Skip to main content
. Author manuscript; available in PMC: 2025 Jan 12.
Published in final edited form as: Cytometry B Clin Cytom. 2024 Jan 12;106(1):25–34. doi: 10.1002/cyto.b.22155

Table 4.

Comparison of different antibody clones in ABC value determination

Conjugate/Mfr. ABC value (F/P adjusted)
CD19 BC clone J3-119: F/P 1.19
Mean 9,941 (8,354)
S.D. 95
CV 0.0096
CD19 BDIS clone SJ25C1: F/P 1.018
Mean 6,789 (6,669)
S.D. 185
CV 0.0272
CD20 BC clone B9E9: F/P 1.0
Mean 12,398 (12,398)
S.D. 430
CV 0.0346
CD20 BDIS clone L27: F/P 0.897
Mean 19,480 (21,717)
S.D. 397
CV 0.0204
CD22 BC clone SJ10.1H11: F/P 0.8
Mean 5,066 (6,333)
S.D. 69
CV 0.0136
CD22 BDIS clone S-HCL-1: F/P 1.03
Mean 9,710 (9,427)
S.D. 14
CV 0.0014
CD4 BC clone 13B8.2: F/P 1.2
Mean 33,715 (28,096)
S.D. 340
CV 0.0101
CD4 BDIS clone SK3: F/P 1.094
Mean 29,721 (27,167)
S.D. 160
CV 0.0054

Mean, Standard Deviation (SD), and Coefficient of Variation (CV) from triplicate Cytotrol cells stained with different antibody clones from BC-Beckman Coulter and BDIS-BD Bioscences and run on a BDIS FACSCanto.