. 2024 Mar 13;8:45. doi: 10.1186/s41747-024-00435-3

Table 2.

Results from PC-MRI-measured flow rate with different acquisition parameters

Protocol Flow rate
(mean and range, mL/s)
SD CoV (%) LDD (mL/s)
Voxel (mm3) NSA VENC (mL/s)
2.5 1 200 7.30 (7.05–7.61) 0.16 2 0.32 (4%)
5.0 1 200 8.74 (8.67–8.83) 0.05 1 0.10 (1%)
7.5 1 200 9.49 (9.20–9.59) 0.12 1 0.24 (3%)
2.5 2 200 7.54 (7.36–7.71) 0.12 2 0.23 (3%)
2.5 3 200 7.73 (7.58–7.90) 0.09 1 0.18 (2%)
2.5 1 300 7.38 (7.08–7.70) 0.21 3 0.41 (5%)
2.5 1 400 7.38 (7.04–7.52) 0.14 2 0.28 (4%)
5.0 3 200 8.41 (8.33–8.50) 0.07 1 0.13 (2%)
7.5 3 200 9.48 (9.33–9.58) 0.08 1 0.15 (2%)
Reference standard 6.59 (6.50–6–70) 0.08 1 0.15 (2%)

CoV Coefficient of variation, LDD Least detectable difference, in both absolute and in relative values (given as % of the mean), NSA Number of signal averages, PC-MRI Phase-contrast magnetic resonance imaging, SD Standard deviation, VENC Velocity encoding