Skip to main content
. 2022 Jun 13;18(1):143–147. doi: 10.1177/19322968221099879

Table 2.

Alert Performance of the G5, G6, and G7 CGM Systems.

Location Abdomen Arm
System G5 G6 G7 G7
Missed hypoglycemia detection rate (threshold of 55 mg/dL), % 32 36.1 26.7 24.2
Missed hypoglycemia detection rate (threshold of 70 mg/dL), % 9 14.0 12.4 11.2
False hypoglycemia alert rate (threshold of 55 mg/dL) a , % 3.2 9.2 14.8 8.7
a

False hypoglycemia alert rate: No YSI values <70 mg/dL within 15 minutes of CGM value(s) ≤55 mg/dL.