Skip to main content
. 2024 Mar 7;6(2):dlae029. doi: 10.1093/jacamr/dlae029

Table 1.

Results of the NG-Test DetecTool OXA-23 in blood cultures positive with A. baumannii

Hospital True positive True negative False positive False negative
Attikon Athens, clinical 83 0 0 0
Attikon Athens, spiked 0 17 0 0
HC Barcelona,
clinical
HC Barcelona,
spiked
5 19 0 0
Firenze, clinical 5 0 0 0
Firenze, spiked
Paris Saclay, clinical 0 18 0 0
Paris Saclay, spiked 13 19 0 0
Bellvitge, clinical
Bellvitge, spiked 27 23 0 0
Semmelweis Budapest, clinical 4 7 0 0
Semmelweis Budapest, spiked 8 0 0 0
Amiens, clinical 1 0 0 0
Amiens, spiked 50 0 0 0
Total 196 103 0 0