Table 2.
Corrosion values for AS and PF Co–Cr–Mo alloys (SBF, 37 °C)
| Alloy | Ecorr (mV, Ag/AgCl) | Eb (mV, Ag/AgCl) | icorr (mA/cm2) | ip (mA/cm2) | rcorr (mm/y) | Pit |
|---|---|---|---|---|---|---|
| AS | − 191 ± 31 | 514 ± 31 | 3.1 × 10–4 ± 0.8 × 10–4 | 8 × 10–4 ± 2 × 10–4 | 0.0030 ± 0.0008 | No (8/10) |
| PF | − 190 ± 39 | 487 ± 23 | 2.1 × 10–4 ± 0.6 × 10–4 | 7 × 10–4 ± 1 × 10–4 | 0.0021 ± 0.0006 | No (7/10) |
Ecorr corrosion potential, Eb breakdown potential, icorr corrosion current density, ip passive current density and rcorr corrosion rate