. 2023 Dec 10;9(4):101416. doi: 10.1016/j.adro.2023.101416

Table 1.

MR imaging and radiation specifications between HiRes Ultra cochlear implant and off-label usage on MRIdian

Condition Advanced Bionic HiRes Ultra MR specification ViewRay MRIdian Linac specification
Static B0 field 1.5 T 0.35 T
Maximum spatial gradient 20 T/m 7 T/m (static gradient outside covers)
18 mT/m (dynamic gradient field strength)
Max SAR 2.0 W/kg 1.2 W/kg
Max slew rate 200 T/m/sec per axis 160 T/m/sec per axis
Duration after implantation 4 weeks after implant N/A
Radiation maximum energy 15 MV 6 MV
Radiation maximum dose 250 Gy N/A

Abbreviations: MR = magnetic resonance; SAR = specific absorption rate.