Repeated measures ANOVA comparing the within-participant effect of modality (OPM, SQUID-MEG) for the data quality measures, separately for the SQUID-MEG-compatible and -incompatible participants.

Measure Mean [SD] Statisticsa
OPM-MEG SQUID-MEG F p-value ηp2
SQUID-MEG-compatible # trials 77.5 [6.0] 77.9 [2.7] 0.09 .766 .01
Head motion (mm) 1.0 [0.3] 1.4 [1.3] 2.26 .157 .15
SQUID-MEG-incompatible # trials 67.8 [13.7] 0.0 [0.0] 122.45  < .001 .97
Head motion (mm) 0.9 [0.2]

SD standard deviation; OPM-MEG: optically- pumped magnetometer-magnetoencephalography; SQUID-MEG: superconducting quantum interference device-magnetoencephalography aRepeated measures ANOVA.