Table 1.
Metrics | T1w | CE-T1w | FLAIR |
---|---|---|---|
Quantitative contrast metrics | |||
Signal-to-noise ratio
|
13.73 ± 4.89 | 23.77 ± 12.02 | 13.09 ± 6.15 |
Contrast-to-noise ratio
|
7.44 ± 2.87 | 18.49 ± 9.57 | 10.77 ± 5.38 |
Quantitative segmentation metrics | |||
Dice similarity coefficient
|
0.67 ± 0.05 | - | 0.68 ± 0.05 |
Absolute volume [mL] | 3.073 ± 0.563 mL | 2.984 ± 0.506 mL | 3.787 ± 0.679 mL |
Percentage volume difference
|
3.52 ± 12.61 | - | 28.02 ± 19.02 |
Absolute percentage volume difference
|
10.57 ± 7.60 | - | 28.43 ± 18.40 |
Pearson’ correlation | 0.77 | - | 0.67 |
Spatial variability metrics | |||
Error distribution per slice |
Data are presented as mean ± standard deviation. Quantitative contrast metrics: contrast-to-noise ratio and signal-to-noise ratio between choroid plexus (ChP) manual segmentation (MSeg) obtained from each sequence and a reference region obtained excluding the ChP from the ventricles (for details, see the “Quantitative contrast metrics” section). Quantitative segmentation metrics: Dice similarity coefficient, absolute volume, absolute percentage volume difference, and Pearson volume correlation analysis between MSegs obtained from non-contrast enhanced sequences (T1w and FLAIR MSeg) and the reference contrast-enhanced T1-weighted MSeg. FN, False negative; FP, False positive; TP, True positive