promotion of the plan to SOP. Once sufficient information has been gathered or tested in preliminary studies, these considerations must be written into the validation plan with specific goals for the assay to accomplish.

Finally, analysis of CAR-specific expression during product transduction necessitates a customized validation unique to the antigenic epitopes of the selected CAR receptor. Antibodies specific for the CAR idiotype in use may have less background staining as compared with reagents detecting other regions of the CAR receptor, but require their own lead time for development and testing[104]. Transduced patient samples also may show substantially different background and antigen specific staining patterns as compared with transduced cell lines. Time spent developing these CAR specific panels should be considered separate and performed in parallel to pre-clinical analysis of CAR function.

Training for personnel

Surveys of clinical flow cytometry laboratories identified several core competencies deemed necessary when staffing flow core services[105]. Flow cytometry technicians should understand basic concepts of size and viability gating to differentiate cells from debris. In addition, technicians should understand fundamental lymphocyte lineage markers, including size and staining differences between T cell, B cell, monocyte, NK cell, and granulocyte populations. These lineage markers would include CD45, CD19, CD20, CD3, CD4, CD8, CD14, CD56, and CD16. Specific knowledge of equipment software was not identified within the surveys as critical, however a working understanding of equipment cleaning, maintenance, and QC is beneficial to identify normal operating parameters for the instruments.

Credentialed certification in clinical flow cytometry is currently offered as a program jointly managed by the International Society for the Advancement of Cytometry (ISAC), the International Clinical Cytometry Society (ICCS), and credentialed by the American Society of Clinical Pathologists (ASCP) Board of Certification. This credential is for a Specialist in Cytometry (SYCM)[106], and has four routes to fulfill experience requirements. In addition to specialist certification, technicians may also choose to enroll in institution-specific flow cytometry training courses, courses offered by other societies such as American Clinical Chemistry Society (ACCS), or manufacturer-specific training courses which are primarily focused on cytometry equipment training. Certification, however, is an important resource to ensure a basic level of competence for diagnostic-focused flow cytometry, as training for research-focused flow cytometry has often relied upon past experience and not necessarily upon standardized credentials. Finally, the proper creation of compensation controls and the setting of gains to match the controls is frequently an area of confusion for technicians without such prior experience in flow cytometry[88]. Staff should be provided with in-depth training about applying compensation to samples. Firsthand empirical training will identify any product-specific issues more generalized flow training courses can miss. Once properly trained, QC can perform annualized competency exams by including both review questions and proficiency tests on control samples of unknown origin to help technicians maintain skills learned.

Summary

Flow cytometry has evolved over the last 55 years, starting as a tool with a limited range of detection to using mass/charge ratios as a way to massively expand the number of markers capable of being detected in a single sample. In many ways, flow cytometry is an ideal tool for cellular therapies: it is capable of detecting as narrow or nearly as broad a range of markers as required, it is highly sensitive, and allows for discriminating the functional parameters of a single cell of a given phenotype. While these characteristics provide significant advantages over other assays, these properties also make operation, training, and analysis quite complex. Moreover, high dimensional analysis of patient and product samples will gradually become standard operation across production labs, much as single cell sequencing for DNA and mRNA expression have become commonplace. The emergence of more automated instruments may mitigate some of these challenges, but they have yet to be widely adopted. Nevertheless, in more regulated settings such as products manufactured under good manufacturing practices, automated instruments with validated processes might further enable the rapid proliferation of cell and gene therapies.

Acknowledgements:

The Flow Cytometry Core, the Cellular Therapy Laboratory (CTL) Core, and the members of the Cell Enhancement and Technologies for Immunotherapy are supported by the Children’s National Research Institute, and the CTL is also supported by NCI 2P01CA148600–07A1 and NCI 5P01CA148600–09. The authors would also like to thank our donors, both known and anonymous, whom help to advocate for unrelenting improvements in our specialized care for all of our pediatric patients.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Conflict of Interest Disclosure:

PJH is a co-founder and on the board of directors of Mana Therapeutics. He is or was an advisor or on the scientific advisory board of: Cellenkos, Cellevolve, Discovery Life Sciences, Capsida, Autolomous, Microfluidx, and March Biosciences.

CAL has no disclosures.

References

  • [1].Jaye DL, Bray RA, Gebel HM, Harris WA, Waller EK, Translational applications of flow cytometry in clinical practice, J Immunol 188(10) (2012) 4715–9. [DOI] [PubMed] [Google Scholar]
  • [2].Cory JM, Ohlsson-Wilhelm BM, Brock EJ, Sheaffer NA, Steck ME, Eyster ME, Rapp F, Detection of human immunodeficiency virus-infected lymphoid cells at low frequency by flow cytometry, Journal of immunological methods 105(1) (1987) 71–8. [DOI] [PubMed] [Google Scholar]
  • [3].Fischl MA, Richman DD, Grieco MH, Gottlieb MS, Volberding PA, Laskin OL, Leedom JM, Groopman JE, Mildvan D, Schooley RT, et al. , The efficacy of azidothymidine (AZT) in the treatment of patients with AIDS and AIDS-related complex. A double-blind, placebo-controlled trial, The New England journal of medicine 317(4) (1987) 185–91. [DOI] [PubMed] [Google Scholar]
  • [4].Barlogie B, Latreille J, Freireich EJ, Fu CT, Mellard D, Meistrich M, Andreeff M, Characterization of hematologic malignancies by flow cytometry, Blood Cells 6(4) (1980) 719–44. [PubMed] [Google Scholar]
  • [5].Civin CI, Strauss LC, Brovall C, Fackler MJ, Schwartz JF, Shaper JH, Antigenic analysis of hematopoiesis. III. A hematopoietic progenitor cell surface antigen defined by a monoclonal antibody raised against KG-1a cells, J Immunol 133(1) (1984) 157–65. [PubMed] [Google Scholar]
  • [6].Sutherland DR, Anderson L, Keeney M, Nayar R, Chin-Yee I, The ISHAGE guidelines for CD34+ cell determination by flow cytometry. International Society of Hematotherapy and Graft Engineering, J Hematother 5(3) (1996) 213–26. [DOI] [PubMed] [Google Scholar]
  • [7].Herzenberg LA, Parks D, Sahaf B, Perez O, Roederer M, Herzenberg LA, The history and future of the fluorescence activated cell sorter and flow cytometry: a view from Stanford, Clin Chem 48(10) (2002) 1819–27. [PubMed] [Google Scholar]
  • [8].McKinnon KM, Flow Cytometry: An Overview, Curr Protoc Immunol 120 (2018) 5 1 1–5 1 11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Dunphy CH, Applications of flow cytometry and immunohistochemistry to diagnostic hematopathology, Arch Pathol Lab Med 128(9) (2004) 1004–22. [DOI] [PubMed] [Google Scholar]
  • [10].Blouin AG, Ye F, Williams J, Askar M, A practical guide to chimerism analysis: Review of the literature and testing practices worldwide, Hum Immunol 82(11) (2021) 838–849. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].De Rosa SC, Herzenberg LA, Herzenberg LA, Roederer M, 11-color, 13-parameter flow cytometry: identification of human naive T cells by phenotype, function, and T-cell receptor diversity, Nature medicine 7(2) (2001) 245–8. [DOI] [PubMed] [Google Scholar]
  • [12].Perez OD, Nolan GP, Simultaneous measurement of multiple active kinase states using polychromatic flow cytometry, Nature biotechnology 20(2) (2002) 155–62. [DOI] [PubMed] [Google Scholar]
  • [13].Demaret J, Varlet P, Trauet J, Beauvais D, Grossemy A, Hego F, Yakoub-Agha I, Labalette M, Monitoring CAR T-cells using flow cytometry, Cytometry B Clin Cytom 100(2) (2021) 218–224. [DOI] [PubMed] [Google Scholar]
  • [14].Lamoreaux L, Roederer M, Koup R, Intracellular cytokine optimization and standard operating procedure, Nat Protoc 1(3) (2006) 1507–16. [DOI] [PubMed] [Google Scholar]
  • [15].Lovelace P, Maecker HT, Multiparameter intracellular cytokine staining, Methods Mol Biol 699 (2011) 165–78. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Krutzik PO, Trejo A, Schulz KR, Nolan GP, Phospho flow cytometry methods for the analysis of kinase signaling in cell lines and primary human blood samples, Methods Mol Biol 699 (2011) 179–202. [DOI] [PubMed] [Google Scholar]
  • [17].Kim KH, Sederstrom JM, Assaying Cell Cycle Status Using Flow Cytometry, Curr Protoc Mol Biol 111 (2015) 28 6 1–28 6 11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Tario JD Jr., Muirhead KA, Pan D, Munson ME, Wallace PK, Tracking immune cell proliferation and cytotoxic potential using flow cytometry, Methods Mol Biol 699 (2011) 119–64. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Hoppner M, Luhm J, Schlenke P, Koritke P, Frohn C, A flow-cytometry based cytotoxicity assay using stained effector cells in combination with native target cells, Journal of immunological methods 267(2) (2002) 157–63. [DOI] [PubMed] [Google Scholar]
  • [20].Venken K, Thewissen M, Hellings N, Somers V, Hensen K, Rummens JL, Stinissen P, A CFSE based assay for measuring CD4+CD25+ regulatory T cell mediated suppression of auto-antigen specific and polyclonal T cell responses, Journal of immunological methods 322(1–2) (2007) 1–11. [DOI] [PubMed] [Google Scholar]
  • [21].Riccardi C, Nicoletti I, Analysis of apoptosis by propidium iodide staining and flow cytometry, Nat Protoc 1(3) (2006) 1458–61. [DOI] [PubMed] [Google Scholar]
  • [22].Ferrer-Font L, Pellefigues C, Mayer JU, Small SJ, Jaimes MC, Price KM, Panel Design and Optimization for High-Dimensional Immunophenotyping Assays Using Spectral Flow Cytometry, Curr Protoc Cytom 92(1) (2020) e70. [DOI] [PubMed] [Google Scholar]
  • [23].Lin D, Maecker HT, Mass Cytometry Assays for Antigen-Specific T Cells Using CyTOF, Methods Mol Biol 1678 (2018) 37–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Davies R, Vogelsang P, Jonsson R, Appel S, An optimized multiplex flow cytometry protocol for the analysis of intracellular signaling in peripheral blood mononuclear cells, Journal of immunological methods 436 (2016) 58–63. [DOI] [PubMed] [Google Scholar]
  • [25].De Rosa SC, Brenchley JM, Roederer M, Beyond six colors: a new era in flow cytometry, Nature medicine 9(1) (2003) 112–7. [DOI] [PubMed] [Google Scholar]
  • [26].Perfetto SP, Chattopadhyay PK, Roederer M, Seventeen-colour flow cytometry: unravelling the immune system, Nature reviews. Immunology 4(8) (2004) 648–55. [DOI] [PubMed] [Google Scholar]
  • [27].Nolan JP, Condello D, Spectral flow cytometry, Curr Protoc Cytom Chapter 1 (2013) 1 27 1–1 27 13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Hont AB, Cruz CR, Ulrey R, O’Brien B, Stanojevic M, Datar A, Albihani S, Saunders D, Hanajiri R, Panchapakesan K, Darko S, Banerjee P, Fortiz MF, Hoq F, Lang H, Wang Y, Hanley PJ, Dome JS, Bollard CM, Meany HJ, Immunotherapy of Relapsed and Refractory Solid Tumors With Ex Vivo Expanded Multi-Tumor Associated Antigen Specific Cytotoxic T Lymphocytes: A Phase I Study, J Clin Oncol 37(26) (2019) 2349–2359. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Dave H, Terpilowski M, Mai M, Toner K, Grant M, Stanojevic M, Lazarski C, Shibli A, Bien SA, Maglo P, Hoq F, Schore R, Glenn M, Hu B, Hanley PJ, Ambinder R, Bollard CM, Tumor-associated antigen-specific T cells with nivolumab are safe and persist in vivo in relapsed/refractory Hodgkin lymphoma, Blood Adv 6(2) (2022) 473–485. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Kinoshita H, Cooke KR, Grant M, Stanojevic M, Cruz CR, Keller M, Fortiz MF, Hoq F, Lang H, Barrett AJ, Liang H, Tanna J, Zhang N, Shibli A, Datar A, Fulton K, Kukadiya D, Zhang A, Williams KM, Dave H, Dome JS, Jacobsohn D, Hanley PJ, Jones RJ, Bollard CM, Outcome of donorderived TAA-T cell therapy in patients with high-risk or relapsed acute leukemia post allogeneic BMT, Blood Adv 6(8) (2022) 2520–2534. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Tanna JG, Ulrey R, Williams KM, Hanley PJ, Critical testing and parameters for consideration when manufacturing and evaluating tumor-associated antigen-specific T cells, Cytotherapy 21(3) (2019) 278–288. [DOI] [PubMed] [Google Scholar]
  • [32].Lazarski CA, Datar AA, Reynolds EK, Keller MD, Bollard CM, Hanley PJ, Identification of new cytokine combinations for antigen-specific T-cell therapy products via a high-throughput multiparameter assay, Cytotherapy 23(1) (2021) 65–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Lyons AB, Analysing cell division in vivo and in vitro using flow cytometric measurement of CFSE dye dilution, Journal of immunological methods 243(1–2) (2000) 147–54. [DOI] [PubMed] [Google Scholar]
  • [34].Lyons AB, Blake SJ, Doherty KV, Flow cytometric analysis of cell division by dilution of CFSE and related dyes, Curr Protoc Cytom Chapter 9 (2013) 9 11 1–9 11 12. [DOI] [PubMed] [Google Scholar]
  • [35].Wallace PK, Tario JD Jr., Fisher JL, Wallace SS, Ernstoff MS, Muirhead KA, Tracking antigendriven responses by flow cytometry: monitoring proliferation by dye dilution, Cytometry A 73(11) (2008) 1019–34. [DOI] [PubMed] [Google Scholar]
  • [36].Pitsillides CM, Runnels JM, Spencer JA, Zhi L, Wu MX, Lin CP, Cell labeling approaches for fluorescence-based in vivo flow cytometry, Cytometry A 79(10) (2011) 758–65. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [37].Dawson AE, Norton JA, Weinberg DS, Comparative assessment of proliferation and DNA content in breast carcinoma by image analysis and flow cytometry, The American journal of pathology 136(5) (1990) 1115–24. [PMC free article] [PubMed] [Google Scholar]
  • [38].Darzynkiewicz Z, Bedner E, Smolewski P, Flow cytometry in analysis of cell cycle and apoptosis, Seminars in hematology 38(2) (2001) 179–93. [DOI] [PubMed] [Google Scholar]
  • [39].Darzynkiewicz Z, Juan G, Bedner E, Determining cell cycle stages by flow cytometry, Curr Protoc Cell Biol Chapter 8 (2001) Unit 8 4. [DOI] [PubMed] [Google Scholar]
  • [40].Estevam J, Danaee H, Liu R, Ecsedy J, Trepicchio WL, Wyant T, Validation of a flow cytometry based G(2)M delay cell cycle assay for use in evaluating the pharmacodynamic response to Aurora A inhibition, Journal of immunological methods 363(2) (2011) 135–42. [DOI] [PubMed] [Google Scholar]
  • [41].Roederer M, Interpretation of cellular proliferation data: avoid the panglossian, Cytometry A 79(2) (2011) 95–101. [DOI] [PubMed] [Google Scholar]
  • [42].Dunkley M, Miller RG, Shortman K, A modified 51Cr release assay for cytotoxic lymphocytes, Journal of immunological methods 6(1–2) (1974) 39–51. [DOI] [PubMed] [Google Scholar]
  • [43].Miller RG, Dunkley M, Quantitative analysis of the 51Cr release cytotoxicity assay for cytotoxic lymphocytes, Cellular immunology 14(2) (1974) 284–302. [DOI] [PubMed] [Google Scholar]
  • [44].Bach FH, Voynow NK, One-way stimulation in mixed leukocyte cultures, Science 153(3735) (1966) 545–7. [DOI] [PubMed] [Google Scholar]
  • [45].Lazarski CA, Toner K, Keller MD, Luban N, Young PP, Bollard CM, Wagner SJ, Hanley PJ , Flow-based analysis of cell division identifies highly active populations within plasma products during mixed lymphocyte cultures, Blood Transfus 19(6) (2021) 456–466. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [46].Sprent J, Mixed Lymphocyte Reaction, in:Delves P (Ed.), Encyclopedia of Immunology (Second Edition), Elsevier; 1998, pp. 1733–1736. [Google Scholar]
  • [47].Michalek J, Collins RH, Hill BJ, Brenchley JM, Douek DC, Identification and monitoring of graftversus-host specific T-cell clone in stem cell transplantation, Lancet 361(9364) (2003) 1183–5. [DOI] [PubMed] [Google Scholar]
  • [48].Scheinberg P, Price DA, Ambrozak DR, Barrett AJ, Douek DC, Alloreactive T cell clonotype recruitment in a mixed lymphocyte reaction: implications for graft engineering, Experimental hematology 34(6) (2006) 788–95. [DOI] [PubMed] [Google Scholar]
  • [49].Zarcone D, Tilden AB, Cloud G, Friedman HM, Landay A, Grossi CE, Flow cytometry evaluation of cell-mediated cytotoxicity, Journal of immunological methods 94(1–2) (1986) 247–55. [DOI] [PubMed] [Google Scholar]
  • [50].Lecoeur H, Fevrier M, Garcia S, Riviere Y, Gougeon ML, A novel flow cytometric assay for quantitation and multiparametric characterization of cell-mediated cytotoxicity, Journal of immunological methods 253(1–2) (2001) 177–87. [DOI] [PubMed] [Google Scholar]
  • [51].Kim GG, Donnenberg VS, Donnenberg AD, Gooding W, Whiteside TL, A novel multiparametric flow cytometry-based cytotoxicity assay simultaneously immunophenotypes effector cells: comparisons to a 4 h 51Cr-release assay, Journal of immunological methods 325(1–2) (2007) 51–66. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [52].Kiesgen S, Messinger JC, Chintala NK, Tano Z, Adusumilli PS, Comparative analysis of assays to measure CAR T-cell-mediated cytotoxicity, Nat Protoc 16(3) (2021) 1331–1342. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [53].Wagner AK, Alici E, Lowdell MW, Characterization of human natural killer cells for therapeutic use, Cytotherapy 21(3) (2019) 315–326. [DOI] [PubMed] [Google Scholar]
  • [54].Godoy-Ramirez K, Makitalo B, Thorstensson R, Sandstrom E, Biberfeld G, Gaines H, A novel assay for assessment of HIV-specific cytotoxicity by multiparameter flow cytometry, Cytometry A 68(2) (2005) 71–80. [DOI] [PubMed] [Google Scholar]
  • [55].I.O.f. Standardization, Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity, 2009, p. 34. [Google Scholar]
  • [56].McMurchy AN, Levings MK, Suppression assays with human T regulatory cells: a technical guide, European journal of immunology 42(1) (2012) 27–34. [DOI] [PubMed] [Google Scholar]
  • [57].Yin Y, Mitson-Salazar A, Prussin C, Detection of Intracellular Cytokines by Flow Cytometry, Curr Protoc Immunol 110 (2015) 6 24 1–6 24 18. [DOI] [PubMed] [Google Scholar]
  • [58].Jaimes MC, Maecker HT, Yan M, Maino VC, Hanley MB, Greer A, Darden JM, D’Souza MP, Quality assurance of intracellular cytokine staining assays: analysis of multiple rounds of proficiency testing, Journal of immunological methods 363(2) (2011) 143–57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [59].Keller MD, Harris KM, Jensen-Wachspress MA, Kankate VV, Lang H, Lazarski CA, Durkee-Shock J, Lee PH, Chaudhry K, Webber K, Datar A, Terpilowski M, Reynolds EK, Stevenson EM, Val S, Shancer Z, Zhang N, Ulrey R, Ekanem U, Stanojevic M, Geiger A, Liang H, Hoq F, Abraham AA, Hanley PJ, Cruz CR, Ferrer K, Dropulic L, Gangler K, Burbelo PD, Jones RB, Cohen JI, Bollard CM, SARS-CoV-2-specific T cells are rapidly expanded for therapeutic use and target conserved regions of the membrane protein, Blood 136(25) (2020) 2905–2917. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [60].Manz R, Assenmacher M, Pfluger E, Miltenyi S, Radbruch A, Analysis and sorting of live cells according to secreted molecules, relocated to a cell-surface affinity matrix, Proceedings of the National Academy of Sciences of the United States of America 92(6) (1995) 1921–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [61].Fujiki Y, Tao K, Bianchi DW, Giel-Moloney M, Leiter AB, Johnson KL, Quantification of green fluorescent protein by in vivo imaging, PCR, and flow cytometry: comparison of transgenic strains and relevance for fetal cell microchimerism, Cytometry A 73(2) (2008) 11–118. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [62].Baerlocher GM, Vulto I, de Jong G, Lansdorp PM, Flow cytometry and FISH to measure the average length of telomeres (flow FISH), Nat Protoc 1(5) (2006) 2365–76. [DOI] [PubMed] [Google Scholar]
  • [63].Freen-van Heeren JJ, Addressing HIV-1 latency with Flow-FISH: Finding, characterizing and targeting HIV-1 infected cells, Cytometry A 99(9) (2021) 861–865. [DOI] [PubMed] [Google Scholar]
  • [64].Krutzik PO, Clutter MR, Trejo A, Nolan GP, Fluorescent cell barcoding for multiplex flow cytometry, Curr Protoc Cytom Chapter 6 (2011) 6 31 1–6 31 15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [65].Krutzik PO, Nolan GP, Fluorescent cell barcoding in flow cytometry allows high-throughput drug screening and signaling profiling, Nat Methods 3(5) (2006) 361–8. [DOI] [PubMed] [Google Scholar]
  • [66].Vignali DA, Multiplexed particle-based flow cytometric assays, Journal of immunological methods 243(1–2) (2000) 243–55. [DOI] [PubMed] [Google Scholar]
  • [67].Chowdhury F, Williams A, Johnson P, Validation and comparison of two multiplex technologies, Luminex and Mesoscale Discovery, for human cytokine profiling, Journal of immunological methods 340(1) (2009) 55–64. [DOI] [PubMed] [Google Scholar]
  • [68].Curpan RF, Simons PC, Zhai D, Young SM, Carter MB, Bologa CG, Oprea TI, Satterthwait AC, Reed JC, Edwards BS, Sklar LA, High-throughput screen for the chemical inhibitors of antiapoptotic bcl-2 family proteins by multiplex flow cytometry, Assay Drug Dev Technol 9(5) (2011) 465–74. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [69].Wu M, Piccini M, Koh CY, Lam KS, Singh AK, Single cell microRNA analysis using microfluidic flow cytometry, PloS one 8(1) (2013) e55044. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [70].Lee WM, Grindle K, Pappas T, Marshall DJ, Moser MJ, Beaty EL, Shult PA, Prudent JR, Gern JE, High-throughput, sensitive, and accurate multiplex PCR-microsphere flow cytometry system for large-scale comprehensive detection of respiratory viruses, J Clin Microbiol 45(8) (2007) 2626–34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [71].Luminex, xMAP Intelliflex Systems. https://www.luminexcorp.com/intelliflex/#overview.
  • [72].Bandura DR, Baranov VI, Ornatsky OI, Antonov A, Kinach R, Lou X, Pavlov S, Vorobiev S, Dick JE, Tanner SD, Mass cytometry: technique for real time single cell multitarget immunoassay based on inductively coupled plasma time-of-flight mass spectrometry, Anal Chem 81(16) (2009) 6813–22. [DOI] [PubMed] [Google Scholar]
  • [73].Iyer A, Hamers AAJ, Pillai AB, CyTOF((R)) for the Masses, Frontiers in immunology 13 (2022) 815828. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [74].Bendall SC, Simonds EF, Qiu P, Amir el AD, Krutzik PO, Finck R, Bruggner RV, Melamed R, Trejo A, Ornatsky OI, Balderas RS, Plevritis SK, Sachs K, Pe’er D, Tanner SD, Nolan GP, Single-cell mass cytometry of differential immune and drug responses across a human hematopoietic continuum, Science 332(6030) (2011) 687–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [75].Abdelaal T, Hollt T, van Unen V, Lelieveldt BPF, Koning F, Reinders MJT, Mahfouz A, CyTOFmerge: integrating mass cytometry data across multiple panels, Bioinformatics 35(20) (2019) 4063–4071. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [76].Ferrer-Font L, Small SJ, Lewer B, Pilkington KR, Johnston LK, Park LM, Lannigan J, Jaimes MC, Price KM, Panel Optimization for High-Dimensional Immunophenotyping Assays Using Full-Spectrum Flow Cytometry, Curr Protoc 1(9) (2021) e222. [DOI] [PubMed] [Google Scholar]
  • [77].Sahir F, Mateo JM, Steinhoff M, Siveen KS, Development of a 43 color panel for the characterization of conventional and unconventional T-cell subsets, B cells, NK cells, monocytes, dendritic cells, and innate lymphoid cells using spectral flow cytometry, Cytometry A (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [78].Park LM, Lannigan J, Jaimes MC, OMIP-069: Forty-Color Full Spectrum Flow Cytometry Panel for Deep Immunophenotyping of Major Cell Subsets in Human Peripheral Blood, Cytometry A 97(10) (2020) 1044–1051. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [79].Morales-Kastresana A, Jones JC, Flow Cytometric Analysis of Extracellular Vesicles, Methods Mol Biol 1545 (2017) 215–225. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [80].Edwards BS, Oprea T, Prossnitz ER, Sklar LA, Flow cytometry for high-throughput, high-content screening, Curr Opin Chem Biol 8(4) (2004) 392–8. [DOI] [PubMed] [Google Scholar]
  • [81].Edwards BS, Young SM, Saunders MJ, Bologa C, Oprea TI, Ye RD, Prossnitz ER, Graves SW, Sklar LA, High-throughput flow cytometry for drug discovery, Expert Opin Drug Discov 2(5) (2007) 685–96. [DOI] [PubMed] [Google Scholar]
  • [82].Sklar LA, Carter MB, Edwards BS, Flow cytometry for drug discovery, receptor pharmacology and high-throughput screening, Current opinion in pharmacology 7(5) (2007) 527–34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [83].Bhagwat N, Dulmage K, Pletcher CH Jr., Wang L, DeMuth W, Sen M, Balli D, Yee SS, Sa S, Tong F, Yu L, Moore JS, Stanger BZ, Dixon EP, Carpenter EL, An integrated flow cytometry-based platform for isolation and molecular characterization of circulating tumor single cells and clusters, Scientific reports 8(1) (2018) 5035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [84].Owens MA, Vall HG, Hurley AA, Wormsley SB, Validation and quality control of immunophenotyping in clinical flow cytometry, Journal of immunological methods 243(1–2) (2000) 33–50. [DOI] [PubMed] [Google Scholar]
  • [85].Hulspas R, Titration of fluorochrome-conjugated antibodies for labeling cell surface markers on live cells, Curr Protoc Cytom Chapter 6 (2010) Unit 6 29. [DOI] [PubMed] [Google Scholar]
  • [86].Hulspas R, O’Gorman MR, Wood BL, Gratama JW, Sutherland DR, Considerations for the control of background fluorescence in clinical flow cytometry, Cytometry B Clin Cytom 76(6) (2009) 355–64. [DOI] [PubMed] [Google Scholar]
  • [87].Erez A, Vogel R, Mugler A, Belmonte A, Altan-Bonnet G, Modeling of cytometry data in logarithmic space: When is a bimodal distribution not bimodal?, Cytometry A 93(6) (2018) 611–619. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [88].Roederer M, Spectral compensation for flow cytometry: visualization artifacts, limitations, and caveats, Cytometry 45(3) (2001) 194–205. [DOI] [PubMed] [Google Scholar]
  • [89].Green CL, Brown L, Stewart JJ, Xu Y, Litwin V, Mc Closkey TW, Recommendations for the validation of flow cytometric testing during drug development: I instrumentation, Journal of immunological methods 363(2) (2011) 104–19. [DOI] [PubMed] [Google Scholar]
  • [90].Saeys Y, Van Gassen S, Lambrecht BN, Computational flow cytometry: helping to make sense of high-dimensional immunology data, Nature reviews. Immunology 16(7) (2016) 449–62. [DOI] [PubMed] [Google Scholar]
  • [91].den Braanker H, Bongenaar M, Lubberts E, How to Prepare Spectral Flow Cytometry Datasets for High Dimensional Data Analysis: A Practical Workflow, Frontiers in immunology 12 (2021) 768113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [92].O’Hara DM, Xu Y, Liang Z, Reddy MP, Wu DY, Litwin V, Recommendations for the validation of flow cytometric testing during drug development: II assays, Journal of immunological methods 363(2) (2011) 120–34. [DOI] [PubMed] [Google Scholar]
  • [93].C.L.S. Institute, Validation of Assays Performed by Flow Cytometry, General Laboratory, Hematology, Standards, 2021.
  • [94].N.I.o.S.a. Technology, Quantitatve Flow Cytometry Standards, in: M.M. Laboratory; (Ed.). [Google Scholar]
  • [95].Mfarrej B, Gaude J, Couquiaud J, Calmels B, Chabannon C, Lemarie C, Validation of a flow cytometry-based method to quantify viable lymphocyte subtypes in fresh and cryopreserved hematopoietic cellular products, Cytotherapy 23(1) (2021) 77–87. [DOI] [PubMed] [Google Scholar]
  • [96].Selliah N, Nash V, Eck S, Green C, Oldaker T, Stewart J, Vitaliti A, Litwin V, Flow Cytometry Method Validation Protocols, Curr Protoc 3(8) (2023) e868. [DOI] [PubMed] [Google Scholar]
  • [97].Maecker HT, McCoy JP Jr., Consortium FHI, Amos M, Elliott J, Gaigalas A, Wang L, Aranda R, Banchereau J, Boshoff C, Braun J, Korin Y, Reed E, Cho J, Hafler D, Davis M, Fathman CG, Robinson W, Denny T, Weinhold K, Desai B, Diamond B, Gregersen P, Di Meglio P, Nestle FO, Peakman M, Villanova F, Ferbas J, Field E, Kantor A, Kawabata T, Komocsar W, Lotze M, Nepom J, Ochs H, O’Lone R, Phippard D, Plevy S, Rich S, Roederer M, Rotrosen D, Yeh JH, A model for harmonizing flow cytometry in clinical trials, Nature immunology 11(11) (2010) 975–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [98].Finak G, Langweiler M, Jaimes M, Malek M, Taghiyar J, Korin Y, Raddassi K, Devine L, Obermoser G, Pekalski ML, Pontikos N, Diaz A, Heck S, Villanova F, Terrazzini N, Kern F, Qian Y, Stanton R, Wang K, Brandes A, Ramey J, Aghaeepour N, Mosmann T, Scheuermann RH, Reed E, Palucka K, Pascual V, Blomberg BB, Nestle F, Nussenblatt RB, Brinkman RR, Gottardo R, Maecker H, McCoy JP, Standardizing Flow Cytometry Immunophenotyping Analysis from the Human ImmunoPhenotyping Consortium, Scientific reports 6 (2016) 20686. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [99].Rosenblatt JI, Hokanson JA, McLaughlin SR, Leary JF, Theoretical basis for sampling statistics useful for detecting and isolating rare cells using flow cytometry and cell sorting, Cytometry 27(3) (1997) 233–8. [DOI] [PubMed] [Google Scholar]
  • [100].Wang L, Hoffman RA, Standardization, Calibration, and Control in Flow Cytometry, Curr Protoc Cytom 79 (2017) 1 3 1–1 3 27. [DOI] [PubMed] [Google Scholar]
  • [101].Kalina T, Reproducibility of Flow Cytometry Through Standardization: Opportunities and Challenges, Cytometry A 97(2) (2020) 137–147. [DOI] [PubMed] [Google Scholar]
  • [102].Mizrahi O, Ish Shalom E, Baniyash M, Klieger Y, Quantitative Flow Cytometry: Concerns and Recommendations in Clinic and Research, Cytometry B Clin Cytom 94(2) (2018) 211–218. [DOI] [PubMed] [Google Scholar]
  • [103].Berg Luecke L, Waas M, Gundry RL, Reliable Protocols for Flow Cytometry Analysis of Intracellular Proteins in Pluripotent Stem Cell Derivatives: A Fit-For-Purpose Approach, Curr Protoc Stem Cell Biol 50(1) (2019) e94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [104].Sarikonda G, Pahuja A, Kalfoglou C, Burns K, Nguyen K, Ch’en IL, Pollner R, Tangri S, Dakappagari N, Monitoring CAR-T cell kinetics in clinical trials by multiparametric flow cytometry: Benefits and challenges, Cytometry B Clin Cytom 100(1) (2021) 72–78. [DOI] [PubMed] [Google Scholar]
  • [105].Davis D, Nadder T, CLS entry level competencies in flow cytometry, Clin Lab Sci 24(1) (2011) 29–34. [PubMed] [Google Scholar]
  • [106].A.S.o.C. Pathology, Specialist in Cytometry (ASCP), 2023. https://www.ascp.org/content/docs/default-source/boc-pdfs/boc-usguidelines/scym_content_outline.pdf?sfvrsn=10.

RESOURCES