Skip to main content
. Author manuscript; available in PMC: 2025 May 1.
Published in final edited form as: Magn Reson Med. 2023 Dec 26;91(5):2057–2073. doi: 10.1002/mrm.29981

Table 6:

Image quality metrics examining SNR and CNR in the renal vein or inferior vena cava for K-MOTIVE and db-K-MOTIVE sequences at various locations. SNR was computed in the image corresponding to TE=0ms, where SNR=μROI/μbackground, i.e., the mean signal intensity in ROI (vein or static tissue) divided by the mean signal intensity of background air. Contrast-to-noise ratio (CNR) was calculated as a difference between the SNR of vein and surrounding static tissue. Metrics were not significantly different among all pulse sequences.

Image Quality SNR=μROIμbackground CNR=SNRvein–SNRtissue
Vein Surrounding tissue
K-MOTIVE-left 15 ± 8 2.6 ± 0.9 13 ± 7
K-MOTIVE-right 15 ± 6 2.6 ± 1.1 12 ± 6
K-MOTIVE-supra 18 ± 9 3.0 ± 1.0 15 ± 8
K-MOTIVE-infra 24 ± 10 3.6 ± 1.7 20 ± 9
db-K-MOTIVE-supra 17 ± 9 3.4 ± 1.3 14 ± 9
db-K-MOTIVE-infra 23 ± 14 3.6 ± 1.7 19 ± 14