Table 3.

Comparison of ABC Values Obtained with Two Different Antibody Lots

Antibody (Lot) ABC Value (Values Adjusted for F/P)
CD19PE (9151856): F/P 1.018
Mean 3,854 (3,785)
SD 163
CV 0.0423
CD19PE (9066849): F/P 0.975
Mean 4,073 (4,177)
SD 83
CV 0.0204
CD20 PE (9176021): F/P 0.897
Mean 14,597 (16,273)
SD 282
CV 0.0193
CD20 PE (9044581): F/P 1.11
Mean 13,001 (11,712)
SD 749
CV 0.0576
CD22 PE (9207363): F/P 1.03
Mean 7,114 (6,906)
SD 64
CV 0.0090
CD22 PE (9038654) F/P 1.11
Mean 6,908 (6,224)
SD 61
CV 0.0088

Mean,Standard Deviation (SD), and Coefficient of Variation (CV) from triplicate Cytotrol cells stained with 2 different lots of CD19PE (clone SJ25C1), CD20PE (clone L27), and CD22PE (clone S-HCL-1) from BD Biosciences. Specimens run on the same day using the same Attune NxT cytometer.