Table 4.

Bland–Altman analysis and paired t-tests show the level of agreement between OCTAVA and in-built commercial OCTA software packages for measuring retinal vessel area density, vessel length, and foveal avascular zone metrics.

Instrument and participant count Metric OCTAVA In-built software Bland–Altman analysis Spearman correlation (p-value)
Mean (SD) Mean (SD) Mean SD Lower LA Upper LA tstat p-value

Cirrus 5000

n = 32

Total VAD [%] 36.5 (1.9) 35.6 (2.1)  +0.9 1.72 -2.4  +4.3  +3.1  < 0.01 0.630 (p < 0.01)
Parafoveal VAD [%] 37.4 (2.2) 37.9 (2.1) -0.5 1.77 -3.9  +3.0  -1.5  < 0.01 0.667 (p < 0.01)
Foveal VAD [%] 19.1 (4.2) 17.5 (4.5)  +1.6 0.17   +0.1  +1.0  +21.6  < 0.01 0.905 (p < 0.01)
Total VL [mm] 185 (11) 183 (10)  +2 7.5 -12.8  +16.5  +1.4 0.17 0.724 (p < 0.01)
Parafoveal VL [mm] 132.5 (8.6) 152.3 (7.9) -19.8 5.5 -30.5 -8.9 -20.3  < 0.01 0.792 (p < 0.01)
Foveal VL [mm] 7.7 (5.6) 8.4 (2.1) -0.7 5.5 -11.4  +10.0  -0.7 0.48 0.388 (p = 0.28)
FAZA [mm2] 0.275 (0.107) 0.275 (0.11)  -0.001 0.019 -0.04  +0.04  -2.7 0.79 0.923 (p < 0.01)
FAZ perimeter [mm] 2.29 (0.52) 2.17 (0.43)  +0.11 0.28 -0.43  +0.66  +2.3 0.03 0.799 (p < 0.01)
FAZ circularity 0.67 (0.15) 0.71 (0.07) -0.04 0.11 -0.25  +0.16  -2.3 0.03 0.708 (p < 0.01)
Revo NX 130 n = 30 Total VAD [%] 31.9 (3.7) 40.3 (1.2) -8.3 3.4 -15.4 -1.3  -12.7  < 0.01 0.192 (p = 0.31)
Total VL [mm] 178 (20) 208 (11) -30 19 -68  +6  -8.9  < 0.01 0.406 (p = 0.03)
FAZA [mm2] 0.322 (0.144) 0.342 (0.163) -0.02 0.04 -0.091  +0.051  -3.0  < 0.01 0.955 (p < 0.01)
FAZ perimeter [mm] 2.28 (0.59) 2.48 (0.69) -0.20 0.38 -0.95  +0.55  -2.9  < 0.01 0.843 (p < 0.01)
FAZ circularity 0.77 (0.17) 0.58 (0.097)  +0.20 0.19  -0.17  +0.56 +5.75  < 0.01 0.176 (p = 0.35)

RTVue-XR

n = 31

Total VAD [%] 30 (3) 50 (2) -20 1.7 -23.7 -17.1 -43.6  < 0.01 0.705 (p < 0.01)
FAZA [mm2] 0.213 (0.111) 0.212 (0.11)  +0.001 0.015 -0.027  +0.030  +0.5 0.63 0.992 (p < 0.01)
FAZ perimeter [mm] 2.09 (0.7) 1.77 (0.49)  +0.32 0.37 -0.40  +1.04  +4.8  < 0.01 0.884 (p < 0.01)
FD-300 [%] 34.3 (4) 50 (4)  -15.6 2.1 -19.7 -116  -42.2  < 0.01 0.783 (p < 0.01)

OCTAVA metrics were generated with Frangi filter maximum kernel size 4 pixels and twig size 8 pixels.

VAD vessel area density, VL vessel length, FAZA foveal avascular zone area.