Comparison of the electrochemical sensors for the detection of ammonium ions.

Sensing material Detection range Detection limit Method of detection Detection medium Types of sample Ref.
PMB@GLDH/AuNPs/SPEC 0.65–300 μM N/A Amperometry PBS (pH 9.5) Water 148
Ag-CNT 0.2–1 mM 1 μM DPV 0.1 M Na2SO4 Groundwater 141
Ppy COSANE-Au microelectrode 1 μM to 40 mM 0.04 mM Potentiometric Buffer Tris-HCl/Tris Tap/sewage water 142
NH4+ selective ASS electrode 1 μM to 0.1 M <1 μM Potentiometric N/A Tap/well water 143
Solid state ISM 1–64 mg N/L 1 × 10−6.5 M Potentiometric N/A Wastewater 144
Natural zeolite clinoptilolite-Au interdigitated microelectrode (IDμE) 0.01 μM to 1 mM 0.01 μM Conductometric 5 mM KH2PO4/Na2HPO4 (pH 6.2) N/A 145
25,27-Di-(5-thio-octyloxy)calix[4] arene-crown-6-Au interdigitated electrode (IDE) 0.01–1 mM 0.01 mM Conductometric 5 mM KH2PO4/Na2HPO4 (pH 6.2) River water 146
Alanine dehydrogenase/SPE 10–100 mM 0.18 mM Amperometric PBS (pH 7) N/A 150
PET microporous membrane 4.2–66 μM 0.12 μM CV, DPV, SWV Water/1,6-dichlorohexane Water 149