Skip to main content
. 2022 Jun 13;18(1):143–147. doi: 10.1177/19322968221099879

Table 1.

Unadjusted and Propensity Score–Adjusted Accuracy Metrics for the G5, 19 G6, 27 and G7 17 CGM Systems.

Location Unadjusted Adjusted
Abdomen Arm Abdomen Arm
System G5 G6 G7 G7 G5 G6 G7 G7
%15/15 86.1 82.1 85.5 89.6 86.1 81.9 85.5 89.6
%20/20 93.2 92.5 93.2 95.3 93.1 92.3 93.2 95.3
%30/30 98.1 98.9 98.1 98.8 98.2 98.8 98.1 98.8
MARD 9.0 9.9 9.1 8.2 9.0 9.9 9.1 8.2

%XX/XX, percentage of values within XX mg/dL or within XX% of corresponding YSI values ≤ or >100 mg/dL, respectively.

Abbreviations: CGM, continuous glucose monitoring; MARD, mean absolute relative difference.