Table 1.

Experimentally measured and simulated heating at the end of TSE sequence, ΔTexp and ΔTsim, and the induced current along the DBS electrodes, Is, for both DBS electrodes in each trajectory and excitation (for electrodes in Traj #1).

Electrode 1 Electrode 2

ΔTexp (°C) ΔTsim (°C) Is1 (mA) ΔTexp (°C) ΔTsim (°C) Is2 (mA)

CP mode
Traj #1 3.8 3.8 ± 1.0 205.9 ± 15.8 0.7 0.8 ± 0.4 79.4 ± 17.3
Traj #2 2.4 2.3 ± 0.6 160.0 ± 14.2 4.8 5.0 ± 1.3 197.8 ± 16.8
Traj #3 1.1 0.7 ± 0.3 85.8 ± 15.8 0.5 0.3 ± 0.2 47.9 ± 15.2
Traj #4 1.7 1.6 ± 0.5 124.8 ± 14.9 1.2 1.2 ± 0.4 111.3 ± 15.9
Traj #5 1.7 2.1 ± 0.6 145.3 ± 14.2 0.2 0.5 ± 0.3 68.6 ± 18.1
Traj #6 0.1 0.1 ± 0.1 29.8 ± 14.9 1.4 1.3 ± 0.7 68.0 ± 17.1
Traj #7 0.1 0.3 ± 0.2 53.7 ± 15.8 0.0 0.2 ± 0.2 42.6 ± 16.6
Traj #8 3.5 3.4 ± 1.0 150.9 ± 15.6 0.3 0.4 ± 0.2 75.1 ± 15.6

Traj #1
Ch 1-only 5.0 5.2 ± 1.1 238.6 ± 9.7 2.6 2.5 ± 0.7 138.8 ± 12.3
Ch 2-only 0.3 0.2 ± 0.1 46.3 ± 12.2 2.4 2.6 ± 0.6 143.2 ± 10.3