Table 2.

Evaluation of ventilator-sharing system (VSS) in benchtop testing using two test lungs under volume control mode.

Simulated patient 1 Simulated patient 2 Simulated patient 1 Simulated patient 2
PB-840 VC settings Test lung VSS setting NM3 measurements Test lung VSS setting NM3 measurements Error Vt %Error Vt Error Vt %Error Vt
IP PEEP Ti RR Crs R Vt PEEP Vt Crs R Vt PEEP Vt
22 10 1.25 24 25 5 300 9.4 297 25 5 500 9.8 503 −3 1.0% 3 0.6%
22 10 1.25 24 25 5 300 10.8 303 50 20 720 16 718 3 1.0% −2 0.3%
22 10 1.25 24 25 20 300 10 302 25 20 500 10.8 500 2 0.7% 0 0.0%
22 10 1.25 24 25 20 300 10.6 296 50 5 720 13.7 727 −4 1.3% 7 1.0%
22 10 1.25 24 50 5 300 12.4 308 50 5 720 13.6 721 8 2.7% 1 0.1%
22 10 1.25 24 50 5 300 11 293 25 20 500 10.7 502 −7 2.3% 2 0.4%
26 10 1.25 24 50 20 300 12.5 291 50 20 720 16.2 712 −9 3.0% −8 1.1%
26 10 1.25 24 50 20 300 11.8 302 25 5 500 10.4 501 2 0.7% 1 0.2%
22 10 1.25 12 25 5 300 9.7 298 25 5 500 10.1 498 −2 0.7% −2 0.4%
22 10 1.25 12 25 5 300 9.6 308 50 20 720 10.1 722 8 2.7% 2 0.3%
22 10 1.25 12 25 20 300 9.7 299 25 20 500 10.1 500 −1 0.3% 0 0.0%
22 10 1.25 12 25 20 300 9.8 296 50 5 720 10.3 728 −4 1.3% 8 1.1%
22 10 1.25 12 50 5 300 9.7 303 50 5 720 10.1 715 3 1.0% −5 0.7%
22 10 1.25 12 50 5 300 9.6 301 25 20 500 10.1 498 1 0.3% −2 0.4%
22 10 1.25 12 50 20 300 9.6 298 50 20 720 10.1 718 −2 0.7% −2 0.3%
22 10 1.25 12 50 20 300 9.8 301 25 5 500 10.2 498 1 0.3% −2 0.4%
Mean −0.25 0.06
Stdev 4.73 4.01
Absolute mean % error 1.3% 0.5%

VC = volume control ventilation mode; IP = inspiratory pressure (cmH2O); PEEP = positive end expiratory pressure (cmH2O); Ti = inspiratory time (s); RR = respiratory rate (breaths per minute, bpm); Crs = dynamic compliance (mL/cmH2O); R = resistance (cmH2O/L/s); Vt = tidal volume (mL); stdev = standard deviation.