Skip to main content
. 2023 Dec 19;21(1):e00295. doi: 10.1016/j.neurot.2023.10.005

Table 2.

Charge density stimulation parameters.

Pulse width (μs) Pulse amplitude (mA) Charge density (μC/cm2)
30 5 2.5
30 10 5
60 2.5 2.5
60 5 5
60 10 10
120 2.5 5
120 5 10
120 10 20
250 1.25 5
250 2.5 10
250 5 20

Stimulation was varied in pulse width and amplitude, and multiple combinations of both were chosen to test set charge density levels. Given that the system has a total resistance of approximately 1 ​kOhm, a 1 ​mA pulse has a potential of 1 ​V.