Skip to main content
. 2023 Jun 30;50(1):11–19. doi: 10.1007/s00068-023-02313-w

Table 2.

Difference in position of the plate (mm) and the deviation in screw directions (°) between the planned and the achieved position of plate and screws

Case Fracture classification Position plate Direction screws
∆ Distance (mm) N ∆ Direction (°)
Case 1–Left knee Schatzker 6 2.9 5 3.1 (1.7)
Case 1–Right knee Schatzker 4 3.9 8 3.3 (1.3)
Case 2–Left knee Schatzker 6 4.5 6 2.7 (0.6)
Case 2–Right knee Schatzker 4 2.8 7 6.1 (3.7)
Case 3–Left knee Schatzker 6 3.1 8 3.9 (2.2)
Case 3–Right knee Schatzker 4 2.8 7 4.1 (3.5)
Median (IQR) 3.0 (2.8–3.7) 7 (6.3–7.8) 3.3 (2.5–5.1)