Skip to main content
. 2024 Feb 5;10(3):e25512. doi: 10.1016/j.heliyon.2024.e25512

Table 3.

Summary of pH sensor absolute discrepancies, limits of agreement, and correlations of [H+] discrepancy to pH, osmolality, and batch age.

pH Sensor type Average absolute discrepancy (pH units)a 95% Limits of Agreement
Correlationb of discrepancy to:
From To CCA pH Bioreactor Osmolality Batch Age
STR EP 0.047 −0.129 0.049 Yes (***) No Yes (**)
SU EP 0.044 −0.172 0.089 Yes (****) No Yes (**)
SU OPT 0.072 −0.201 0.192 Yes (****) Yes (*) No
a

Calculated using the average discrepancy in [H+] measurements and transformed into pH difference from pH 7.1.

b

Correlation refers to the statistical significance by Kruskal-Wallis test; p value summaries are p < 0.05 (*), p < 0.01(**), p < 0.001 (***) and p < 0.0001(****).