Ensure adequate initial training and periodic review of the healthcare staff measuring BP
Use validated BP monitors listed by the British and Irish Hypertension Society
Ensure BP devices are appropriately validated, maintained, and regularly recalibrated
Measure BP in a relaxed, temperate setting, with the person quiet and seated, legs uncrossed, and their arm outstretched and supported at heart level
Use BP cuff of size appropriate to the arm circumference
On initial assessment, measure BP in both arms; repeat the measurements if the difference is >15 mmHg. If the difference remains, measure further BP in the arm with the higher reading
If office BP is ≥140/90 mmHg, repeat the measurement. If the second measurement is substantially different from the first, take a third measurement and record the lower of the last two measurements as the office BP
|