Skip to main content
. 2024 Feb 19;18:1321872. doi: 10.3389/fnint.2024.1321872

Table 2.

In-vitro electrochemical properties of materials used in bioelectronic medicines.

Material Application |Z|@1kHz (Ω cm2) Electrolysis window (V) CSCc (mC/cm2) Max Qinj (mC/cm2) Uses in bioelectronic medicines References
Platinum (Pt) and Platinum/Iridium (Pt/Ir) Stimulation/Recording 0.4–4.7 −0.6–0.8 4.4–8.0 0.1–0.3 Cochlear stimulation for deafness+, deep brain stimulation for Parkinson’s disease and essential tremor+, sacral nerve stimulation for urinary and fecal incontinence+, spinal cord stimulation for chronic pain+, etc. Rose and Robblee, 1990; de Boer et al., (1978); Negi S et al., (2010); Fan B. et al., (2021)
Titanium Nitride (TiN) Stimulation/Recording 0.3–4.8 −0.9–0.9 2.5–3.0 0.5–0.9 Brain-machine interfaces*, sciatic nerve stimulation and recording*, etc. Gonzalez-Gonzalez et al., (2021); Weiland et al., (2002)
Sputtered Iridium oxide (SIROF) Stimulation 0.1–3.8 −0.6–0.8 18–34 0.4–3.8 Intracortical microstimulation for somatosensory perceptions*, brain-machine interfaces*. Cogan et al., (2004); Negi S et al., (2010)
Activated iridium oxide (AIROF) Stimulation 0.1–3.8 −0.6–0.8 3.0–30 0.3–3.8 Cochlear stimulation for deafness+, intracortical stimulation for somatosensory perceptions+, occipital cortex stimulation for blindness+. Beebe X et al., (1988) Cogan et al. (2004, 2006) Frederick et al., (2020)
Gold (Au) Recording 6.9–33 −0.6–0.9 0.1 1.0–3.2 Motor cortex brain-machine interfaces+, sciatic nerve recordings*, hippocampal recordings*, etc. Fan et al., (2020) Gonzalez-Gonzalez et al., (2021) Kim et al., (2013) Suyatin et al., (2013)
PEDOT Recording 0.1–0.2 −0.9–0.6 0.8–7.0 2.3–15 Motor cortex brain-machine interfaces*, peripheral nerve recordings*. Cogan et al., (2007a) Venkatraman et al., (2011) Hagler et al., (2022) Cui et al., (2007)
Graphene and Graphene oxide Recording 22.5–215 −0.6–0.6 8.0–31 56–106 Peripheral motor interfaces*, In-vitro recording, bidirectional motor cortex brain-machine interfaces*. Chen et al., (2011) Du et al., (2015)

All measurements are vs. Ag|AgCl. Data reported as range min – max; CSC reported at slow sweep rates ≤ 100 mV/s. + indicates clinical and preclinical uses, * indicates preclinical uses only.