Acquisition parameters for the study sequences

Sequence name qDixon VIBE
Sequence type 3D multi-echo Dixon 3D spoiled gradient-echo
Orientation Axial Axial
Repetition time (ms) 15.6 10 (4.2)
Echo time (ms) Range 1.10 to 14.28 with 12 echoes 1.27 (1.48)
Flip angle (°) 4 2/18
Bandwidth (Hz/pixel) 1,090 399 (313)
Matrix 96 × 96 128 × 128
Field of view (mm2) 200 × 200 200 × 200
Slice thickness (mm) 2.5 2.5
Number slices 72 72
Number acquisitions 1 pre- and post-oxygen

Proton density: 3 (5)

Dynamic: 40 (60)

Oxygen on (min:s) - 3:32 (2:09)
Total acquisition time (min:s) 0:58 8:27 (5:35)
Time for single acquisition (s) 29 11.8 (5.15)

Values in parentheses represent the initial imaging protocol reviewed after the first three nonpatient volunteers and two patients