Table 4. Comparison of Protein Quantification Methods Applied to NIST SRM BSA.
| method | target compound | matrix | relative uncertainty U, k = 2 (%) | reference |
|---|---|---|---|---|
| online ID CE/ICP–MS | albumin | serum | 6.7 | this work |
| ID ICP–MS | albumin | aqueous standard solution | 3.6 | Lemke et al.30 |
| AAA ID LC–MS | albumin | aqueous standard solution | 2.0 | NIST SRM 927e BSA |
| AAA ID LC–MS | albumin | aqueous standard solution | 0.8 | NIST SRM 927f BSA |