Table 2.
Results of syringe pump tests using the adjusted volumes derived from linear regressions 2 with n=10 for each tested volume. The error quickly decreases as the volume increases.
Syringe/Pipette tip | Target volume (L) | Adjusted volume (L) | Accuracy (%) | Precision (%) |
---|---|---|---|---|
50 L + P20 tip | 1.00 | 1.45 | 5.00% | 11.22% |
50 L + P20 tip | 2.00 | 2.52 | 12.50% | 7.03% |
50 L + P20 tip | 5.00 | 5.13 | 2.60% | 6.03% |
50 L + P20 tip | 10.00 | 11.13 | 0.80% | 5.17% |
50 L + P20 tip | 20.00 | 21.89 | 1.05% | 0.97% |
250 L + P200 tip | 10.00 | 11.12 | 2.60 % | 1.47 % |
250 L + P200 tip | 25.00 | 26.62 | 0.32 % | 0.39 % |
250 L + P200 tip | 50.00 | 52.47 | 0.74 % | 0.36 % |
250 L + P200 tip | 100.00 | 104.15 | 0.93 % | 4.80 % |
250 L + P200 tip | 200.00 | 207.52 | 0.95 % | 0.32 % |