Table 2.

Comparison of eight FeLV test systems in

Tests Witness Snap Combo Plus Fastest Duo Speed Virachek FeLV PetChek FeLV One-Step Mapic FeLV
Companies Synbiotics IDEXX MegaCor Bio Veto Test Synbiotics IDEXX Biotech EVL
Countries USA USA Germany France USA USA USA Netherlands
Invalid tests (%) 1.3 0.6 0.2 1.9 0.2 0.4 30.4 13.3
Tests difficult to interpret (%) 13.8 0.4 1.3 1.5 1.5 1.4 7.2 8.4
Sensitivity (%) 92.1 92.3 94.7 94.7 94.9 92.1 nd 96.8
95% CI (sensitivity) 79.7–97.3 79.7–97.3 82.7–98.5 87.2–98.5 83.1–98.6 79.2–97.3 nd 83.9–100
Specificity (%) 97.5 97.3 98.8 99.2 98.4 99.8 nd 95.4
95% CI (specificity) 95.7–98.6 95.5–98.4 97.3–99.4 97.9–99.7 96.8–99.2 97.3–99.4 nd 93.2–96.8
Positive predictive value (%) 74.5 73.5 85.7 90.0 82.2 85.4 nd 62.0
Negative predictive value (%) 99.4 99.4 99.6 99.6 99.6 99.4 nd 99.7
n 528 528 528 528 528 528 378 517

CI=confidence interval; nd=not determined.