Table 2.
SPECT/CT registration QA results before and after alignment of the hybrid-imaging system evaluated with “COR calibration check” and MEGP collimator.
| Date | X | Y | Z |
|---|---|---|---|
| Mean (Standard Deviation) | |||
| 5/2/2022 | 3.04 (0.95) | 0.87 (0.48) | 1.24 (0.68) |
| 6/6/2022 | 0.82 (0.39) | 0.63 (0.45) | 0.91 (0.06) |
| Limit Specs | 2.2 | ||
CT to NM difference in mm. Out-of-tolerance values are in bold letters.