Skip to main content
. Author manuscript; available in PMC: 2025 Jan 12.
Published in final edited form as: Cytometry B Clin Cytom. 2024 Jan 12;106(1):25–34. doi: 10.1002/cyto.b.22155

Table 6.

Comparison of CD4 Raw GeoMFI and ABC Value Variability Across Instruments

Flow Cytometer CD4 Raw GeoMFI CD4 ABC value
CytoFLEX LX
Mean 75294 24,472
SD 1204 385
CV 0.0160 0.0157
Attune NxT
Mean 47602 23,129
SD 363 181
CV 0.0076 0.0078
CANTO
Mean 18717 25,166
SD 29 39
CV 0.0015 0.0015
Lyric
Mean 27134 26,089
SD 509 488
CV 0.0188 0.0187

Mean, Standard Deviation (SD), and Coefficient of Variation (CV) from triplicate Cytotrol cells stained with CD4PE, clone SK3 and CD45 v500, clone H130 (BDIS). No F/P ratio adjustment was made to the ABC values.