Skip to main content
. 2024 Feb 19;8:24. doi: 10.1186/s41747-024-00425-5

Table 3.

Characteristics of self-expanding metal stents based on study findings

Uncovered Wallflex™ EGIS single bare Zilver 635® E-Luminexx™
Radial resistive force Low High Moderate High
Chronic outward force Low High Moderate High
Crush resistance High High Low Moderate
Axial force High Low Low High
Conformability Good Moderate Moderate Poor
Surface quality Poor Good Moderate Poor
Foreshortening High High None None
Radiopacity Good Poor Good Good