Table 2.
Location | Abdomen | Arm | ||
---|---|---|---|---|
System | G5 | G6 | G7 | G7 |
Missed hypoglycemia detection rate (threshold of 55 mg/dL), % | 32 | 36.1 | 26.7 | 24.2 |
Missed hypoglycemia detection rate (threshold of 70 mg/dL), % | 9 | 14.0 | 12.4 | 11.2 |
False hypoglycemia alert rate (threshold of 55 mg/dL) a , % | 3.2 | 9.2 | 14.8 | 8.7 |
False hypoglycemia alert rate: No YSI values <70 mg/dL within 15 minutes of CGM value(s) ≤55 mg/dL.
Abbreviation: CGM, continuous glucose monitoring.