. 2024 Feb 6;129(3):401–410. doi: 10.1007/s11547-024-01773-3

Table 2.

Measured dose parameters for the coronary vessel scanning with and without iodinated lumen, including three repetitions and three scan modes per heart rate

Phantom With iodine Without iodine
Heart rates 0 bpm 60 bpm 80 bpm 100 bpm 0 bpm
Pitch factor 0.8 (0.2–3.2) 0.8 (0.2–3.2) 0.8 (0.3–3.2) 0.8 (0.3–3.2) 0.8 (0.2–3.2)
Eff. mAs 38 38 38 38 38
CTDIvol (mGy) 5.9 (2.7–13.7) 6.2 (2.8–43.8) 6.2 (2.7–33.7) 6.1 (2.7–27.4) 6.1 (2.8–43.8)
DLP (mGy*cm) 55.8 (47.8–176.0) 86.9 (48.0–480.0) 86.9 (47.9–507.0) 86.5 (47.9–415.0) 86.6 (61.3–652.0)
SSDE (mGy) 9.7 (4.4–24.3) 10.1 (4.4–70.1) 10.1 (4.4–53.9) 10.1 (4.4–43.8) 10.1 (4.5–69.9)

Values are median (interquartile range). CTDIvol = volumetric CT dose index, DLP = dose length product, SSDE = size-specific dose estimate