Skip to main content
. 2024 Mar 14;15:200–218. doi: 10.18632/oncotarget.28565

Table 6. Precision summary.

Sample Total runs Mean concentration, measured (PPM) Standard deviation (PPM) Coefficient of variation (%)
High concentration (~25,000 PPM) 24 25,017 1,038.2 4.15%
Medium concentration (~1,000 PPM) 24 991.0 35.18 3.55%
Low concentration (~25 PPM) 24 26.02 3.323 12.77%